Regular physical activity and exercise (PA) are cornerstones of diabetes care for individuals with type 1 diabetes. In recent years, the availability of automated insulin delivery (AID) systems has improved the ability of people with type 1 diabetes to achieve the recommended glucose target ranges. PA provides additional health benefits but can cause glucose fluctuations, which challenges current AID systems. While an increasing number of clinical trials and reviews are being published on different AID systems and PA, it seems prudent at this time to collate this information and develop a position statement on the topic. This joint European Association for the Study of Diabetes (EASD)/International Society for Pediatric and Adolescent Diabetes (ISPAD) position statement reviews current evidence on AID systems and provides detailed clinical practice points for managing PA in children, adolescents and adults with type 1 diabetes using AID technology. It discusses each commercially available AID system individually and provides guidance on its use in PA. Additionally, it addresses different glucose responses to PA and provides stratified therapy options to maintain glucose levels within the target ranges for these age groups.

AID

Automated insulin delivery

CGM

Continuous glucose monitoring

CHO

Carbohydrates

IOB

Insulin on board

ISPAD

International Society for Pediatric and Adolescent Diabetes

PA

Physical activity and exercise

TAR>10.0

Time above range (>10.0 mmol/L)

TBR<3.9

Time below range (<3.9 mmol/L)

TBR<3.0

Time below range (<3.0 mmol/L)

TIR3.9–10.0

Time in range (3.9–10.0 mmol/L)

Regular moderate-to-vigorous physical activity and exercise (PA) can be beneficial for managing type 1 diabetes [1, 2]. While previous position statements have provided guidance on glucose management during exercise based on glycaemic trends and continuous glucose monitoring (CGM) (Fig. 1 [1‒4]), recommendations on using current commercially available automated insulin delivery (AID) systems for PA are limited [5‒8]. The general glucose responses to PA and considerations for insulin dose changes and carbohydrate (CHO) intake, as shown in Figure 1, lay the foundation for the general principles of AID use described below.

Fig. 1.

A general overview of glucose trends and exogenous insulin and CHO intake requirements in response to PA in people with type 1 diabetes and not a complete reflection of the variability that may exist within and between each individual and in different PA types. The average glucose responses to exercise (top row) are highly variable based on several factors including IOB, baseline glucose, glucose rate of change, time of day, fitness level, prandial state, fasted state, and menstrual cycle phase [21, 67, 69]. People with type 1 diabetes should understand their individual responses to different types of activity [69] and in different settings (e.g., morning vs. afternoon [22], practice vs. competition [70]). Strategies can then be individualised based on their average glucose responses. No one PA can be associated with one glucose trend; however, activities shown in the upper left panel tend to result in the glucose trends in the first two columns; activities shown in the upper right panel tend to result in the glucose trends in the last two columns; and activities shown in the middle panel can result in the glucose trends in the middle three columns. When considering an increase in insulin doses around PA, this should be discussed with the healthcare professional and care team as only a few studies have investigated higher insulin doses for exercise. This figure is available as part of a downloadable slideset.

Fig. 1.

A general overview of glucose trends and exogenous insulin and CHO intake requirements in response to PA in people with type 1 diabetes and not a complete reflection of the variability that may exist within and between each individual and in different PA types. The average glucose responses to exercise (top row) are highly variable based on several factors including IOB, baseline glucose, glucose rate of change, time of day, fitness level, prandial state, fasted state, and menstrual cycle phase [21, 67, 69]. People with type 1 diabetes should understand their individual responses to different types of activity [69] and in different settings (e.g., morning vs. afternoon [22], practice vs. competition [70]). Strategies can then be individualised based on their average glucose responses. No one PA can be associated with one glucose trend; however, activities shown in the upper left panel tend to result in the glucose trends in the first two columns; activities shown in the upper right panel tend to result in the glucose trends in the last two columns; and activities shown in the middle panel can result in the glucose trends in the middle three columns. When considering an increase in insulin doses around PA, this should be discussed with the healthcare professional and care team as only a few studies have investigated higher insulin doses for exercise. This figure is available as part of a downloadable slideset.

Close modal

In general, people with type 1 diabetes with lower incomes often face numerous challenges that limit their opportunities to adopt technology, including access to insulin pump therapy and CGM systems, not to mention AID systems [9]. Although the prevalence of type 1 diabetes is increasing globally, it is estimated that only 800,000 individuals are currently using AID systems [10], with significant regional differences in access and insurance support. People with type 1 diabetes using AID technology often face significant challenges around meals [11, 12] and PA, both planned and unplanned [8]. Furthermore, several barriers to PA exist (e.g., fear of hypoglycaemia) that may increase the risk of diabetes distress [13]. At present, AID users are required to manually announce meals and adjust for anticipated PA. Some of the technical limitations include a CGM “lag time” between blood and interstitial glucose concentrations with rapid changes in glucose levels [14‒19]. In this joint European Association for the Study of Diabetes (EASD)/International Society for Pediatric and Adolescent Diabetes (ISPAD) position statement, we discuss the existing evidence on commercially available AID systems around PA and provide recommendations for managing a physically active lifestyle in children, adolescents, and adults with type 1 diabetes. For additional information on emerging AID technology for use in type 1 diabetes, see electronic online supplementary ESM 1 (for all online suppl. material, see https://doi.org/10.1159/000542287). This position statement is intended for both healthcare professionals and individuals with type 1 diabetes and aims to provide strategies for effective glucose management around planned and unplanned PA. Self-management of PA is often challenging for individuals with type 1 diabetes, no matter what type of insulin therapy they are using. This document provides a comprehensive overview of PA and current AID systems that will serve as a starting point to better manage PA safely and effectively.

The writing group members were selected by O.M. and M.C.R. (October 4, 2023) based on publication record and/or clinical experience in the field of AID and PA and approved by the EASD (May 7, 2024; see online suppl. ESM 2). Following initial discussions with specific members of the writing group (O.M., D.P.Z., S.E.H., J.K.M., C.M., H.S., M.C.R.), a first draft was produced by the co-first authors (O.M. and D.P.Z.) and circulated to the writing group for further discussions and feedback (April 5, 2024). Consensus meetings were held online on May 21 and 22, 2024, and consensus was obtained by means of the Delphi technique. After consensus feedback from co-authors was addressed and consensus was met, an updated version of the position statement was sent to three individuals with type 1 diabetes, three parents of children/adolescents with type 1 diabetes and three experts working in the field of PA and type 1 diabetes (June 14, 2024). After consideration of their comments, the final version of the joint EASD/ISPAD position statement was sent to the writing group for approval (August 30, 2024). The document was then reviewed by EASD’s Committee on Clinical Affairs (CCA) and endorsed by the Boards of EASD and ISPAD.

The writing group used previous position statements as guidance for the current position statement [1‒4]. A literature search was conducted by two independent researchers (O.M. and D.P.Z.) in PubMed, EMBASE, and The Cochrane Library for publications involving AID systems and PA in children/adolescents and/or adults with type 1 diabetes. Details on the keywords and the search strategies are available in online supplementary ESM 3.

The strengths of the recommendations in this position statement are categorised as A–D. Additionally, “consensus D” reflects clinical experience of respected authorities (see online suppl. ESM 1 for further details).

We recommend five key strategies for PA and diabetes management when using AID technology (see Text Box, “Consensus recommendations for PA and AID in type 1 diabetes”). The authors generalise that these consensus recommendations will work for a majority but not all types of PA [20] (consensus D).

