Background: So far, 3 randomized controlled trials have shown that the endobronchial treatment using coils is safe and effective. However, the more exact underlying mechanism of the treatment and best predictors of response are unknown. Objectives: The aim of the study was to gain more knowledge about the underlying physiological mechanism of the lung volume reduction coil treatment and to identify potential predictors of response to this treatment. Methods: This was a prospective nonrandomized single-center study which included patients who were bilaterally treated with coils. Patients underwent an extensive number of physical tests at baseline and 3 months after treatment. Results: Twenty-four patients (29% male, mean age 62 years, forced expiratory volume in 1 s [FEV1] 26% pred, residual volume (RV) 231% pred) were included. Three months after treatment, significant improvements were found in spirometry, static hyperinflation, air trapping, airway resistance, treated lobe RV and treated lobes air trapping measured on CT scan, exercise capacity, and quality of life. The change in RV and airway resistance was significantly associated with a change in FEV1, forced vital capacity, air trapping, maximal expiratory pressure, dynamic compliance, and dynamic hyperinflation. Predictors of treatment response at baseline were a higher RV, larger air trapping, higher emphysema score in the treated lobes, and a lower physical activity level. Conclusions: Our results confirm that emphysema patients benefit from endobronchial coil treatment. The primary mechanism of action is decreasing static hyperinflation with improvement of airway resistance which consequently changes dynamic lung mechanics. However, the right patient population needs to be selected for the treatment to be beneficial which should include patients with severe lung hyperinflation, severe air trapping, and significant emphysema in target lobes.

Emphysema is characterized by gradual destruction of alveolar walls. This results in reduced lung elasticity and recoil pressures and allows the smaller airways to collapse prematurely during exhalation, resulting in hyperinflation, air trapping, and diaphragm flattening [1]. In the past decade, less invasive bronchoscopic treatments have been developed, and the lung volume reduction coil (LVRC) is one of those approaches [2].

Currently, 3 randomized controlled trials have been performed which showed that the coil treatment was safe and effective [3-5]. However, the exact underlying working mechanism of the coil treatment is unknown. The proposed mechanism of action is that the compression of the lung parenchyma by the coils reduces air flow to treated portions of the lung, allowing enhanced airflow to healthier untreated portions of the lung. Recently, we found that the reduction in lobar residual volume (RV) was the driving mechanism of action [6]. This lobar compression also reduces the volume of the hyperinflated emphysematous lung, resulting in lung volume reduction with possible improved diaphragmatic efficiency. Additionally, by gathering up the loose parenchyma of the most severely damaged segments, the coil may restore elasticity and recoil to the whole lung, improving expiratory flow rates, reducing small airway collapse with subsequent air trapping, and reducing dynamic hyperinflation. Therefore, to advance the understanding of changes in lung physiology, it would be very useful to investigate dynamic hyperinflation, compliance/elastic recoil, diaphragm functioning, ventilation and small airways/air trapping parameters more in depth in patients who are treated with LVRCs.

Another less well-understood aspect of the LVRC treatment is which group of patients benefits of this treatment. So far, a clinical significant response to the coil treatment has been reported in about 5–6 of every 10 patients treated [7]. To maximize the responder rate and avoid unnecessary treatments, it would be very useful if we could identify the responders beforehand. Recently, the correct choice of the most destructed target lobes to treat, significant hyperinflation (RV ≥200% of predicted), emphysema score >20% low attenuation area at −950 Hounsfield units, and absence of airway disease were found to be predictors of better response [8]. Therefore, the aim of this study was to gain more knowledge on the underlying physiological mechanism of the LVRC treatment and to identify potential predictors of response to this treatment.

