Background: Bubble CPAP may be used in infants with suspected or confirmed COVID-19. Electrostatic filters may reduce cross infection. This study aims to determine if including a filter in the bubble CPAP circuit impacts stability of pressure delivery. Methods: A new electrostatic filter was placed before (pre) or after (post) the bubble CPAP generator, or with no filter (control) in an in vitro study. Pressure was recorded at the nasal interface for 18 h (6 L/min; 7 cm H2O) on 3 occasions for each configuration. Filter failure was defined as pressure >9 cm H2O for 60 continuous minutes. The filter was weighed before and after each experiment. Results: Mean (SD) time to reach the fail point was 257 (116) min and 525 (566) min for filter placement pre- and post-CPAP generator, respectively. Mean pressure was higher throughout in the pre-generator position compared to control. The filter weight was heavier at end study in the pre- compared to the post-generator position. Conclusions: Placement of the filter at the pre-generator position in a bubble CPAP circuit should be avoided due to unstable mean pressure. Filters are likely to become saturated with water over time. The post-generator position may accommodate a filter, but regular pressure monitoring and early replacement are required.

SARS-CoV-2 originated in Wuhan, China, in November 2020 [1]. The clinical disease, COVID-19, was declared a global pandemic by WHO in March 2020 [2]. Data on neonatal infection with SARS-CoV-2 are limited [3-5]. The need for respiratory support in infants with COVID-19 is uncertain.

Continuous positive airway pressure (CPAP) has been a cornerstone treatment of neonatal respiratory distress syndrome for >40 years. CPAP use has extended in recent years to include infants with viral respiratory disease [6]. Bubble CPAP is a popular means of providing respiratory support in these infants. The expiratory limb of the bubble CPAP circuit vents through an underwater seal [7], creating bubbles. The resulting pressure oscillations are transmitted back to the nares and to the lung; hence, the bubbling delivers a variable rather than constant pressure [8]. These pressure oscillations (frequency) may improve acute respiratory mechanics compared to constant delivery pressure consequent to recruitment of atelectatic alveoli [7-11].

CPAP generates aerosols and therefore may promote transmission of viral particles [12, 13]. Electrostatic filters remove viruses from expired gas, offering protection to health care staff and are recommended in international guidelines [14]. Manufacturers of electrostatic filters recommend electrostatic filter replacement every 24 h [15]. One reason for these recommendations is the possibility of moisture saturation of the electrostatic filter which would increase resistance to airflow, increasing circuit pressure and work of breathing. We hypothesized that progressive moisture saturation of an electrostatic filter placed in humidified bias flow would result in increased circuit pressure. We aimed to determine how the placement and position of an electrostatic filter in a bubble CPAP circuit impacts the magnitude and stability of the delivered pressure in an in vitro model.

An in vitro study was conducted at the University of Western Australia in April 2020. A simulated in vitro neonatal lung on a bubble CPAP model was created by connecting a test lung (Dräger, Lubeck, Germany) to a bubble CPAP generator (Fisher & Paykel, Auckland, New Zealand). An infant nasal CPAP cannula (Size 4, Hudson RCI, Teleflex, Morrisville, NC, USA) was interfaced between the bias flow circuit and the test lung. Humidified bias flow (MR290; Fisher & Paykel, Auckland, New Zealand) was set at 6 L/min (of air). CPAP was measured at the proximal pressure line connector of the CPAP cannula using a physiological pressure transducer and dome (MLT844; ADInstruments, Sydney, Australia). The bubble CPAP generator and humidifier were set below the test lung, which was maintained on a radiant-heated neonatal cot (Cosy-Cot; Fisher and Paykel, Auckland, New Zealand). The CPAP generator was positioned below the circuit to promote free water drainage into the generator throughout the experiment without regular intervention. The cot warmer was set at 75% of maximum heat setting on the manual mode. Environmental ambient temperature was 23.9–24.7°C throughout the recording periods.