As a wide range of exercise types (and activity settings) exist, different recommendations for PA are necessary depending on the type of activity performed and whether that activity is planned or unplanned (Table 1) [6‒8]. These general recommendations can serve as starting points that can be incorporated into each specific AID system. Most AID systems have the option to alter (i.e., raise and/or lower) the target glucose value before, during, and after PA, which can help maintain glucose levels in the target range. Depending on the AID system being used, the feature that raises the glucose target (or range) is sometimes called exercise mode, activity mode, activity feature, physical activity mode, or temporary target; however, for simplicity, we use the term “higher glucose target” for this feature in this position statement.

Table 1.

General considerations for PA with AID systems

InformationConsiderations for PA
CGM trend arrow 
  • Try to start an activity under stable glucose conditions (e.g., horizontal trend arrow) [63] (D)

  • Immediate postprandial PA increases hypoglycaemia risk [26, 31, 50, 66]; pay close attention to CGM trend arrows and initiate CHO intake if glucose drops <7.0 mmol/L (C)

 
Modification of glucose target (i.e., exercise announcement) 
  • Aim to set a higher glucose target 1–2 h before PA if a decrease in glucose is expected [25, 26, 46, 51] (A)

  • Setting a higher glucose target at activity onset is less likely to be effective at reducing hypoglycaemia risk, but is still encouraged, particularly when activity is prolonged (e.g., >30 min) (consensus D)

  • Exercise announcement (i.e., higher glucose target) may not be needed in all situations (i.e., if an increase in glucose is expected) [22, 23, 41] (D)

 
Insulin on board (IOB) 
  • Higher IOB is a predictor of hypoglycaemia risk during PA [63, 67, 68]; however, the calculation and display of IOB depends on the AID system used and various user settings

  • IOB displayed on an AID system does not accurately reflect peak insulin action, which typically occurs 1–2 h after the prandial bolus dose (see “Exercise time of day and prandial status” below)

  • In general, aim for low IOB at onset of PA if a decrease in glucose is expected during activity [63, 67, 68] (C)

  • Consider the unadjusted/regular prandial dose and/or insulin delivery settings if an increase in glucose is expected [23, 41] (D)

 
Carbohydrates (CHO) 
  • Optimise the type of CHO that works best for an individual to maintain glucose levels between 7.0 and 10.0 mmol/L (e.g., fast-acting CHO to treat or prevent hypoglycaemia; low glycaemic index CHO for meal pre-exercise) (consensus D)

  • Consume CHO in small amounts (10–20 g) just before PA if SG is <7.0 mmol/L before unplanned PA or as needed during PA (e.g., every 30 min)

  • Determine the threshold at which to consume CHO (e.g., <7.0 mmol/L) [4, 7, 28, 30, 31]. Overconsumption of CHO to treat hypoglycaemia may result in rebound hyperglycaemia followed by AID-induced hypoglycaemia during activity; consider consuming around 12–20 g CHO if hypoglycaemia occurs during activity (consensus D)

  • Always have fast-acting CHO available to prevent or treat hypoglycaemia

  • Consider carrying emergency supplies (e.g., glucagon) to treat severe hypoglycaemia

  • Various factors can impact the intra-individual amount of CHO intake needed around PA (e.g., menstrual cycle, hormones, puberty)

 
Exercise time of day and prandial status 
  • Reduce the prandial bolus insulin dose (by 25–33%) for meals consumed <2 h prior to PA when a decrease in glucose is expected during the activity [25, 26, 50] (A)

  • Employ the usual prandial bolus insulin dose for meals before PA when an increase in glucose is expected during the activity [22, 23, 41] (D)

  • In general, morning fasted PA is not associated with a large drop in glucose levels and may promote a rise compared with other times of the day (i.e., may be safer for hypoglycaemia risk mitigation) [67] (C)

 
Placement of insulin pump and CGM devices 
  • Based on activity type and user experience, consider location of the infusion set and CGM placement in areas less likely to fail (e.g., fall off) [4] (D)

  • Consider additional adhesives or overlay tape to protect devices during activity [4] (D)

  • Consider placing the insulin infusion set further away from actively working skeletal muscle (consensus D)

 
Blood glucose and blood ketone monitoring 
  • Have a glucose metre readily available (e.g., in case of CGM failure or malfunction, to confirm SG level) [4, 14, 15, 17, 19] (D)

  • Have a blood ketone metre (or urine ketone strips) available if the PA is prolonged or intense [4] (D)

 
InformationConsiderations for PA
CGM trend arrow 
  • Try to start an activity under stable glucose conditions (e.g., horizontal trend arrow) [63] (D)

  • Immediate postprandial PA increases hypoglycaemia risk [26, 31, 50, 66]; pay close attention to CGM trend arrows and initiate CHO intake if glucose drops <7.0 mmol/L (C)

 
Modification of glucose target (i.e., exercise announcement) 
  • Aim to set a higher glucose target 1–2 h before PA if a decrease in glucose is expected [25, 26, 46, 51] (A)

  • Setting a higher glucose target at activity onset is less likely to be effective at reducing hypoglycaemia risk, but is still encouraged, particularly when activity is prolonged (e.g., >30 min) (consensus D)

  • Exercise announcement (i.e., higher glucose target) may not be needed in all situations (i.e., if an increase in glucose is expected) [22, 23, 41] (D)

 
Insulin on board (IOB) 
  • Higher IOB is a predictor of hypoglycaemia risk during PA [63, 67, 68]; however, the calculation and display of IOB depends on the AID system used and various user settings

  • IOB displayed on an AID system does not accurately reflect peak insulin action, which typically occurs 1–2 h after the prandial bolus dose (see “Exercise time of day and prandial status” below)

  • In general, aim for low IOB at onset of PA if a decrease in glucose is expected during activity [63, 67, 68] (C)

  • Consider the unadjusted/regular prandial dose and/or insulin delivery settings if an increase in glucose is expected [23, 41] (D)

 
Carbohydrates (CHO) 
  • Optimise the type of CHO that works best for an individual to maintain glucose levels between 7.0 and 10.0 mmol/L (e.g., fast-acting CHO to treat or prevent hypoglycaemia; low glycaemic index CHO for meal pre-exercise) (consensus D)

  • Consume CHO in small amounts (10–20 g) just before PA if SG is <7.0 mmol/L before unplanned PA or as needed during PA (e.g., every 30 min)

  • Determine the threshold at which to consume CHO (e.g., <7.0 mmol/L) [4, 7, 28, 30, 31]. Overconsumption of CHO to treat hypoglycaemia may result in rebound hyperglycaemia followed by AID-induced hypoglycaemia during activity; consider consuming around 12–20 g CHO if hypoglycaemia occurs during activity (consensus D)

  • Always have fast-acting CHO available to prevent or treat hypoglycaemia

  • Consider carrying emergency supplies (e.g., glucagon) to treat severe hypoglycaemia

  • Various factors can impact the intra-individual amount of CHO intake needed around PA (e.g., menstrual cycle, hormones, puberty)

 
Exercise time of day and prandial status 
  • Reduce the prandial bolus insulin dose (by 25–33%) for meals consumed <2 h prior to PA when a decrease in glucose is expected during the activity [25, 26, 50] (A)

  • Employ the usual prandial bolus insulin dose for meals before PA when an increase in glucose is expected during the activity [22, 23, 41] (D)