Study Population

This was a prospective nonrandomized intervention study (“REACTION-study”: NCT02179125) including symptomatic patients with severe emphysema (forced expiratory volume in 1 s [FEV1]% pred <45%, RV% pred >175%) between March 9, 2015 and July 4, 2018. The study was originally designed as a multicenter trial but due to regulatory issues in the UK performed as 2 separate trials. The complete list of inclusion and exclusion criteria can be found in the online suppl. Table 1; see www.karger.com/doi/10.1159/000514319 for all online suppl. material. Patients were included in the analyses when they were bilaterally treated and visited our hospital for the study tests at baseline and 3 months after the second treatment. The study was approved by the Local Medical Ethics Committee (METc#2014.447), and all patients provided informed consent.

Study Procedure

Patients visited our hospital for the study measurements described below at baseline and at 3 months after the second treatment. After screening, patients underwent the first coil treatment and 6–8 weeks later the second treatment. The coils were placed in the most damaged lobes per lung based on a visual CT analysis by the treating physician.

Study Measurements

Spirometry, body plethysmography, helium dilution, and diffusion capacity for carbon monoxide (DLCO) were performed according to the ATS/ERS guidelines [9-11]. An esophageal balloon was used to measure static and dynamic lung compliance (Cooper-Surgical, Berlin, Germany). Furthermore, maximal inspiratory pressure and maximal expiratory pressure (MEP) were measured using body plethysmography [12, 13] and dynamic hyperinflation with a metronome-paced test (used protocol is described in reference [14]). Exercise capacity was measured by a 6-minute walk distance (6MWD) test and an incremental cycle ergometry test according to the ATS guidelines [15, 16]. Physical activity was measured for 1 week by the Dynaport Movemonitor (McRoberts, Hague, The Netherlands) [17]. St. George’s Respiratory Questionnaire (SGRQ) was administered to measure quality of life [18]. Lung perfusion was measured by a perfusion scan using 99mTc albumin. An inspiratory and expiratory high-resolution computed tomography scan was performed, and quantitative analyses were performed by FLUIDDA Inc., (Kontich, Belgium). Outcomes were lobar volumes, emphysema destruction scores, airway radius, and air trapping (defined as all intrapulmonary voxels between −1,024 and −850 Hounsfield units, on the expiratory high-resolution computed tomography scan). Furthermore, on CT scan, a blinded visual assessment of the presence or absence of coil-associated opacity (CAO) was performed by DJS.

Statistical Analyses

All analyses were performed with IBM-SPSS statistics version 23(IBM, Armonk, NY, USA). p values <0.05 were considered statistically significant. Safety was reported by the number and type of adverse events. A paired t test (normally distributed data) or a Wilcoxon signed rank test (non-normal data distribution) was performed to investigate the difference between baseline and 3-month follow-up in all clinical variables. To calculate the number of responders, we used the following established minimal important differences: FEV1: 100 mL [19], RV: 310 mL [20], 6MWD: 26 m [21], SGRQ: 8 units [22], steps per day: 600 steps [23, 24]. Differences between responders in RV were tested by an independent t test or Mann-Whitney U test (non-normal distributed data). To investigate the mechanism of action, we calculated the association between the change in RV and change in airway resistance with change in other clinical outcomes using Pearson’s correlation coefficient or Spearman’s rho (non-normal distributed data). Furthermore, we performed an independent t test to investigate whether patients with CAO have better response in clinical outcomes compared to patients without CAO. To investigate the predictors of response, we calculated the association between change in RV and change in airway resistance and potential predictors of response at baseline using Pearson’s correlation coefficient or Spearman’s rho.

In total, 24 patients (29% male, mean age 62 years, FEV1 26% pred, RV 231% pred) were included in the analyses. Patient characteristics can be found in Table 1, with the study flowchart in Figure 1.

Table 1.

Baseline characteristics and change after 3-month follow-up (n = 24)

Baseline characteristics and change after 3-month follow-up (n = 24)
Baseline characteristics and change after 3-month follow-up (n = 24)
Fig. 1.

Study flowchart. RV, residual volume; 6MWD, 6-min walk distance; TLC, total lung capacity.

