Objective: To establish the impact of the imaging modality, registration method and use of intraoperative computed tomography (CT) scan on the accuracy of the ROSA® stereotactic robot. Methods: Using a dedicated phantom device, we measured the accuracy of the stereotactic robot for 20 targets as a function of the registration method (frameless, FL, or frame-based, FB) and the reference imaging modality (3T magnetic resonance imaging, MRI, CT scanner or flat-panel CT, fpCT). We performed a retrospective study of the accuracy of the first 26 FB and 31 FL robotized stereotactic surgeries performed in our department. Results: In a phantom study, the mean target accuracy was 1.59 mm for 3T MRI-guided FL surgery, 0.3 mm for fpCT-guided FL surgery and 0.3 mm for CT-guided FB surgery. In our retrospective series, the mean accuracy was 0.81 mm for FB stereotactic surgery, 1.22 mm for our 24 stereotactic surgery procedures with FL (surface recognition) registration and 0.7 mm for our 7 stereotactic surgery procedures with FL fiducial marker registration. Intraoperative fpCT fully corrected all the registration errors. Conclusions: The ROSA stereotactic robot is highly accurate. Robotized FB stereotactic surgery is more accurate than robotized FL stereotactic surgery.

1.
Lozano AM, Gildenberg PL, Tasker RR (eds): Textbook of Stereotactic and Functional Neurosurgery. Berlin, Springer, 2009.
2.
Dorward NL, Alberti O, Palmer JD, Kitchen ND, Thomas DG: Accuracy of true frameless stereotaxy: in vivo measurement and laboratory phantom studies. J Neurosurg 1999;90:160-168.
3.
Eljamel MS: Robotic neurological surgery applications: accuracy and consistency or pure fantasy? Stereotact Funct Neurosurg 2009;87:88-93.
4.
Gralla J, Nimsky C, Buchfelder M, Fahlbusch R, Ganslandt O: Frameless stereotactic brain biopsy procedures using the Stealth Station: indications, accuracy and results. Zentralbl Neurochir 2003;64:166-170.
5.
Hartov A, Roberts DW: Accuracy in stereotactic and image guidance; in Lozano AM, Gildenberg PL, Tasker RR (eds): Textbook of Stereotactic and Functional Neurosurgery. Berlin, Springer, 2009, pp 443-452.
6.
Landi A, Marina R, DeGrandi C, Crespi A, Montanari G, Sganzerla EP, Gaini SM: Accuracy of stereotactic localisation with magnetic resonance compared to CT scan: experimental findings. Acta Neurochir 2001;43:593-601.
7.
Pfisterer WK, Papadopoulos S, Drumm DA, Smith K, Preul MC: Fiducial versus nonfiducial neuronavigation registration assessment and considerations of accuracy. Neurosurgery 2008;62:201-207.
8.
Pezeshkian P, DeSalles AAF, Gorgulho A, Behnke E, McArthur D, Bari A: Accuracy of frame-based stereotactic magnetic resonance imaging versus frame-based stereotactic head computed tomography fused with recent magnetic resonance imaging for postimplantation deep brain stimulator lead localization. Neurosurgery 2011;69:1299-1306.
9.
Novotný J, Vymazal J, Liščák R, Vladyka V: Assessment of the accuracy of stereotactic target localization using magnetic resonance imaging: a phantom study. J Radiosurg SBRT 1998;1:99-111.
10.
Maurer CR, Aboutanos GB, Dawant BM, Gadamsetty S, Margolin RA, Maciunas RJ, Fitzpatrick JM: Effect of geometrical distortion correction in MR on image registration accuracy. J Comput Assist Tomogr 1996;20:666-679.
11.
Maciunas RJ, Galloway RLJ, Latimer JW: The application accuracy of stereotactic frames. Neurosurgery 1994;35:682-695.
12.
Li QH, Zamorano L, Pandya A, Perez R, Gong J, Diaz F: The application accuracy of the NeuroMate robot - a quantitative comparison with frameless and frame-based surgical localization systems. Comput Aided Surg 2002;7:90-98.
13.
Hariz MI, Zrinzo L: CT/MRI technology: basic principles; in Lozano AM, Gildenberg PL, Tasker RR (eds): Textbook of Stereotactic and Functional Neurosurgery. Berlin, Springer, 2009, pp 269-278.
14.
Eljamel MS: Validation of the PathFinder neurosurgical robot using a phantom. Int J Med Robot 2007;3:372-377.
15.
Lefranc M, Le Gars D: Robotic implantation of deep brain stimulation leads, assisted by intra-operative, flat-panel CT. Acta Neurochir 2012;154:2069-2074.