  • In general, morning fasted PA is not associated with a large drop in glucose levels and may promote a rise compared with other times of the day (i.e., may be safer for hypoglycaemia risk mitigation) [67] (C)

 
Placement of insulin pump and CGM devices 
  • Based on activity type and user experience, consider location of the infusion set and CGM placement in areas less likely to fail (e.g., fall off) [4] (D)

  • Consider additional adhesives or overlay tape to protect devices during activity [4] (D)

  • Consider placing the insulin infusion set further away from actively working skeletal muscle (consensus D)

 
Blood glucose and blood ketone monitoring 
  • Have a glucose metre readily available (e.g., in case of CGM failure or malfunction, to confirm SG level) [4, 14, 15, 17, 19] (D)

  • Have a blood ketone metre (or urine ketone strips) available if the PA is prolonged or intense [4] (D)

 

Overall, a challenge of AID systems around PA is preventing an increase in algorithm-derived AID before the onset of exercise when the sensor glucose (SG) value is rising or already elevated because of a CHO snack or a reduction in prandial insulin delivery before the activity. However, if a higher glucose target is set prior to performing a manual prandial (bolus) insulin reduction and/or consuming an “uncovered snack” (i.e., a snack without prandial insulin administered), this effect is likely to be attenuated. Another challenge during PA and instances of hyperglycaemia is that some AID systems do not restrain insulin delivery effectively enough and/or they continue to give automatic insulin correction doses during PA, even if a higher glucose target is set. While we recommend keeping the AID system in automated mode during PA, in cases where the device still results in PA-related hypoglycaemia, we acknowledge that placing the AID system in manual mode before the activity begins may be necessary. We also acknowledge that suspending with or without disconnecting the AID system may be required in some settings, which may require an alternative insulin delivery method (such as taking insulin by injection or by reconnecting the insulin pump and giving a manual bolus intermittently).

Strategies for glucose management around PA with AID systems may also differ based on the timing and nature of the activity and whether that activity is planned or unplanned (Tables 1-3). For example, following an overnight fast or before a high-intensity sprint activity, a higher glucose target may be set close to the start of PA or may not be necessary at all [21‒24] (D); however, more research in this area is warranted [21]. For planned PA after a meal (up to 2 h after a meal), a higher glucose target should be set first, where possible, followed by performing a prandial bolus insulin reduction (e.g., around 25–33% reduction) to help reduce prandial insulin on board (IOB) and the risk of hypoglycaemia (see online suppl. ESM 1 for more details on IOB). In situations where the planned activity occurs more than 2 h after a meal, the higher glucose target should be set between 1 and 2 h beforehand and maintained until the end of the activity [25, 26] (A). For unplanned activity, AID systems may provide some protection against exercise-induced hypoglycaemia relative to other insulin delivery modalities when basal insulin delivery is fixed, but CHO intake is typically required, and to a greater extent, compared with planned activity [27]. As such, a recommendation for unplanned activity is still to set a higher glucose target from the start until the end of activity. If glucose levels drop below 7.0 mmol/L during PA, even with a higher glucose target set, we recommend that small amounts of fast-acting CHO be consumed based on the CGM trend arrow (see below), without announcing it to the AID system [7, 28‒31] (C).

  • 3–6 g for a horizontal trend arrow

  • 6–9 g for a slightly decreasing trend arrow

  • 9–12 g for a decreasing trend arrow

  • 12–20 g for two or three decreasing trend arrows

Table 2.

General AID strategies for planned PA based on pre-exercise starting glucose concentrations

Before PA (1–2 h before)During PA (every 20–30 min)After PAa
glucose levelstrategy
>15.0 mmol/L or previous history where increase in glucose expected during PA 
  • May not need higher glucose target

  • Usual prandial insulin dose before PA

  • If glucose >15.0 mmol/L (and symptoms of DKA exist), check infusion site for any noticeable causes of blockages, such as kinked tubing or pressure/discomfort at infusion site, and test for ketones

 
  • If glucose <7.0 mmol/L, consume 3–20 g CHO depending on CGM trend arrow

 
  • If glucose >15.0 mmol/L after PA, resume usual settings and continue to monitor ketones and glucose levels

 
5.0–15.0 mmol/L 
  • Start higher glucose targetb

  • If PA occurs <2 h after meal, perform 25–33% prandial bolus reduction

 
 
  • If glucose 5.0–15.0 mmol/L after PA, resume usual settings

 
<5.0 mmol/L or previous history where decrease in glucose expected during PA 
  • Start higher glucose targetb

  • If PA occurs <2 h after meal, perform 25–33% prandial bolus reduction

  • CHO (10–20 g) snack at PA onset with no prandial insulin

 
 
  • If glucose <5.0 mmol/L, consider 3–20 g CHO and maintain higher glucose target for up to 2 h post PA

 
Before PA (1–2 h before)During PA (every 20–30 min)After PAa
glucose levelstrategy
>15.0 mmol/L or previous history where increase in glucose expected during PA 
  • May not need higher glucose target

  • Usual prandial insulin dose before PA

  • If glucose >15.0 mmol/L (and symptoms of DKA exist), check infusion site for any noticeable causes of blockages, such as kinked tubing or pressure/discomfort at infusion site, and test for ketones

 
  • If glucose <7.0 mmol/L, consume 3–20 g CHO depending on CGM trend arrow

 
  • If glucose >15.0 mmol/L after PA, resume usual settings and continue to monitor ketones and glucose levels

 
5.0–15.0 mmol/L 
  • Start higher glucose targetb

  • If PA occurs <2 h after meal, perform 25–33% prandial bolus reduction

 
 
  • If glucose 5.0–15.0 mmol/L after PA, resume usual settings

 
<5.0 mmol/L or previous history where decrease in glucose expected during PA 
  • Start higher glucose targetb

  • If PA occurs <2 h after meal, perform 25–33% prandial bolus reduction

  • CHO (10–20 g) snack at PA onset with no prandial insulin

 
 
  • If glucose <5.0 mmol/L, consider 3–20 g CHO and maintain higher glucose target for up to 2 h post PA

 

CGM, continuous glucose monitoring (A–D); DKA, diabetic ketoacidosis.

aConsider 25–33% reduction in prandial bolus insulin with the following meal after PA. For unplanned PA, the frequency of CHO intake is generally expected to be higher and closer to the upper limit of CHO amount recommendations than that for planned PA.

bIf PA occurs after a meal where a prandial bolus dose is delivered, set the higher glucose target before performing the prandial bolus reduction.

Table 3.