Fig. 1.

Study flowchart. RV, residual volume; 6MWD, 6-min walk distance; TLC, total lung capacity.

Close modal

Procedure Characteristics and Safety

Twenty-one patients (88%) were treated in both upper and 3 patients in both lower lobes. The median procedure time was 35 (range: 20–90) min, and the median number of coils placed was 21.5 (19–25) (right lung: 11 [9–13], left lung: 11 [10–12]). Median time between treatments was 48 (35–63) days. All reported (serious) adverse events of all patients treated unilaterally or bilaterally (n = 29) can be found in Table 2. In total, 11 serious adverse events occurred during the study period, of which 3 were probably or possibly related to the treatment (2 COPD exacerbations and 1 pneumonia). One patient died 84 days after the first treatment due to progressive hypercapnic respiratory failure secondary to disease progression which was not deemed to be related to the treatment.

Table 2.

Overview of all reported (serious) adverse events (n = 29)

Overview of all reported (serious) adverse events (n = 29)
Overview of all reported (serious) adverse events (n = 29)

Efficacy

Changes between baseline and 3-month follow-up for all clinical variables can be found in Table 1. Significant improvements were found in spirometry (FEV1 and forced vital capacity [FVC]), static hyperinflation, air trapping (measured by the total lung capacity (TLC) difference between helium and body plethysmography), airway resistance, treated lobes RV and treated lobes air trapping (both measured on CT scan), exercise capacity (6MWD and cycle ergometry), and quality of life (SGRQ). No significant changes were found in DLCO, maximal inspiratory pressure and MEP, lung compliance, dynamic hyperinflation, lung perfusion, physical activity, and airway radius. The responder rates were RV: 67%, FEV1: 46%, 6MWD: 48%, SGRQ: 58%, and steps/day: 33%. The responders based on RV had a significant better improvement than the nonresponders in FVC, air trapping (measured by the TLC difference between helium and body plethysmography), alveolar volume/TLC, MEP, dynamic compliance, airway resistance, and exercise capacity (cycle ergometry) (online suppl. Table 2).

Mechanism of Action

The change in RV and airway resistance was significantly associated with a change in FEV1, FVC, air trapping (TLC difference between body plethysmography and helium), MEP, dynamic compliance, and dynamic hyperinflation (Table 3). In 15 patients (62.5%), CAO was visible on the CT thorax. Patients with CAO experienced greater improvement after treatment in FEV1, FVC, RV, dynamic and static compliance, and 6MWD than patients without CAO (Table 4; online suppl. Table 3).

Table 3.

Association between change in RV or airway resistance and change in clinical outcomes (n = 24)

Association between change in RV or airway resistance and change in clinical outcomes (n = 24)
Association between change in RV or airway resistance and change in clinical outcomes (n = 24)
Table 4.

Significant differences between patients with or without visible CAO

Significant differences between patients with or without visible CAO
Significant differences between patients with or without visible CAO

Predictors of Response

A larger decrease in RV in response to the LVRC treatment was significantly associated with the following clinical variables at baseline: a higher RV (measured by body plethysmography and on CT scan), larger air trapping (measured on CT scan and the TLC difference between body plethysmography and helium dilution), higher treated lobes emphysema score, and lower physical activity level (Table 5; online suppl. Table 4).

Table 5.

Significant association between change in RV and potential predictors of response at baseline

Significant association between change in RV and potential predictors of response at baseline
Significant association between change in RV and potential predictors of response at baseline

This study aimed to gain more knowledge on the underlying physiological mechanism and potential predictors of response to the LVRC treatment. Consequently, patients underwent a large number of physiological measurements before and after treatment. Our results showed that patients can benefit from the treatment in well-known domains like static hyperinflation, exercise capacity, and quality of life but also in lesser known domains, notably air trapping and airway resistance. Furthermore, both a decrease in airway resistance and RV was significantly associated with an improvement in spirometry, air trapping, ventilation heterogeneity, dynamic compliance, dynamic hyperinflation, and diaphragm function. At baseline, more severe static hyperinflation, more air trapping, more severe emphysema destruction in the treatment target lobes, and a lower physical activity level were predictors of successful LVRC treatment.