Filter Placement

A DARTM pediatric-neonatal electrostatic filter (size: small; Covidien, Mansfield, MA, USA) was placed at one of 2 points in the breathing circuit: on the CPAP probe immediately before the exit of the bias flow into the water column (pre-generator; Fig. 1b) or at the bias flow exit point, normally used as the water inlet (post-generator; Fig. 1c). A new filter was used for each experimental repeat and was weighed before and after each trial. A bubble CPAP circuit without a filter was used as the control (Fig. 1a) for the 2 experimental arms. Each filter position and control was repeated 3 times.

Fig. 1.

Filter placement. Control, standard setup (a); pre-generator expiratory limb (b); and post-generator at point of flow exit into environment (c).

Fig. 1.

Filter placement. Control, standard setup (a); pre-generator expiratory limb (b); and post-generator at point of flow exit into environment (c).

Close modal

Signal Recording

The pressure signal was amplified, digitized (ML880; PowerLab ADInstruments, Bella Vista, NSW, Australia), and recorded continuously for 18 h (LabChart 8, ADInstruments, Bella Vista, NSW, Australia) onto a laptop computer. This duration was chosen to reflect likely clinical exposure while remaining within the manufacturer’s maximum recommended period for use (24 h). Pressure and frequency data were collected at 5-min intervals in 10-s epochs at a sampling frequency of 1 kHz.

Analysis

The filter position fail point was defined as a mean pressure exceeding 9 cm H2O for >60 continuous minutes. Mean pressure variability in a no-leak bubble CPAP system is 25% greater than the set pressure [16]. Our fail point allowed a 25% increase on top of the set pressure to the nearest integer. The mean pressure and pressure difference (∆P) of the bubble oscillations for each group (pre-generator, post-generator, and control) were compared across the 18 h using a two-way repeated measures ANOVA with Holm-Sidak post hoc correction. The frequency oscillations, unique to bubble CPAP, were assessed by transforming the mean pressure time series recording into the frequency domain using power spectral analysis. The mean frequency at the maximum power signal of the 3 circuit configurations was compared using a two-way repeated measures ANOVA with Holm-Sidak post hoc correction. Filter weight and water volume in the circuit and overflow were compared by an unpaired Student’s t test and one-way ANOVA, respectively. p values were assessed at a 0.05 significance level by SigmaPlot (Systat, Inc., Santa Clara, CA, USA).

CPAP Fail Point

Failure to maintain mean pressure below 9 cm H2O occurred for 3 tests of the pre-generator and 2 of the 3 tests of the post-generator position. The mean (SD) time to the fail point in pre-generator position was 257 (116) min and 525 (566) min for post-generator position. Mean pressure did not reach the fail point in the control group.

Mean Pressure, ∆P, and Frequency

The mean pressure with the filter in the pre-generator position increased progressively throughout the experiment and was significantly greater than control in 15 of the 18 h of the study. ∆P was lower with the filter in the pre-generator position from 3 h after study commencement. In contrast, the increase in mean pressure with the filter at the post-generator position was not significant compared to control. Post-generator ∆P was significantly lower than control only at hours 8, 12, and 17. The mean frequency at maximum power was similar for all circuit configurations (Fig. 2a–f).

Fig. 2.

Mean pressure (a–b), ΔP (c–d), and frequency (e–f) over time for the filter positions and control: no filter (none, black square), pre-generator (pre, open triangle), post-generator (post, open circle). Data points are mean (SD) for n = 3/group. The fail points for the respective filter position are represented by a vertical dashed line. *p < 0.05.

Fig. 2.

Mean pressure (a–b), ΔP (c–d), and frequency (e–f) over time for the filter positions and control: no filter (none, black square), pre-generator (pre, open triangle), post-generator (post, open circle). Data points are mean (SD) for n = 3/group. The fail points for the respective filter position are represented by a vertical dashed line. *p < 0.05.