General AID strategies for unplanned PA based on pre-exercise starting glucose concentrations

At PA onsetDuring PA (every 20–30 min)After PAa
glucose levelstrategy
>15.0 mmol/L or previous history where increase in glucose expected during PA 
  • May not need higher glucose target

  • If glucose >15.0 mmol/L (and symptoms of DKA exist), check infusion site for any noticeable causes of blockages, such as kinked tubing or pressure/discomfort at infusion site, and consider testing for ketones

 
  • If glucose <7.0 mmol/L, consume 3–20 g CHO depending on CGM trend arrow

 
• If glucose >15.0 mmol/L after PA, resume usual settings and continue to monitor ketones and glucose levels 
5.0–15.0 mmol/L 
  • Start higher glucose target until end of PAb

  • May consider CHO (10–20 g) snack at PA onset with no prandial bolus insulin

 
 
  • If glucose 5.0–15.0 mmol/L after PA, resume usual settings

 
<5.0 mmol/L or previous history where decrease in glucose expected during PA 
  • Start higher glucose target until end of PAb

  • CHO (10–20 g) snack at PA onset with no prandial bolus insulin

 
 
  • If glucose <5.0 mmol/L, consider 3–20 g CHO and maintain higher glucose target for up to 2 h post PA

 
At PA onsetDuring PA (every 20–30 min)After PAa
glucose levelstrategy
>15.0 mmol/L or previous history where increase in glucose expected during PA 
  • May not need higher glucose target

  • If glucose >15.0 mmol/L (and symptoms of DKA exist), check infusion site for any noticeable causes of blockages, such as kinked tubing or pressure/discomfort at infusion site, and consider testing for ketones

 
  • If glucose <7.0 mmol/L, consume 3–20 g CHO depending on CGM trend arrow

 
• If glucose >15.0 mmol/L after PA, resume usual settings and continue to monitor ketones and glucose levels 
5.0–15.0 mmol/L 
  • Start higher glucose target until end of PAb

  • May consider CHO (10–20 g) snack at PA onset with no prandial bolus insulin

 
 
  • If glucose 5.0–15.0 mmol/L after PA, resume usual settings

 
<5.0 mmol/L or previous history where decrease in glucose expected during PA 
  • Start higher glucose target until end of PAb

  • CHO (10–20 g) snack at PA onset with no prandial bolus insulin

 
 
  • If glucose <5.0 mmol/L, consider 3–20 g CHO and maintain higher glucose target for up to 2 h post PA

 

CGM, continuous glucose monitoring (A–D); DKA, diabetic ketoacidosis.

aConsider 25–33% reduction in prandial bolus insulin with the following meal after PA. For unplanned PA, the frequency of CHO intake is generally expected to be higher and closer to the upper limit of CHO amount recommendations than that for planned PA.

bIf PA occurs after a meal where a prandial bolus dose is delivered, set the higher glucose target before performing the prandial bolus reduction.

We also recommend checking the SG around 20–30 min after CHO consumption and repeating treatment if necessary [4] (D). For the post-exercise period, the higher glucose target should be stopped at the end of PA [32, 33]; however, exceptions to this rule may exist, such as after an unusually active day or when post-exercise late-onset hypoglycaemia persists [31, 34] (C).

AID-specific recommendations are listed alphabetically by company.

Beta Bionics iLet Bionic Pancreas

The iLet insulin-only Bionic Pancreas system’s glucose targets can be set to 6.1 (lower), 6.7 (usual), or 7.2 mmol/L (higher). Unlike other AID systems, the iLet is initialised based only on bodyweight and does not require discrete CHO input for meals; instead, users employ a qualitative approach to meals indicating if meal sizes are “Usual for me,” “More,” or “Less.” The system delivers ∼75% of the estimated insulin needs for a meal immediately and will automatically increase or decrease additional basal or correction insulin dosing in the postprandial period as needed. Correction doses are provided by the system and the user cannot override it to give a manual dose of insulin. In this AID system, IOB is estimated using a fixed model of insulin absorption into the blood and clearance from the blood that considers all correction boluses and meal boluses. IOB is computed every 5 min based on an assumed peak time of insulin in the blood after administration (tmax) of 65 min. Unlike most other AID systems, the duration of insulin action cannot be adjusted by the user.

Evidence on Glucose Management around PA with the iLet System

To date, the iLet system has been tested only in clinical trials of physically active youth and adults with type 1 diabetes, with no formal evaluation of how it performs during and after PA [35, 36]. Nonetheless, several exercise studies on earlier system designs, including dual-hormone (i.e., glucagon and insulin) configurations, have been conducted [37‒39]. Therefore, recommendations (Fig. 2) are given based primarily on studies from other AID systems with considerations on how the iLet system may be adjusted for PA (all recommendations are level D).

Fig. 2.

Recommendations for use of the iLet Bionic Pancreas system to manage glucose outcomes during PA. Consider insulin delivery suspension with or without disconnecting the iLet system 30 min prior to activity to help mitigate hypoglycaemia risk. If ingesting undeclared CHO and disconnecting the iLet before activity, ensure that the device is already suspended and disconnected prior to CHO ingestion. The prandial bolus insulin dose can be reduced only by “underestimating” CHO (i.e., entering a smaller meal size). SG, sensor glucose. Glucose values: 6.1 mmol/L = 110 mg/dL, 6.7 mmol/L = 120 mg/dL, 7.0 mmol/L = 126 mg/dL, 7.2 mmol/L = 130 mg/dL, 8.3 mmol/L = 150 mg/dL. See online supplementary ESM 1 for version of this figure with glucose concentrations in mg/dL. This figure is available as part of a downloadable slideset.

Fig. 2.

Recommendations for use of the iLet Bionic Pancreas system to manage glucose outcomes during PA. Consider insulin delivery suspension with or without disconnecting the iLet system 30 min prior to activity to help mitigate hypoglycaemia risk. If ingesting undeclared CHO and disconnecting the iLet before activity, ensure that the device is already suspended and disconnected prior to CHO ingestion. The prandial bolus insulin dose can be reduced only by “underestimating” CHO (i.e., entering a smaller meal size). SG, sensor glucose. Glucose values: 6.1 mmol/L = 110 mg/dL, 6.7 mmol/L = 120 mg/dL, 7.0 mmol/L = 126 mg/dL, 7.2 mmol/L = 130 mg/dL, 8.3 mmol/L = 150 mg/dL. See online supplementary ESM 1 for version of this figure with glucose concentrations in mg/dL. This figure is available as part of a downloadable slideset.

Close modal

Recommendations for Glucose Management around PA with the iLet System

The iLet system currently does not have a feature to allow a higher glucose target to be set prior to PA to help reduce the risk of hypoglycaemia during an activity. However, if using a glucose target of 6.1 or 6.7 mmol/L as the “usual target,” one option for PA may be to switch the glucose target to 7.2 mmol/L, ideally 1–2 h before the activity (consensus D). Users must remember to return the glucose target back to the usual target following PA. The prandial bolus insulin dose can be reduced for a pre-exercise meal bolus only by entering a smaller meal size into the device (i.e., select “Less” rather than “Usual for me”), which effectively reduces the bolus insulin dose by 50%.

One other point to consider is whether to leave the iLet connected during PA or whether it should be suspended with or without disconnecting the pump during some PA where the risk for hypoglycaemia is elevated. This approach may need to be personalised to the individual and the PA type and intensity (consensus D). Without the current option of setting a glucose target >7.2 mmol/L, in instances of increased hypoglycaemia risk, key strategies for this AID system around PA include (1) frequent checking and monitoring of real-time CGM values and trends pre-, during, and post-exercise; (2) having fast-acting CHO readily available to prevent or treat hypoglycaemia; and (3) aiming to limit the amount of CHO on board before PA when possible, to avoid increases in AID (consensus D). For individuals aiming to consume uncovered CHO before PA, one strategy is to consume CHO after suspending and disconnecting the iLet system, to avoid increases in AID (consensus D).

CamDiab Mylife CamAPS FX

The mylife CamAPS FX system allows a glucose target between 4.4 and 11.0 mmol/L to be set, with a default target of 5.8 mmol/L. Insulin delivery using auto-modulated insulin release based on the algorithm and manual correction doses is possible in auto-mode but is not recommended unless following an infusion set occlusion or similar.