In accordance with multiple previous studies, we found that the LVRC treatment had an acceptable safety profile and led to significant improvements in lung function, static hyperinflation, exercise capacity, and quality of life [3-5, 7, 25, 26]. We investigated a broad range of clinical outcomes, some of them for the first time, whereas others were previously tested in small case series. For example, we found for the first time significant improvements in air trapping, both measured with body plethysmography and using quantitative CT scan analysis, as well as maximal exercise capacity measured by incremental cycle ergometry. The significant change in airway resistance was in line with 2 smaller previous studies [27, 28]. In contrast, we did not find significant improvements in DLCO, dynamic hyperinflation, inspiratory and expiratory muscle strength, lung compliance, lung perfusion, airway radius measured on CT scan, and physical activity. Opposite to our results, Makris et al. [29] found a significant decrease in dynamic compliance in a small group of patients (n = 10) compared with controls (n = 11). The nonsignificant change in physical activity was in line with our previous pilot study [30] but not with a recent published pilot study [31]. This difference could be caused by the extreme low baseline physical activity levels in the other study, with only 2 days of measurement compared to 7 and the shorter follow-up after coil treatment (6 vs. 12 weeks). We also did not find a change in perfusion in the (un)treated lobes, while Lador et al. [32] did find an improvement in perfusion in the untreated lobes after treatment. This difference could be caused by the fact that they measured perfusion by dual-energy computed tomography.

Figure 2 shows a schematic overview of the potential mechanism of action based on our results. We found that both a decrease in airway resistance and RV was associated with a significant change in FEV1 and FVC, air trapping, ventilation heterogeneity, dynamic compliance, dynamic hyperinflation, and diaphragm function. Our results confirm previous studies that have shown that the LVRC treatment leads to reduction of the lung volume, which can lead to clinical benefits [6, 27]. It seems that the primary driver of action is the actual lung volume reduction, together with an accompanied decrease in airway resistance. Consequently, the decrease in RV can lead to an increase in vital capacity and the inspiratory capacity, which increased the ability to develop dynamic hyperinflation. This suggests that the coil treatment does not directly improve the elasticity and recoil of the lung but that the lung mechanics change due to the reduction in lung volume and airway resistance. It could be that due to the stiffness of the current coil design, the coil is not able to reduce the elasticity. Future trials using new to be designed coils with greater intrinsic flexibility should show whether these can lead to a direct improvement in elasticity and recoil. Our results also showed that patients in whom CAO was visible had additional benefits compared to patients who did not, which is in line with results from the RENEW study [4]. It seems that these reactive changes are associated with compression of the surrounding parenchyma and therefore contributes to volume reduction in the treated lobe.

Fig. 2.

Schematic overview of the potential mechanism of action of the coil treatment. FEV1, forced expiratory volume in 1 s; FVC, forced vital capacity; TLC, total lung capacity difference; VA, alveolar volume; IC, inspiratory capacity; MPT, metronome paced test; 6MWD, 6-min walk distance; TLC, total lung capacity; MEP, maximal expiratory pressure; CAO, coil-associated opacity.

Fig. 2.

Schematic overview of the potential mechanism of action of the coil treatment. FEV1, forced expiratory volume in 1 s; FVC, forced vital capacity; TLC, total lung capacity difference; VA, alveolar volume; IC, inspiratory capacity; MPT, metronome paced test; 6MWD, 6-min walk distance; TLC, total lung capacity; MEP, maximal expiratory pressure; CAO, coil-associated opacity.