Close modal

Filter Weight

Filter weights before and after use are reported in Table 1. for both circuit filter positions. The filter weight was significantly heavier in the pre-generator compared to the post-generator position at the end of the study. Significantly more water accumulated in the generator overflow in the post-generator positions compared to the pre-generator position and the control position. Inclusion of a filter did not alter the volume of water that accumulated in the CPAP circuit.

Table 1.

Filter weights and volumes of water removed from the circuit

Filter weights and volumes of water removed from the circuit
Filter weights and volumes of water removed from the circuit

The placement of an electrostatic filter in a bubble CPAP circuit changes the stability of pressure delivery. Positioning of the filter either immediately before or after the bubble generator increases the mean pressure delivered above a predefined fail point. Filter failure occurs earlier and is sustained when the filter is positioned immediately before the CPAP generator, compared to later and intermittent failure when placed in the post-generator position (point of exit of flow into the environment). Incorporation of a filter did not change the mean frequency of the maximum power in the oscillatory pressure waveform. The filter become more saturated with water when positioned in the pre-generator versus post-generator position.

Placement of a filter in a bubble CPAP circuit reduces water accumulation within the circuit; however, the filter increases in weight over time, indicative of water absorption by the filter. Filter saturation increases resistance [17]. This may explain the increased mean pressure and dampened pressure waveform with the filter in the pre-generator position. Microbial filters marginally increase the imposed work of breathing measured over 18 breaths across a range of non-humidified respiratory support devices [18]. In comparison, our recordings were in a humidified system and were conducted over an extended period (18 h); using this more clinically applicable setting we clearly show that time and humidification affect mean pressure when the circuit configuration includes an electrostatic filter, especially when the filter is positioned prior to the bubble generator.

Few studies report the volume of airway condensate in infants on CPAP. The estimated median exhaled rate of breath condensate production in moderately preterm infants receiving nasal CPAP is 0.07 mL/min [19]. If extrapolated over 18 h, the exhaled breath condensate would equate to 75 mL of additional liquid into the breathing circuit. This estimated condensate volume is substantially more than that measured in the experiments reported here. The lower volume of breath condensate in our study may be due to the in vitro nature of the study, which did not include a contribution of expired gas from the lung. The added contribution of humidified exhaled air may saturate the electrostatic filter sooner when used in the clinical setting. Thus, the clinician needs to be aware that filter changes may be required even sooner than suggested by our study to avoid potentially adverse increases in circuit pressure due to increased expiratory resistance through the saturated filter.

A highly variable or noisy pressure waveform is characteristic of bubble CPAP. Similar to high-frequency oscillatory ventilation, the amplitude of the bubble CPAP pressure waveform (∆P) is a fundamental determinant of carbon dioxide clearance [7, 20]. The variable pressure is also fundamental for stochastic recruitment of collapsed alveoli [7-11]. Bubble CPAP improved measures of acute respiratory mechanics and reduced lung injury compared to constantly delivered CPAP in preterm lambs [7]. Damping of the ∆P with the filter in the pre-generator position would reduce the efficacy of bubble CPAP for gas exchange and alveolar recruitment.

Although the in vitro nature of this study may be considered a limitation, such models are used widely for describing the performance of respiratory support devices, including neonatal respiratory support modalities [16, 18, 21-24]. Our in vitro experimental setup was a sealed circuit, measuring the stability of pressure without considering leak. Clinically, bubble CPAP pressure delivery is determined by the bias flow and degree of leak at the nares, as well as the degree of mouth closure [16]. Many neonatal units aim to minimize leak by using chin straps (to keep the mouth closed) and colloid dressings over the nares that not only reduce nasal trauma but also markedly reduce leak at the nasal interface. The instability of the delivered pressure over time should concern clinicians considering the use of such filters when treating premature infants with bubble CPAP, particularly when positioning the filter immediately before the bubble CPAP generator chamber.