In this system, any bolus insulin given through the bolus calculator (correction or meal related) counts toward IOB (displayed as “Active Insulin”). The active insulin time that is displayed to the user can be set between 2 and 8 h; however, the real active insulin time used by the algorithm is subject to adaptive learning and is automatically adjusted. Basal rate or algorithm-directed insulin delivery does not count toward IOB, and the programmed duration of insulin action does not affect the algorithm-directed insulin delivery. A realistic view of IOB can be visualised by turning the mobile phone to landscape (horizontal mode), which allows the last bolus dosing and pharmacokinetic profile of the basal rate of insulin delivery to be seen.

Two additional features are available in the mylife CamAPS FX system: the “Ease-off” mode, which delivers less insulin, raises the glucose target and suspends insulin delivery if glucose levels are <7.0 mmol/L; and the “Boost” mode, which increases the algorithm responsiveness to higher glucose levels by up to ∼35% while maintaining the same glucose target.

Evidence on Glucose Management during PA with the Mylife CamAPS FX System

Previous studies performed on children and adolescents with type 1 diabetes using the mylife CamAPS FX system demonstrated that use of the Ease-off mode for PA resulted in safe glucose levels during PA [31, 34]. Specifically, increasing the glucose target to 8.3 mmol/L and simultaneously starting the Ease-off mode 2 h before maximum cardiopulmonary exercise testing resulted in stable glucose levels in young people with type 1 diabetes (start 10.7 ± 3.1 mmol/L vs. end 10.5 ± 3.1 mmol/L; p = 0.69) [34]. In a ski camp study performed in children and adolescents with type 1 diabetes, it was also shown that starting the Ease-off mode 2 h before exercise was suitable for avoiding hypoglycaemia [31].

Recommendations for Glucose Management around PA with the Mylife CamAPS FX System

To reduce the risk of hypoglycaemia during activity, we suggest setting the Ease-off mode and/or increasing the glucose target 1–2 h before PA [31, 34] (C); this may be especially relevant in instances of high IOB or during aerobic exercise [40] (D). We recommend using the Boost mode if an increase in glucose is expected during PA [8] (D) (e.g., during high-intensity sprinting in the fasted state [41] (D)). If deemed useful by the user, caregiver or healthcare professional, both the Ease-off and the Boost mode can be pre-programmed in this system to automatically start and end at a predefined time when PA is expected, as described in Figure 3.

Fig. 3.

Illustration of how to set a new personal glucose target and how to set the Ease-off (now or later) mode when using the mylife CamAPS FX system. See online supplementary ESM 1 for version of this figure with glucose concentrations in mg/dL. This figure is available as part of a downloadable slideset.

Fig. 3.

Illustration of how to set a new personal glucose target and how to set the Ease-off (now or later) mode when using the mylife CamAPS FX system. See online supplementary ESM 1 for version of this figure with glucose concentrations in mg/dL. This figure is available as part of a downloadable slideset.

Close modal

For unplanned, low-to moderate-intensity PA, where a decrease in glucose levels is expected and the glucose level is already in a reasonable target range for PA (e.g., 5.0–7.0 mmol/L), the Ease-off mode and/or a higher glucose target should be set immediately, followed by consumption of 10–20 g of CHO at exercise onset [7, 42] (D). The suggestion is to announce this meal or snack as “hypoglycaemia treatment” in the “Add meal” function and not as a regular meal, otherwise the system will likely deliver insulin [31, 34] (C). As with any AID system, more CHO can be consumed during prolonged PA based on observed glucose trends and for performance reasons [43]. In contrast, for instances where a rise in glucose levels is expected during PA (e.g., during high-intensity PA in the overnight fasted state) [41], we recommend starting the Boost mode with the regular or lower glucose target at the onset of PA to help limit activity-related hyperglycaemia. We advise not starting the Boost mode well in advance of the onset of the activity as this might result in pre-exercise hypoglycaemia (consensus D).

As the Ease-off and Boost modes contribute to a lesser extent to algorithmic learning, these modes may be considered for individuals who perform more irregular PA (consensus D). For individuals who exercise more regularly with respect to specific days and times (e.g., Mon, Wed, Fri and Sun at ∼17:00), one option may be to set a specific glucose target depending on the time of day and type of PA that is typically performed, as described in Figure 3,(consensus D). For example, when glucose levels are expected to decrease during PA, consider setting a glucose target ≥8.3 mmol/L ∼ 2 h before activity. We also recommend that users set an individualised glucose target for PA when they are using either the Ease-off mode (i.e., higher glucose target) or the Boost mode (i.e., lower glucose target) to help achieve their desired glucose level [34] (D). All recommended adaptations concerning the Ease-off and Boost modes, as well as glucose targets, are provided in Figure 4.

Fig. 4.

Recommendations for use of the mylife CamAPS FX system to manage glucose outcomes during PA. Insulin delivery suspension with or without disconnection for prolonged periods (up to 120 min) may be required under some circumstances (e.g., swimming, diving, contact sports), although it is generally not recommended for most activities. SG, sensor glucose. Glucose values: 4.4 mmol/L = 80 mg/dL, 5.6 mmol/L = 100 mg/dL, 6.7 mmol/L = 120 mg/dL, 7.0 mmol/L = 126 mg/dL, 8.3 mmol/L = 150 mg/dL. See online supplementary ESM 1 for version of this figure with glucose concentrations in mg/dL. This figure is available as part of a downloadable slideset.

Fig. 4.

Recommendations for use of the mylife CamAPS FX system to manage glucose outcomes during PA. Insulin delivery suspension with or without disconnection for prolonged periods (up to 120 min) may be required under some circumstances (e.g., swimming, diving, contact sports), although it is generally not recommended for most activities. SG, sensor glucose. Glucose values: 4.4 mmol/L = 80 mg/dL, 5.6 mmol/L = 100 mg/dL, 6.7 mmol/L = 120 mg/dL, 7.0 mmol/L = 126 mg/dL, 8.3 mmol/L = 150 mg/dL. See online supplementary ESM 1 for version of this figure with glucose concentrations in mg/dL. This figure is available as part of a downloadable slideset.

Close modal

Diabeloop Generation 1

The Generation 1 (DBLG1) system’s default glucose target is 6.1 mmol/L, but the glucose target can be set between 5.6 and 7.2 mmol/L. The low glucose threshold when insulin delivery is stopped can be set between 3.3 and 4.7 mmol/L and the algorithm hyperglycaemia threshold is 10.0 mmol/L. The aggressiveness of insulin delivery of the DBLG1 system can be modified to deliver 59–147% of the typical basal rate when glucose is between 3.9 and 10.0 mmol/L. When SG is >10.0 mmol/L, the automated correction bolus can be set to deliver within the range of 43–186% of the typical automated correction bolus dose. The prandial insulin dose can also be set to deliver insulin in the range of 50–200% for breakfast, lunch and dinner. These functions may be used to adapt the prandial insulin dose for post-meal activity; however, this needs to be discussed, individualised and, in some cases, modified with support from the healthcare professional team. In this system, IOB (displayed as “Active Insulin”), as shown in the interface, corresponds to the IOB provided by regulation, including any insulin source confirmed by the pump (basal rate and bolus insulin).

