Abstract
An automated stereotactic machine has been interfaced to a surgical computer to complete a totally interactive surgical system capable of locating tumor volumes. Stepper motors, activated by the host computer, drive a three-dimensional slide to position the patient''s head with respect to a fixed arc, locating the surgical target. Linear encoders on each axis create a closed-loop positioning system and a digital display for visual inspection of the slide''s position. The 160-mm arc directs all instrumentation to its isocenter, regardless of the two angular settings, providing maximum freedom in selecting a safe trajectory to the target. Phantom test points compatible with computerized tomographic and magnetic resonance imaging were repeatedly scanned to determine the overall system accuracy, which approached 0.6 mm, depending on the spatial resolution of the image. This stereotactic device may be used to perform stereotactic laser craniotomies, biopsies, 192Ir implants for interstitial radiation, third ventriculostomies and functional procedures.