Close modal

We found that greater static hyperinflation, more air trapping, higher emphysema destruction score, and a lower physical activity level at baseline were significantly associated with a larger improvement in RV after treatment. Higher static hyperinflation and emphysema destruction as predictors of response are in line with a recent publication [8]. In addition to this, we have also found larger air trapping (measured on CT scan and the TLC difference between body plethysmography and helium dilution) at baseline to be a possible predictor of response. Furthermore, the RENEW post hoc study did demonstrate the negative influence of the presence of airway disease on treatment outcome. However, in the present study, we did not enroll patients with severe airway disease, and therefore, this could not be evaluated in our analyses as a potential predictor. Combining both results show that it is important to only treat patients who have severe lung hyperinflation and emphysema. In that perspective, our inclusion criterion of a RV below 175% of predicted was probably too low, as reflected by our responder rates of 33–58%. It would be interesting to establish a cutoff value of RV as an indicator of response, but unfortunately, our sample size was too low for this kind of analysis. Significant air trapping could be added to these predictors of response, which can be measured on CT scan or the TLC difference between body plethysmography and helium dilution. The addition of significant air trapping is also in line with the recommendations of an expert panel [2]. A low physical activity level was also a predictor of better response in our population, which is less clear. More severe static hyperinflation could cause a lower level of physical activity. But it could also be an indication that it is important to treat only symptomatic patients with limitations, in line with the general recommendations for bronchoscopic lung volume reduction [33]. However, it is also strange that a low physical activity level was a predictor, while 6-minute walk distance and maximal workload on the cycle ergometer test were not.

Our study has some limitations. The sample size was relatively low, and we did not include a control group. However, 3-coil RCTs did not show any benefit in the control groups 3–12 months after randomization [3-5]. Furthermore, the follow-up was relatively short (3 months), which could have been too short to detect improvements in, for example, physical activity and inspiratory and expiratory muscle strength. A strength of our study is the extensive amount of physiological measurements that have been performed. Consequently, we were able to investigate and combine for the first time a broad range of potential mechanisms of action and predictors of response.

Unfortunately, the future of the coil treatment is uncertain as the current owner of the treatment (Boston Scientific Corp., Marlborough, MA, USA) has decided to discontinue with the treatment. Therefore, it is not known whether patients can still be treated with coils in the future. However, currently, a new generation of the device is developed with an improved design (patent number: US20190328400A1). We believe the results of our study could be applicable to this new-generation device as well, but future research will need to confirm this.

Our results confirm that patients benefit from the bronchoscopic LVRC treatment. Furthermore, it seems that the primary mechanism of action is decreasing static hyperinflation with improvement of airway resistance, which consequently changes dynamic lung mechanics. However, the right patient population needs to be selected for the treatment to be beneficial which should include patients with severe lung hyperinflation, severe air trapping, and severe emphysema in target lobes.

The study was approved by the Medical Ethics Committee of the University Medical Center Groningen (METc#2014.447), and all patients provided informed consent. The study was registered in clinicaltrials.gov with number NCT02179125.

J.E.H., K.K., N.H.T.t.H., S.W.S.A., W.v.G., and J.L.G. have no conflicts of interest to declare. P.S. reports personal fees from Boston Scientific, CSA Medical, Creo Medical, Nuvairia Olympus, Medtronic, and PneumRX/BTG as consultant on scientific advisory board, sponsorship to Imperial College for a bronchoscopy course by from ERBE, Cook medical, Medtronic, Boston Scientific, Broncus, Pulmonx, Olympus, and PneumRX/BTG, all outside the submitted work. He has been an investigator on clinical trials with bronchial thermoplasty, endobronchial valves, endobronchial coils, thermal ablation, and the airway bypass procedure. DJS reports grants, nonfinancial support, and other from Nuvaira Inc., Minneapolis, MN, USA; grants and nonfinancial support from CSA medical, Boston, MA, USA; grants, personal fees, and nonfinancial support from PulmonX Inc., Redwood City, CA, USA; and grants and nonfinancial support from PneumRx/BTG, California, USA, all outside the submitted work.