Evidence on Glucose Management during PA with the DBLG1 System

In a post hoc analysis of an RCT, glycaemic outcomes were compared between days with and days without PA in 56 adults with type 1 diabetes using the DBLG1 system for 12 weeks [44]. Participants announced PA at least 30 min before exercise, which reduced insulin delivery, and, if necessary, a certain amount of CHO was also recommended by the system to avoid hypoglycaemia. Time below range (<3.0 mmol/L; TBR<3.0) was not significantly different between days with and days without PA, independent of exercise duration and intensity (2.0 ± 1.5% vs. 2.2 ± 1.1%; p > 0.05). Ingested CHO as a preventative strategy against hypoglycaemia as recommended by the system were significantly higher on days with PA (41.1 ± 35.5 vs. 21.8 ± 28.5 g/day; p < 0.001), and the AID insulin dose was significantly lower on days with PA (31.5 ± 10.5 vs. 34.0 ± 10.5 U/day; p < 0.001). The time above range (>10.0 mmol/L; TAR>10.0) was 28.7 ± 9.3% on days with PA compared with 26.8 ± 8.6% on days without PA (p = 0.017). Time in range (3.9–10.0 mmol/L; TIR3.9–10.0) was 69.1 ± 8.2% on days with PA versus 70.9 ± 8.2% on days without PA (p = 0.017). The coefficient of variation in glucose was higher on days with PA than days without (32.0 ± 3.7% vs. 30.9 ± 3.7%; p = 0.019), indicating increased glycaemic variability on exercise days. Another study performed on adults with type 1 diabetes showed that the DBLG1 system was superior to open-loop insulin delivery with respect to TIR3.9–10.0 and TAR>10.0 when the “Physical Activity” mode was set 30 min before the start of activity [45].

Recommendations for Glucose Management around PA with the DBLG1 System

The Physical Activity mode can be used to decrease the risk of hypoglycaemia during PA (Fig. 5). In this mode, the glucose target and hypoglycaemia threshold are increased by 3.9 mmol/L, which reduces the aggressiveness of insulin delivery. When the Physical Activity mode is used, the PA intensity can be set to low, moderate, or intense and the planned duration of PA can be set. Both the duration and intensity are considered a matrix, with coefficients modulating the insulin basal rate, corrective bolus, or meal bolus. Another feature of the DBLG1 system is the “ZEN” mode, which increases the glucose target by an increment that is between 0.6 and 2.2 mmol/L for a period of 1–8 h [8] (D).

Fig. 5.

Illustration of how to start, stop, and modify/delete the Physical Activity mode in the DBLG1 system. See online supplementary ESM 1 for version of this figure with glucose concentrations in mg/dL. This figure is available as part of a downloadable slideset.

Fig. 5.

Illustration of how to start, stop, and modify/delete the Physical Activity mode in the DBLG1 system. See online supplementary ESM 1 for version of this figure with glucose concentrations in mg/dL. This figure is available as part of a downloadable slideset.

Close modal

We recommend starting Physical Activity mode at least 30 min before PA as the DBLG1 system suggests consuming a specific amount of CHO to avoid hypoglycaemia [44] (C). However, it is also beneficial to start Physical Activity mode earlier (e.g., between 1 and 2 h before the start of PA) as this has been shown to reduce the risk of hypoglycaemia (Fig. 6) [31, 46] (D). When PA is announced more than 1 h before the start of the activity, the target glucose is increased by 3.9 mmol/L, and the system aims to raise blood glucose prior to the start of PA. However, if glucose levels are <8.9 mmol/L 15 min before the start of PA, a specific CHO intake is recommended by the system. When PA is announced closer to the start of the activity, the system only provides a recommendation for CHO intake 15 min before PA if glucose is <8.9 mmol/L.

Fig. 6.

Recommendations for use of the DBLG1 system to manage glucose outcomes during PA. Insulin delivery suspension with or without disconnection for prolonged periods (up to 120 min) may be required under some circumstances (e.g., swimming, diving, contact sports), although it is generally not recommended for most activities as several of these strategies cannot be implemented and/or require modification. SG, sensor glucose. Glucose values: 5.6 mmol/L = 100 mg/dL, 6.1 mmol/L = 110 mg/dL, 6.7 mmol/L = 120 mg/dL, 7.0 mmol/L = 126 mg/dL. See online supplementary ESM 1 for version of this figure with glucose concentrations in mg/dL. This figure is available as part of a downloadable slideset.

Fig. 6.

Recommendations for use of the DBLG1 system to manage glucose outcomes during PA. Insulin delivery suspension with or without disconnection for prolonged periods (up to 120 min) may be required under some circumstances (e.g., swimming, diving, contact sports), although it is generally not recommended for most activities as several of these strategies cannot be implemented and/or require modification. SG, sensor glucose. Glucose values: 5.6 mmol/L = 100 mg/dL, 6.1 mmol/L = 110 mg/dL, 6.7 mmol/L = 120 mg/dL, 7.0 mmol/L = 126 mg/dL. See online supplementary ESM 1 for version of this figure with glucose concentrations in mg/dL. This figure is available as part of a downloadable slideset.

Close modal

Furthermore, the DBLG1 system automatically reduces the basal rate of insulin delivery for 16 h after Physical Activity mode is enabled to help mitigate the risk of post-exercise hypoglycaemia caused by increased insulin sensitivity. Physical Activity mode also allows the user to name and save the PA session (e.g., football) and provide the duration and intensity (e.g., low, moderate, intense). Figure 6 provides recommendations for managing glucose levels during PA using the DBLG1 system.

Insulet Omnipod 5

The Omnipod 5 system is a tubeless AID system [8, 47] that uses SmartAdjust technology to predict glucose values 60 min in advance and dynamically adjusts basal insulin delivery every 5 min. SmartAdjust targets glucose levels between 6.1 and 8.3 mmol/L, with levels set by the user, caregiver or healthcare professional. Different targets can be programmed for different hours of the day. With each Pod change, usually occurring at least every 72 h, the Omnipod 5 system automatically calculates an adaptive basal rate based on a fading memory of insulin requirements over 6 days. Furthermore, Omnipod 5 is a waterproof patch pump (i.e., Pod) that can provide users with increased flexibility in daily activities, in particular, with water-based activities [48].

In this system, the IOB is the sum of the correction IOB (insulin remaining in the body from previous correction doses), meal IOB (insulin remaining in the body from previous meal boluses) and Omnipod 5 software IOB (i.e., all insulin delivered by the system). IOB is mainly determined by the “Duration of Insulin Action” setting, which ranges from 2 to 6 h. Furthermore, the “Reverse Correction” feature deducts the IOB from the bolus calculation when the current glucose value is below the target glucose value [47].

Evidence on Glucose Management during PA with the Omnipod 5 System

In the pivotal trial of the Omnipod 5 system, an exercise study was conducted in 59 adults with type 1 diabetes. Participants underwent three, 60 min moderate-intensity treadmill exercise sessions in which (1) the “Activity” feature (higher glucose target) was set 30 min prior to exercise; (2) the Activity feature was set 60 min prior to exercise; and (3) usual AID was continued with no adjustment made for exercise [49]. Not surprisingly, at the start of exercise in sessions (1) and (2), insulin delivery was lower and glucose was higher with use of the Activity feature than with usual AID.

Recommendations for Glucose Management around PA with the Omnipod 5 System

For PA, the higher glucose target in the Omnipod 5 system is 8.3 mmol/L. This target attenuates AID and can be programmed to last from 1 to 24 h [6]. For activities that lead to an increased risk of hypoglycaemia, the recommendation with this system is to set the Activity feature 1–2 h before PA until the end of the activity (Fig. 7).

Fig. 7.