16.
Kall BA: Image reconstruction and fusion; in Lozano AM, Gildenberg PL, Tasker RR (eds): Textbook of Stereotactic and Functional Neurosurgery. Berlin, Springer, 2009, pp 335-343.
17.
Hemm S, Coste J, Gabrillargues J, Ouchchane L, Sarry L, Caire F, Vassal F, Nuti C, Derost P, Durif F, Lemaire J-J: Contact position analysis of deep brain stimulation electrodes on post-operative CT images. Acta Neurochir (Wien) 2009;151:823-829.
18.
O'Gorman RL, Jarosz JM, Samuel M, Clough C, Selway RP, Ashkan K: CT/MR image fusion in the postoperative assessment of electrodes implanted for deep brain stimulation. Stereotact Funct Neurosurg 2009;87:205-210.
19.
Slavin KV, Thulborn KR, Wess C, Nersesyan H: Direct visualization of the human subthalamic nucleus with 3T MR imaging. AJNR Am J Neuroradiol 2006;27:80-84.
20.
Langlois S, Desvignes M, Constans JM, Revenu M: MRI geometric distortion: a simple approach to correcting the effects of non-linear gradient fields. J Magn Reson Imaging 1999;9:821-831.
21.
Novotny J, Vymazal J, Novotny J, Tlachacova D, Schmitt M, Chuda P, Urgosik D, Liscak R: Does new magnetic resonance imaging technology provide better geometrical accuracy during stereotactic imaging? J Neurosurg 2005;102:8-13.
22.
Walton L, Hampshire A, Vaughan P, Forster DM, Kemeny AA, Radatz MW: Distortion in magnetic resonance images obtained for stereotactic localization. Case report. J Neurosurg 2000;93:191-192.
23.
Walton L, Hampshire A, Forster DM, Kemeny AA: Stereotactic localization with magnetic resonance imaging: a phantom study to compare the accuracy obtained using two-dimensional and three-dimensional data acquisitions. Neurosurgery 1997;41:131-137.
24.
Wang D, Doddrell DM, Cowin G: A novel phantom and method for comprehensive 3-dimensional measurement and correction of geometric distortion in magnetic resonance imaging. Magn Reson Imaging 2004;22:529-542.
25.
Schicho K, Figl M, Seemann R, Donat M, Pretterklieber ML, Birkfellner W, Reichwein A, Wanschitz F, Kainberger F, Bergmann H, Wagner A, Rolf Ewers R: Comparison of laser surface scanning and fiducial marker-based registration in frameless stereotaxy. J Neurosurg 2007;106:704-709.
26.
Quiñones-Hinojosa A, Ware M, Sanai N, McDermott M: Assessment of image guided accuracy in a skull model: comparison of frameless stereotaxy techniques vs. frame-based localization. J Neurooncol 2006;76:65-70.
27.
Owen CM, Linskey ME: Frame-based stereotaxy in a frameless era: current capabilities, relative role, and the positive and negative predictive values of blood through the needle. J Neurooncol 2009;93:139-149.
28.
Walton L, Hampshire A, Forster DM, Kemeny AA: A phantom study to assess the accuracy of stereotactic localization, using T1-weighted magnetic resonance imaging with the Leksell stereotactic system. Neurosurgery 1996;38:170-176.
29.
Lunsford LD, Kondziolka D, Leksell D: Leksell stereotactic apparatus; in Lozano AM, Gildenberg PL, Tasker RR (eds): Textbook of Stereotactic and Functional Neurosurgery. Berlin, Springer, 2009, pp 469-485.
30.
Holl EM, Petersen EA, Foltynie T, Martinez-Torres I, Limousin P, Hariz MI, Zrinzo L: Improving targeting in image-guided frame-based deep brain stimulation. Neurosurgery 2010;67:437-447.
31.
Mascott CR, Sol J-C, Bousquet P, Lagarrigue J, Lazorthes Y, Lauwers-Cances V: Quantification of true in vivo (application) accuracy in cranial image-guided surgery: influence of mode of patient registration. Neurosurgery 2006;59:146-156.
32.
Giese H, Hoffmann K-T, Winkelmann A, Stockhammer F, Jallo GI, Thomale U-W: Precision of navigated stereotactic probe implantation into the brainstem. J Neurosurg Pediatr 2010;5:350-359.
33.
Widmann G, Schullian P, Ortler M, Bale R: Frameless stereotactic targeting devices: technical features, targeting errors and clinical results. Int J Med Robot 2012;8:1-16.
34.