The authors received an unrestricted grant of PneumRx GmbH (Dusseldorf, Germany) which also provided the treatment devices in kind.

J.E.H. and D.J.S. contributed to the conception of the study, acquisition of data, and data analyses and wrote the first version of the manuscript. K.K., P.L.S., and NHTtH contributed to the conception of the study and acquisition of data and revised the manuscript. S.W.S.A., W.v.G., and J.L.G. contributed to acquisition of data and revised the manuscript. All authors approved the final version of the manuscript.

1.
Kemp
SV
,
Polkey
MI
,
Shah
PL
.
The epidemiology, etiology, clinical features, and natural history of emphysema
.
Thorac Surg Clin
.
2009
;
19
(
2
):
149
58
. .
2.
Slebos
D-J
,
ten Hacken
NH
,
Hetzel
M
,
Herth
FJF
,
Shah
PL
.
Endobronchial coils for endoscopic lung volume reduction: best practice recommendations from an expert panel
.
Respiration
.
2018
;
96
(
1
):
1
11
. .
3.
Shah
PL
,
Zoumot
Z
,
Singh
S
,
Bicknell
SR
,
Ross
ET
,
Quiring
J
, et al.
Endobronchial coils for the treatment of severe emphysema with hyperinflation (RESET): a randomised controlled trial
.
Lancet Respir Med
.
2013
;
1
(
3
):
233
40
. .
4.
Sciurba
FC
,
Criner
GJ
,
Strange
C
,
Shah
PL
,
Michaud
G
,
Connolly
TA
, et al.
Effect of endobronchial coils vs usual care on exercise tolerance in patients with severe emphysema: the renew randomized clinical trial
.
JAMA
.
2016
;
315
(
20
):
2178
89
. .
5.
Deslée
G
,
Mal
H
,
Dutau
H
,
Bourdin
A
,
Vergnon
JM
,
Pison
C
, et al.
Lung volume reduction coil treatment vs. usual care in patients with severe emphysema
.
JAMA
.
2016
;
315
(
2
):
175
84
. .
6.
Hartman
JE
,
Shah
PL
,
Sciurba
F
,
Herth
FJF
,
Slebos
DJ
;
RENEW Study Group
.
Endobronchial coils for emphysema: dual mechanism of action on lobar residual volume reduction
.
Respirology
.
2020
;
25
(
11
):
1160
6
.
7.
Slebos
DJ
,
Hartman
JE
,
Klooster
K
,
Blaas
S
,
Deslee
G
,
Gesierich
W
, et al.
Bronchoscopic coil treatment for patients with severe emphysema: a meta-analysis
.
Respiration
.
2015
;
90
(
2
):
136
45
. .
8.
Slebos
DJ
,
Cicenia
J
,
Sciurba
FC
,
Criner
GJ
,
Hartman
JE
,
Garner
J
, et al.
Predictors of response to endobronchial coil therapy in patients with advanced emphysema
.
Chest
.
2019
;
155
(
5
):
928
37
. .
9.
Wanger
J
,
Clausen
JL
,
Coates
A
,
Pedersen
OF
,
Brusasco
V
,
Burgos
F
, et al.
Standardisation of the measurement of lung volumes
.
Eur Respir J
.
2005 Sep
;
26
(
3
):
511
22
. .
10.
Miller
MR
,
Hankinson
J
,
Brusasco
V
,
Burgos
F
,
Casaburi
R
,
Coates
A
, et al.
Standardisation of spirometry
.
Eur Respir J
.
2005 Aug
;
26
(
2
):
319
38
. .
11.
MacIntyre
N
,
Crapo
RO
,
Viegi
G
,
Johnson
DC
,
van der Grinten
CP
,
Brusasco
V
, et al.
Standardisation of the single-breath determination of carbon monoxide uptake in the lung
.
Eur Respir J
.
2005
;
26
(
4
):
720
35
. .
12.
Sclauser Pessoa
IM
,
Franco Parreira
V
,
Fregonezi
GA
,
Sheel
AW
,
Chung
F
,
Reid
WD
.
Reference values for maximal inspiratory pressure: a systematic review
.
Can Respir J