The transmission of SARS-CoV-2 to health care workers is a major concern in the current pandemic [25]. Neonates are at risk of COVID-19, although evidence of infection is sparse. Newborns of mothers with confirmed or suspected COVID-19 require precautions as recommended in recent published international guidelines [26]. Filters reduced the expired viral load from ventilator circuits when tested in previous pandemics [27, 28] and are recommended by several international societies [29, 30]. However, it is essential to consider how such filters impact the efficacy of the respiratory support and to identify required changes in clinical practice, such as frequency of filter exchange, to ensure safety of both health care workers and the patient.

We conclude that the addition of an electrostatic filter into the circuit of a bubble CPAP system must be undertaken with extreme caution. The clinical impact of such filters may be the unstable and unintended delivery of excessive airway pressure resulting in unwanted side effects such as air leak. Placement of a viral filter in the expiratory circuit immediately before the bubble CPAP generator chamber is not recommended. The excessive drift in mean pressure delivery and damping of the pressure waveform with the filter in this position presents a safety risk for the infant that outweighs the benefit of preventing cross infection. The addition of the filter at the flow exit point for the CPAP generator (post-generator) may be acceptable but requires monitoring of the pressure in the circuit at regular intervals. The manufacturer of this filter recommends changing it every 24 h. Based on the data presented, changing the filter at least every 12 h, and likely every 8 h, should be considered when used in bubble CPAP.

This paper is exempt from ethical committee approval as it was conducted in a laboratory without the participation of human or animal subjects.

J.W. Davis, M.N. Cooper, and M.J. Dahl have no financial ties to the products in this study. J.J. Pillow has received unrestricted grants of equipment (humidifiers) and consumables (ventilator and CPAP circuits) for use in preclinical research from Fisher and Paykel Healthcare.

The authors received no specific funding for this work.

M.J. Dahl and J.W. Davis made substantial contributions to the study conception and design, acquisition, analysis, interpretation of data, drafting and revising the manuscript. M.N. Cooper made a substantial contribution to the analysis, interpretation of the data, and drafting and revising the article for submission. J.J. Pillow contributed to the study design, analysis, and interpretation of data and made a substantial contribution in revising the manuscript and approving for submission.