Illustration of how to set and cancel the Activity feature on the Omnipod 5 system. See online supplementary ESM 1 for version of this figure with glucose concentrations in mg/dL. This figure is available as part of a downloadable slideset.

Fig. 7.

Illustration of how to set and cancel the Activity feature on the Omnipod 5 system. See online supplementary ESM 1 for version of this figure with glucose concentrations in mg/dL. This figure is available as part of a downloadable slideset.

Close modal

If the usual glucose target is set to 6.7, 7.2, 7.8, or 8.3 mmol/L, and glucose is expected to increase during PA (e.g., fasted, high-intensity PA), we recommend lowering the usual glucose target to 6.1 mmol/L prior to the onset of PA and resuming the usual glucose target after the PA event (consensus D) (Fig. 8). With Omnipod 5, up to eight different targets can be programmed throughout the day, so there is some flexibility around what glucose target is set and when. Therefore, for school-aged children, higher glucose targets can be leveraged to account for usual after-school sports by setting the target higher 1–2 h prior to the scheduled activity until the end of the activity (consensus D).

Fig. 8.

Recommendations for use of the Omnipod 5 system to manage glucose outcomes during PA. SG, sensor glucose. Glucose values: 6.1 mmol/L = 110 mg/dL, 6.7 mmol/L = 120 mg/dL, 7.0 mmol/L = 126 mg/dL. See online supplementary ESM 1 for version of this figure with glucose concentrations in mg/dL. This figure is available as part of a downloadable slideset.

Fig. 8.

Recommendations for use of the Omnipod 5 system to manage glucose outcomes during PA. SG, sensor glucose. Glucose values: 6.1 mmol/L = 110 mg/dL, 6.7 mmol/L = 120 mg/dL, 7.0 mmol/L = 126 mg/dL. See online supplementary ESM 1 for version of this figure with glucose concentrations in mg/dL. This figure is available as part of a downloadable slideset.

Close modal

A more general consideration for healthcare professionals is how the “Reverse Correction” feature might impact insulin delivery at the meal before PA. With the Reverse Correction feature on, the prandial bolus dose will be reduced if the pre-meal glucose level is below target. If this feature is combined with a manual prandial bolus insulin reduction initiated by the user (e.g., 25–33% reduction) prior to activity, then glucose will likely rise and result in AID by the system, thereby increasing hypoglycaemia risk during PA. To date, there are no published studies to support specific guidance on using Reverse Correction around PA.

If PA is planned <2 h following a meal and a drop in glucose is anticipated, a 25–33% reduction in prandial bolus insulin is generally recommended [26, 50] (C). Another option is to turn the Reverse Correction feature off when applying a prandial bolus insulin reduction before PA (consensus D). For the meal prior to the onset of PA, the bolus insulin amount on the Omnipod 5 system can be reduced by either (1) entering fewer CHO into the system than the amount being consumed or (2) decreasing the recommended bolus insulin amount by 25% up to 100% (i.e., no bolus) [8]. Importantly, research trials on the amount and timing of prandial bolus insulin reductions before PA in children and adults with type 1 diabetes using Omnipod 5 are not currently available. Figure 8 provides recommendations for managing glucose levels during PA using the Omnipod 5 system.

Medtronic MiniMed 780G

The MiniMed 780G system using SmartGuard technology can set glucose targets of 5.5, 6.1, 6.7, and 8.3 mmol/L (“Temp Target,” exercise mode). One of the major safety features of the Temp Target is the prevention of automatic bolus correction doses in response to rising glucose levels from ingestion of CHO immediately before or during PA. Without this feature, there is likely to be a significant increase in IOB during PA when CHO is given, which can result in a recurrent cycle of hypoglycaemic episodes. The auto-correction bolus, when enabled, automatically delivers bolus insulin doses when the algorithm has been delivering auto-basal insulin at the maximum insulin limit; the SG value is >6.7 mmol/L and the calculated correction bolus is >10% of the maximum insulin limit. Furthermore, the auto-correction bolus can be switched off in the SmartGuard settings. When SmartGuard technology is used for calculating the bolus insulin dose for CHO, the dose suggestion is increased or decreased based on the actual glucose value and the total IOB.

The IOB (displayed as “Active Insulin”) accounts for bolus insulin, including meal boluses, manual correction boluses, and automatic correction boluses. Basal insulin, either from a pre-programmed basal rate or from SmartGuard auto-basal insulin delivery, is excluded from active insulin. The displayed IOB is affected by the Active Insulin time settings (2–8 h, adjustable). Active Insulin is also used in the calculation of correction boluses (both manual and automated).

Evidence on Glucose Management during PA with the MiniMed 780G System

The Medtronic MiniMed AID systems are suitable for use during PA in people with type 1 diabetes and are the systems with the largest body of published literature related to PA [7]. In a trial of ten adults with type 1 diabetes, it was shown that transitioning from open-loop systems to the MiniMed 780G system did not significantly alter glucose levels during and after 45 min of moderate-intensity exercise [51].

McCarthy et al. [26] demonstrated in adults with type 1 diabetes that glucose levels may be optimised during exercise when using the MiniMed 780G system by reducing the pre-exercise prandial bolus insulin dose by 25% for meals consumed up to 90 min before exercise. This study also showed that increasing the glucose target at the onset of exercise or 45 min prior to the start of exercise was less effective for avoiding hypoglycaemia than setting a higher glucose target 90 min before exercise when prandial insulin was reduced by 25%. In a study of youth with type 1 diabetes using the MiniMed 780G system, it was determined that, independent of the type of insulin used (faster-acting insulin aspart vs. standard insulin aspart), exercise was safe, with a TBR (<3.9 mmol/L glucose) of 2.8% versus 2.5%, respectively, when the Temp Target was set at least 1 h before exercise [27].

In a preliminary, controlled, in-clinic research study by Lee et al. [46] TIR3.9–10.0 was 100% for 45 min of high-intensity exercise or moderate-intensity exercise when the Temp Target on the MiniMed 670G system was started 2 h prior to the start of exercise in adults with type 1 diabetes who also had impaired awareness of hypoglycaemia. Use of the MiniMed advanced hybrid closed-loop system with different insulins (faster-acting insulin aspart and insulin aspart) did not significantly alter the risk of nocturnal hypoglycaemia on exercise days compared with non-exercise days [33, 52]. Furthermore, when comparing different types of exercise (high-intensity exercise, resistance exercise, moderate-intensity exercise), there were no differences in glycaemic outcomes [46, 53] or risk of nocturnal post-exercise hypoglycaemia [32].

Recommendations for Glucose Management around PA with the MiniMed 780G System

When using the MiniMed 780G system, we recommend adjusting the glucose target based on the anticipated glucose response to exercise [53‒55] (D). In instances where a glucose decrease is expected during planned PA, one option is to initiate the Temp Target 1–2 h before PA [26, 46]), which will automatically stop after a set duration (Fig. 9, 10). In general, the Temp Target should be timed to stop near the end of the activity [32, 33]. For unplanned PA, when a glucose decrease is expected, CHO consumption (e.g., 10–20 g) will likely be necessary, particularly if glucose levels at exercise onset are <7.0 mmol/L [42] (C). In addition, the Temp Target should be turned on immediately prior to CHO consumption [8] (D) (Fig. 10).

Fig. 9.

Illustration of how to set and cancel the Temp Target on the MiniMed 780G system. See online supplementary ESM 1 for version of this figure with glucose concentrations in mg/dL. This figure is available as part of a downloadable slideset.