.
2014
;
21
(
1
):
43
50
. .
13.
Laveneziana
P
,
Albuquerque
A
,
Aliverti
A
,
Babb
T
,
Barreiro
E
,
Dres
M
, et al.
ERS statement on respiratory muscle testing at rest and during exercise
.
Eur Respir J
.
2019
;
53
(
6
):
1801214
. .
14.
Klooster
K
,
Ten Hacken
NH
,
Hartman
JE
,
Sciurba
FC
,
Kerstjens
HA
,
Slebos
DJ
.
Determining the role of dynamic hyperinflation in patients with severe chronic obstructive pulmonary disease
.
Respiration
.
2015
;
90
(
4
):
306
13
. .
15.
ATS Committee on Proficiency Standards for Clinical Pulmonary Function Laboratories
.
ATS statement: guidelines for the six-minute walk test
.
Am J Respir Crit Care Med
.
2002 Jan
;
166
(
1
):
111
7
. .
16.
American Thoracic Society; American College of Chest Physicians
.
ATS/ACCP statement on cardiopulmonary exercise testing
.
Am J Respir Crit Care Med
.
2003
;
167
(
2
):
211
77
. .
17.
Rabinovich
RA
,
Louvaris
Z
,
Raste
Y
,
Langer
D
,
Van Remoortel
H
,
Giavedoni
S
, et al.
Validity of physical activity monitors during daily life in patients with COPD
.
Eur Respir J
.
2013
;
42
(
5
):
1205
15
. .
18.
Jones
PW
,
Quirk
FH
,
Baveystock
CM
,
Littlejohns
P
.
A self-complete measure of health status for chronic airflow limitation. The St. George’s respiratory questionnaire
.
Am Rev Respir Dis
.
1992
;
145
(
6
):
1321
7
.
19.
Donohue
JF
.
Minimal clinically important differences in COPD lung function
.
COPD
.
2005
;
2
(
1
):
111
24
. .
20.
Hartman
JE
,
Ten Hacken
NH
,
Klooster
K
,
Boezen
HM
,
De Greef
MH
,
Slebos
DJ
.
The minimal important difference for residual volume in patients with severe emphysema
.
Eur Respir J
.
2012
;
40
(
5
):
1137
41
. .
21.
Puhan
MA
,
Chandra
D
,
Mosenifar
Z
,
Ries
A
,
Make
B
,
Hansel
NN
, et al.
The minimal important difference of exercise tests in severe COPD
.
Eur Respir J
.
2011
;
37
(
4
):
784
90
. .
22.
Welling
JBA
,
Hartman
JE
,
Ten Hacken
NHT
,
Klooster
K
,
Slebos
D-J
.
The minimal important difference for the St George’s respiratory questionnaire in patients with severe COPD
.
Eur Respir J
.
2015
;
46
(
6
):
1598
604
.
23.
Demeyer
H
,
Burtin
C
,
Hornikx
M
,
Camillo
CA
,
Van Remoortel
H
,
Langer
D
, et al.
The minimal important difference in physical activity in patients with COPD
.
PLoS One
.
2016
;
11
(
4
):
e0154587
. .
24.
Teylan
M
,
Kantorowski
A
,
Homsy
D
,
Kadri
R
,
Richardson
C
,
Moy
M
.
Physical activity in COPD: minimal clinically important difference for medical events
.
Chron Respir Dis
.
2019 Jan–Dec
;
16
:
1479973118816424
9
. .
25.
Zoumot
Z
,
Kemp
SV
,
Singh
S
,
Bicknell
SR
,
McNulty
WH
,
Hopkinson
NS
, et al.
Endobronchial coils for severe emphysema are effective up to 12 months following treatment: medium term and cross-over results from a randomised controlled trial
.
PLoS One
.
2015
;
10
(
4
):
e0122656
. .
26.
Slebos
DJ
,
Klooster
K
,
Ernst
A
,
Herth
FJF
,
Kerstjens
HAM
.
Bronchoscopic lung volume reduction coil treatment of patients with severe heterogeneous emphysema
.
Chest
.
2012
;
142
(
3
):
574
82
. .
27.
Palamidas
AF
,
Kemp
SV
,
Shen
M