1.
Mimouni
F
,
Lakshminrusimha
S
,
Pearlman
SA
,
Raju
T
,
Gallagher
PG
,
Mendlovic
J
.
Perinatal aspects on the covid-19 pandemic: a practical resource for perinatal-neonatal specialists
.
J Perinatol
.
2020
;
40
(
5
):
820
6
. .
2.
Jin
Y
,
Yang
H
,
Ji
W
,
Wu
W
,
Chen
S
,
Zhang
W
, et al
Virology, epidemiology, pathogenesis, and control of COVID-19
.
Viruses
.
2020
;
12
(
4
):
372
.
3.
Piersigilli
F
,
Carkeek
K
,
Hocq
C
,
van Grambezen
B
,
Hubinont
C
,
Chatzis
O
, et al
COVID-19 in a 26-week preterm neonate
.
Lancet Child Adolesc Health
.
2020
;
4
(
6
):
476
8
. .
4.
Zeng
L
,
Xia
S
,
Yuan
W
,
Yan
K
,
Xiao
F
,
Shao
J
, et al
Neonatal early-onset infection with SARS-CoV-2 in 33 neonates born to mothers with COVID-19 in Wuhan, China
.
JAMA Pediatr
.
2020
;
174
(
7
):
722
5
. .
5.
Zimmermann
P
,
Curtis
N
.
COVID-19 in children, pregnancy and neonates: a review of epidemiologic and clinical features
.
Pediatr Infect Dis J
.
2020
;
39
(
6
):
469
77
. .
6.
Franklin
D
,
Fraser
JF
,
Schibler
A
.
Respiratory support for infants with bronchiolitis, a narrative review of the literature
.
Paediatr Respir Rev
.
2019
;
30
:
16
24
. .
7.
Pillow
JJ
,
Hillman
N
,
Moss
TJ
,
Polglase
G
,
Bold
G
,
Beaumont
C
, et al
Bubble continuous positive airway pressure enhances lung volume and gas exchange in preterm lambs
.
Am J Respir Crit Care Med
.
2007
;
176
(
1
):
63
9
. .
8.
Pillow
JJ
,
Travadi
JN
.
Bubble CPAP: is the noise important? An in vitro study
.
Pediatr Res
.
2005
;
57
(
6
):
826
30
. .
9.
Gupta
S
,
Sinha
SK
,
Tin
W
,
Donn
SM
.
A randomized controlled trial of post-extubation bubble continuous positive airway pressure versus infant flow driver continuous positive airway pressure in preterm infants with respiratory distress syndrome
.
J Pediatr
.
2009
;
154
(
5
):
645
50
. .
10.
Narendran
V
,
Donovan
EF
,
Hoath
SB
,
Akinbi
HT
,
Steichen
JJ
,
Jobe
AH
.
Early bubble CPAP and outcomes in ELBW preterm infants
.
J Perinatol
.
2003
;
23
(
3
):
195
9
. .
11.
Nowadzky
T
,
Pantoja
A
,
Britton
JR
.
Bubble continuous positive airway pressure, a potentially better practice, reduces the use of mechanical ventilation among very low birth weight infants with respiratory distress syndrome
.
Pediatrics
.
2009
;
123
(
6
):
1534
40
. .
12.
Wax
RS
,
Christian
MD
.
Practical recommendations for critical care and anesthesiology teams caring for novel coronavirus (2019-nCoV) patients
.
Can J Anesth
.
2020
;
67
(
5
):
568
76
. .
13.
Wax
RS
,
Christian
MD
.
Practical recommendations for critical care and anesthesiology teams caring for novel coronavirus (2019-nCoV) patients
.
Can J Anaesth
.
2020
;
67
(
5
):
568
76
. .
14.
Nolan
JP
,
Monsieurs
KG
,
Bossaert
L
,
Böttiger
BW
,
Greif
R
,
Lott
C
, et al
European resuscitation council COVID-19 guidelines executive summary
.
Resuscitation
.
2020
;
153
:
45
55
. .
16.
Kahn
DJ
,
Courtney
SE
,
Steele
AM
,
Habib
RH
.
Unpredictability of delivered bubble nasal continuous positive airway pressure: role of bias flow magnitude and nares-prong air leaks
.
Pediatr Res
.
2007
;
62
(
3
):
343
7
. .
17.
Vanderbroucke-Grauls
CM
,
Teeuw
KB
,
Ballemans
K
,
Lavooij
C
,
Cornelisse
PB
,
Verhoef
J
.
Bacterial and viral removal efficiency, heat and moisture exchange properties of four filtration devices
.
J Hosp Infect
.
1995
;
29
(
1
):
45
56
. .
18.
Donaldsson
S
,
Naver
L
,
Jonsson
B
,
Drevhammar
T
.
COVID-19: minimising contaminated aerosol spreading during CPAP treatment
.
Arch Dis Child Fetal Neonatal Ed
.
2020
;
105
(
6
):
669
71
.
19.
Cheah
FC
,
Darlow
BA
,
Winterbourn
CC
.
Problems associated with collecting breath condensate for the measurement of exhaled hydrogen peroxide from neonates on respiratory support
.
Biol Neonate
.
2003
;
84
(
4
):
338
41
. .
20.
Lee
KS
,
Dunn
MS
,
Fenwick
M
,
Shennan
AT
.
