Background: The usage of multichannel brain MRI coils, which have several advantages over single-channel brain coils used for stereotactic radiosurgery (SRS), requires a frame adapter device to fit the frames inside the multichannel brain coils. However, such a frame adapter has not been developed until now. Objective: to develop an SRS frame adapter for multichannel MRI coils and verify the geometrical accuracy and signal-to-noise ratio (SNR) of the MR images obtained using multichannel MRI coils. Methods: We fabricated an SRS frame adapter for a 48-channel MRI coil using a three-dimensional (3D) printer. Furthermore, we obtained phantom and human-brain MR images with a 3.0 Tesla MRI scanner using multi- and single-channel coils. Computed tomography (CT) phantom images were also obtained as reference. We compared the coordinate errors of the multi- and single-channel coils to evaluate the geometrical accuracy. Two neurosurgeons measured the coordinates. In addition, we compared the SNR differences between multi- and single-channel coils using the T1- and T2-weighted brain images. Results: For the CT coordinate measurements, the correlation coefficient r = 1 and p < 0.001 with respect to the 3 axes (Δx, Δy, and Δz) and 3D errors (Δr) showed no interpersonal differences between the 2 neurosurgeons. The results obtained using the T1-weighted images showed that a multichannel coil had smaller coordinate errors in Δx, Δy, Δz, and Δr than that observed in case of a single-channel coil (p < 0.001). In case of the SNR measurements, most of the brain areas showed higher SNRs when using a multichannel coil compared with that observed when using a single-channel coil in the T1- and T2-weighted images. Conclusion: Compared with single-channel coils, the use of multichannel MRI coils with a newly developed frame adapter is expected to ensure successful SRS treatments with improved geometrical accuracy and SNR.

This content is only available via PDF.
Copyright / Drug Dosage / Disclaimer
Copyright: All rights reserved. No part of this publication may be translated into other languages, reproduced or utilized in any form or by any means, electronic or mechanical, including photocopying, recording, microcopying, or by any information storage and retrieval system, without permission in writing from the publisher.
Drug Dosage: The authors and the publisher have exerted every effort to ensure that drug selection and dosage set forth in this text are in accord with current recommendations and practice at the time of publication. However, in view of ongoing research, changes in government regulations, and the constant flow of information relating to drug therapy and drug reactions, the reader is urged to check the package insert for each drug for any changes in indications and dosage and for added warnings and precautions. This is particularly important when the recommended agent is a new and/or infrequently employed drug.
Disclaimer: The statements, opinions and data contained in this publication are solely those of the individual authors and contributors and not of the publishers and the editor(s). The appearance of advertisements or/and product references in the publication is not a warranty, endorsement, or approval of the products or services advertised or of their effectiveness, quality or safety. The publisher and the editor(s) disclaim responsibility for any injury to persons or property resulting from any ideas, methods, instructions or products referred to in the content or advertisements.
You do not currently have access to this content.