,
McNulty
W
,
Zoumot
Z
,
Hopkinson
NS
, et al.
Putative mechanisms of action of endobronchial coils
.
Am J Respir Crit Care Med
.
2017 Jul
;
196
(
1
):
109
15
. .
28.
Klooster
K
,
Ten Hacken
NH
,
Franz
I
,
Kerstjens
HA
,
van Rikxoort
EM
,
Slebos
DJ
.
Lung volume reduction coil treatment in chronic obstructive pulmonary disease patients with homogeneous emphysema: a prospective feasibility trial
.
Respiration
.
2014
;
88
(
2
):
116
25
. .
29.
Makris
D
,
Leroy
S
,
Pradelli
J
,
Benzaquen
J
,
Guenard
H
,
Perotin
JM
, et al.
Changes in dynamic lung mechanics after lung volume reduction coil treatment of severe emphysema
.
Thorax
.
2018 Jun
;
73
(
6
):
584
6
. .
30.
Hartman
JE
,
Boezen
HM
,
Heintzbergen
S
,
De Greef
MH
,
Klooster
K
,
Ten Hacken
NH
, et al.
Daily physical activity after bronchoscopic lung volume reduction: a pilot study
.
Eur Respir J
.
2012
;
40
(
6
):
1566
7
. .
31.
Biener
L
,
Skowasch
D
,
Hollmann
S
,
Schreiber
T
,
Nickenig
G
,
Fimmers
RPC
.
Endoscopic lung volume reduction in COPD: the impact of coil implantation on patients’ physical activity
.
Respiration
.
2020
;
99
(
2
):
177
80
.
32.
Lador
F
,
Hachulla
AL
,
Hohn
O
,
Plojoux
J
,
Ronot
M
,
Montet
X
, et al.
Pulmonary perfusion changes as assessed by contrast-enhanced dual-energy computed tomography after endoscopic lung volume reduction by coils
.
Respiration
.
2016
;
92
(
6
):
404
13
. .
33.
Herth
FJF
,
Slebos
DJ
,
Criner
GJ
,
Valipour
A
,
Sciurba
F
,
Shah
PL
.
Endoscopic lung volume reduction: an expert panel recommendation: update 2019
.
Respiration
.
2019
;
97
(
6
):
548
57
. .
Open Access License / Drug Dosage / Disclaimer
This article is licensed under the Creative Commons Attribution-NonCommercial 4.0 International License (CC BY-NC). Usage and distribution for commercial purposes requires written permission. Drug Dosage: The authors and the publisher have exerted every effort to ensure that drug selection and dosage set forth in this text are in accord with current recommendations and practice at the time of publication. However, in view of ongoing research, changes in government regulations, and the constant flow of information relating to drug therapy and drug reactions, the reader is urged to check the package insert for each drug for any changes in indications and dosage and for added warnings and precautions. This is particularly important when the recommended agent is a new and/or infrequently employed drug. Disclaimer: The statements, opinions and data contained in this publication are solely those of the individual authors and contributors and not of the publishers and the editor(s). The appearance of advertisements or/and product references in the publication is not a warranty, endorsement, or approval of the products or services advertised or of their effectiveness, quality or safety. The publisher and the editor(s) disclaim responsibility for any injury to persons or property resulting from any ideas, methods, instructions or products referred to in the content or advertisements.