A comparison of underwater bubble continuous positive airway pressure with ventilator-derived continuous positive airway pressure in premature neonates ready for extubation
.
Biol Neonate
.
1998
;
73
(
2
):
69
75
. .
21.
Pillow
JJ
,
Neil
H
,
Wilkinson
MH
,
Ramsden
CA
.
Effect of I/E ratio on mean alveolar pressure during high-frequency oscillatory ventilation
.
J Appl Physiol
.
1999
;
87
(
1
):
407
14
.
22.
Pillow
JJ
,
Wilkinson
MH
,
Neil
HL
,
Ramsden
CA
.
In vitro performance characteristics of high-frequency oscillatory ventilators
.
Am J Respir Crit Care Med
.
2001
;
164
(
6
):
1019
24
. .
23.
Fredberg
JJ
,
Glass
GM
,
Boynton
BR
,
Frantz
ID
 3rd
.
Factors influencing mechanical performance of neonatal high-frequency ventilators
.
J Appl Physiol
.
1987
;
62
(
6
):
2485
90
.
24.
Youngquist
TM
,
Richardson
CP
,
Diblasi
RM
.
Effects of condensate in the exhalation limb of neonatal circuits on airway pressure during bubble CPAP
.
Respir Care
.
2013
;
58
(
11
):
1840
6
. .
25.
Barranco
R
,
Ventura
F
.
Covid-19 and infection in health-care workers: an emerging problem
.
Med Leg J
.
2020
;
88
(
2
):
65
6
. .
26.
Lavizzari
A
,
Klingenberg
C
,
Profit
J
,
Zupancic
JAF
,
Davis
AS
,
Mosca
F
, et al
International comparison of guidelines for managing neonates at the early phase of the SARS-CoV-2 pandemic
.
Pediatr Res
.
2020
. .
27.
Heuer
JF
,
Crozier
TA
,
Howard
G
,
Quintel
M
.
Can breathing circuit filters help prevent the spread of influenza A (H1N1) virus from intubated patients?
GMS Hyg Infect Control
.
2013
;
8
(
1
):
Doc09
. .
28.
Mechanical ventilation of SARS patients. Safety issues involving breathing-circuit filters
.
Health Devices
.
2003
;
32
(
6
):
220
2
.
29.
Respiratory Care Committee of Chinese Thoracic Society
.
[Expert consensus on preventing nosocomial transmission during respiratory care for critically ill patients infected by 2019 novel coronavirus pneumonia]
.
Zhonghua Jie He He Hu Xi Za Zhi
.
2020
;
43
(
4
):
288
96
.
30.
Sprung
CL
,
Zimmerman
JL
,
Christian
MD
,
Joynt
GM
,
Hick
JL
,
Taylor
B
, et al
Recommendations for intensive care unit and hospital preparations for an influenza epidemic or mass disaster: summary report of the European Society of Intensive Care Medicine’s Task Force for intensive care unit triage during an influenza epidemic or mass disaster
.
Intensive Care Med
.
2010
;
36
(
3
):
428
43
.
Copyright / Drug Dosage / Disclaimer
Copyright: All rights reserved. No part of this publication may be translated into other languages, reproduced or utilized in any form or by any means, electronic or mechanical, including photocopying, recording, microcopying, or by any information storage and retrieval system, without permission in writing from the publisher.
Drug Dosage: The authors and the publisher have exerted every effort to ensure that drug selection and dosage set forth in this text are in accord with current recommendations and practice at the time of publication. However, in view of ongoing research, changes in government regulations, and the constant flow of information relating to drug therapy and drug reactions, the reader is urged to check the package insert for each drug for any changes in indications and dosage and for added warnings and precautions. This is particularly important when the recommended agent is a new and/or infrequently employed drug.
Disclaimer: The statements, opinions and data contained in this publication are solely those of the individual authors and contributors and not of the publishers and the editor(s). The appearance of advertisements or/and product references in the publication is not a warranty, endorsement, or approval of the products or services advertised or of their effectiveness, quality or safety. The publisher and the editor(s) disclaim responsibility for any injury to persons or property resulting from any ideas, methods, instructions or products referred to in the content or advertisements.