Gupta R, Cheung AC, Bartling SH, Lisauskas J, Grasruck M, Leidecker C, Schmidt B, Flohr T, Brady TJ: Flat-panel volume CT: fundamental principles, technology, and applications. Radiographics 2008;28:2009-2022.
35.
Jonker BP: Image fusion pitfalls for cranial radiosurgery. Surg Neurol Int 2013;4:123-128.
36.
Rahimian J, Chen JC, Rao AA, Girvigian MR, Miller MJ, Greathouse HE: Geometrical accuracy of the Novalis stereotactic radiosurgery system for trigeminal neuralgia. J Neurosurg 2004;101:351-355.
37.
Hamm KD, Surber G, Schmücking M, Wurm RE, Aschenbach R, Kleinert G, Niesen A, Baum RP: Stereotactic radiation treatment planning and follow-up studies involving fused multimodality imaging. J Neurosurg 2004;101:326-333.
38.
Cohen DS, Lustgarten JH, Miller E, Khandji AG, Goodman RR: Effects of coregistration of MR to CT images on MR stereotactic accuracy. J Neurosurg 1995;82:772-779.
39.
Petersen EA, Holl EM, Martinez-Torres I, Foltynie T, Limousin P, Hariz MI, Zrinzo L: Minimizing brain shift in stereotactic functional neurosurgery. Neurosurgery 2010;67:213-221.
40.
Smith JS, Quiñones-Hinojosa A, Barbaro NM, McDermott MW: Frame-based stereotactic biopsy remains an important diagnostic tool with distinct advantages over frameless stereotactic biopsy. J Neurooncol 2005;73:173-179.
41.
Woodworth GF, McGirt MJ, Samdani A, Garonzik I, Olivi A, Weingart JD: Frameless image-guided stereotactic brain biopsy procedure: diagnostic yield, surgical morbidity, and comparison with the frame-based technique. J Neurosurg 2006;104:233-237.
42.
Dammers R, Haitsma IK, Schouten JW, Kros JM, Avezaat CJ, Vincent AJ: Safety and efficacy of frameless and frame-based intracranial biopsy techniques. Acta Neurochir (Wien) 2008;150:23-29.
43.
McGirt MJ, Woodworth GF, Coon AL, Frazier JM, Amundson E, Garonzik I, Olivi A, Weingart JD: Independent predictors of morbidity after image-guided stereotactic brain biopsy: a risk assessment of 270 cases. J Neurosurg 2005;102:897-901.
44.
Benabid A-L, Chabardes S, Mitrofanis J, Pollak P: Deep brain stimulation of the subthalamic nucleus for the treatment of Parkinson's disease. Lancet Neurol 2009;8:67-81.
45.
Deuschl G, Schade-Brittinger C, Krack P, Volkmann J, Schäfer H, Bötzel K, Daniels C, Deutschländer A, Dillmann U, Eisner W, Gruber D, Hamel W, Herzog J, Hilker R, Klebe S, Kloss M, Koy J, Krause M, Kupsch A, Lorenz D, Lorenzl S, Mehdorn HM, Moringlane JR, Oertel W, Pinsker MO, Reichmann H, Reuss A, Schneider G-H, Schnitzler A, Steude U, Sturm V, Timmermann L, Tronnier V, Trottenberg T, Wojtecki L, Wolf E, Poewe W, Voges J: A randomized trial of deep-brain stimulation for Parkinson's disease. N Engl J Med 2006;355:896-908.
Copyright / Drug Dosage / Disclaimer
Copyright: All rights reserved. No part of this publication may be translated into other languages, reproduced or utilized in any form or by any means, electronic or mechanical, including photocopying, recording, microcopying, or by any information storage and retrieval system, without permission in writing from the publisher.
Drug Dosage: The authors and the publisher have exerted every effort to ensure that drug selection and dosage set forth in this text are in accord with current recommendations and practice at the time of publication. However, in view of ongoing research, changes in government regulations, and the constant flow of information relating to drug therapy and drug reactions, the reader is urged to check the package insert for each drug for any changes in indications and dosage and for added warnings and precautions. This is particularly important when the recommended agent is a new and/or infrequently employed drug.
Disclaimer: The statements, opinions and data contained in this publication are solely those of the individual authors and contributors and not of the publishers and the editor(s). The appearance of advertisements or/and product references in the publication is not a warranty, endorsement, or approval of the products or services advertised or of their effectiveness, quality or safety. The publisher and the editor(s) disclaim responsibility for any injury to persons or property resulting from any ideas, methods, instructions or products referred to in the content or advertisements.
You do not currently have access to this content.