Fig. 9.

Illustration of how to set and cancel the Temp Target on the MiniMed 780G system. See online supplementary ESM 1 for version of this figure with glucose concentrations in mg/dL. This figure is available as part of a downloadable slideset.

Close modal
Fig. 10.

Recommendations for use of the MiniMed 780G system to manage glucose outcomes during PA. Insulin delivery suspension with or without disconnection for prolonged periods (up to 120 min) may be required under some circumstances (e.g., swimming, diving, contact sports), although it is generally not recommended for most activities as several of these strategies cannot be implemented and/or require modification. SG, sensor glucose. Glucose values: 5.6 mmol/L = 100 mg/dL, 6.1 mmol/L = 110 mg/dL, 6.7 mmol/L = 120 mg/dL, 7.0 mmol/L = 126 mg/dL. See online supplementary ESM 1 for version of this figure with glucose concentrations in mg/dL. This figure is available as part of a downloadable slideset.

Fig. 10.

Recommendations for use of the MiniMed 780G system to manage glucose outcomes during PA. Insulin delivery suspension with or without disconnection for prolonged periods (up to 120 min) may be required under some circumstances (e.g., swimming, diving, contact sports), although it is generally not recommended for most activities as several of these strategies cannot be implemented and/or require modification. SG, sensor glucose. Glucose values: 5.6 mmol/L = 100 mg/dL, 6.1 mmol/L = 110 mg/dL, 6.7 mmol/L = 120 mg/dL, 7.0 mmol/L = 126 mg/dL. See online supplementary ESM 1 for version of this figure with glucose concentrations in mg/dL. This figure is available as part of a downloadable slideset.

Close modal

If glucose levels are expected to rise during PA, a lower target glucose may be more appropriate (e.g., 5.5 mmol/L) as this will likely result in greater insulin delivery than when a higher target is set. However, setting glucose targets should be based on an individual’s average glucose responses, which may vary depending on PA type, time of day, CHO fuelling strategies, menstrual cycle phase, and other factors [56] (D) (Fig. 10). In general, we recommend keeping the MiniMed 780G system in automated mode during PA when a glucose decrease is expected, in addition to setting a Temp Target, and reducing prandial bolus insulin by 25–33% to minimise hypoglycaemia [50, 51].

If there is going to be a reduction in the prandial bolus insulin for the meal preceding PA, this should always be implemented in conjunction with the Temp Target, which disables the automated bolus function. Otherwise, the meal recognition software will signal that food has been eaten, and the device will try to address the initial rise in glucose levels with automated bolus insulin.

Tandem t:slim X2 with Control-IQ

The t:slim X2 insulin pump using Control-IQ technology predicts glucose levels 30 min ahead and adjusts insulin delivery accordingly, including the delivery of automated correction boluses up to once per hour if needed. The “Personal Profile” settings have a standard glucose target of 6.1 mmol/L, but the system targets a glucose range between 6.2 and 8.9 mmol/L. A higher glucose range between 7.8 and 8.9 mmol/L can be set for PA (referred to as “Exercise” mode). In “Sleep” mode, the target shifts to a tighter range (between 6.3 and 6.7 mmol/L), using basal adjustments, and this mode does not perform any auto-correction boluses. While this feature is designed for overnight glucose management, individuals can create sleep schedules at other times of the day to automatically leverage the transition to this tighter target range.

Non-customisable parameters in the Control-IQ system include the duration of active insulin and the glucose target [57]. If Control-IQ technology is enabled, IOB (displayed as “Insulin on Board”) includes all basal insulin delivered above and below the programmed basal rate, in addition to all bolus insulin delivered (not adjustable; set to 5 h). Up to six Personal Profiles can be programmed in which the individual can adjust the basal insulin doses, insulin-to-CHO ratio and insulin sensitivity factor (ISF) settings. Thus, for those performing different types of PA, different settings can be programmed. In addition, a correction dose (up to 60% of the dose determined by the correction factor or ISF) is delivered a maximum of once per hour if the predicted glucose value 30 min later is anticipated to be > 10.0 mmol/L, the system is not in Sleep mode, and there has been no user-initiated bolus in the past hour [58]. Recently, a new pump design from Tandem called the Tandem Mobi pump has been released. The Mobi pump uses the Control-IQ algorithm but has the benefit of a smaller physical footprint and is fully controllable from a user’s phone via a mobile app (with a button available on the pump to permit bolus delivery without the app).

Evidence on Glucose Management around PA with the t:slim X2 Control-IQ System

The Tandem Control-IQ Artificial Pancreas system was the first AID system tested in adolescents and children (aged 6–18 years) with type 1 diabetes in an outpatient exercise setting (i.e., during winter sports, particularly skiing) [59‒61]. These real-world studies demonstrated that Control-IQ technology improved glycaemic metrics and reduced hypoglycaemia risk during prolonged winter sporting activities in this group compared with a non-AID system. Mameli et al. [62] recently evaluated TIR3.9–10.0 during 2 h of outdoor physical activity, planned 90 min after lunch, in youth aged 9–18 years using the t:slim X2 pump with Control-IQ technology. In this study, group A underwent endurance activities for 60 min (1,000 m run, jump circuit) followed by power activities for 60 min (80 m run and a long jump), and group B underwent power activities for 60 min followed by endurance activities for 60 min. A higher glucose target (Exercise mode, see below) was set 90 min before exercise until dinner time and pre-exercise prandial bolus insulin was reduced by 50%. In this study, group A and group B participants had similar TIR3.9–10.0 during the 2 h of activity (50.4%, 95% CI: 33.8, 75.0 vs. 39.6%, 95% CI: 26.9, 58.3; p = 0.39). No TBR<3.0 occurred during the 2 h of activity (both the endurance and the power workout) [62].

Recommendations for Glucose Management around PA with the t:slim X2 Control-IQ System

Individuals with type 1 diabetes using this system can announce PA using Control-IQ’s Exercise mode, which aims to maintain glucose levels between 7.8 and 8.9 mmol/L (Fig. 11). When using Exercise mode, Control-IQ technology will decrease the algorithm-derived insulin delivery rate (or basal insulin delivery rate) when the glucose level is predicted to be <7.8 mmol/L 30 min in the future and will increase the insulin delivery rate when the glucose is predicted to be > 8.9 mmol/L 30 min in the future [6]. If the glucose prediction 30 min in the future is expected to exceed 10.0 mmol/L, an automated correction bolus equivalent to 60% of the dose calculated by the ISF will be delivered up to once per hour, even if Exercise mode is active [58]. This may increase the risk for activity-related hypoglycaemia in some settings where a rise in glucose occurs prior to PA (e.g., from unannounced CHO intake). Importantly, an automated correction bolus will not occur within 60 min of any bolus (of any amount) that has been delivered or cancelled. With software version 7.7 (introduced in some countries in January 2024), Exercise mode can be set for a duration of 30 min to 8 h. Otherwise, for software versions below version 7.7, the user will need to manually turn Exercise mode off after PA.

Fig. 11.

Illustration of how to set the Exercise mode on the t:slim X2 Control-IQ system. See online supplementary ESM 1 for version of this figure with glucose concentrations in mg/dL. This figure is available as part of a downloadable slideset.

Fig. 11.

Illustration of how to set the Exercise mode on the t:slim X2 Control-IQ system. See online supplementary ESM 1 for version of this figure with glucose concentrations in mg/dL. This figure is available as part of a downloadable slideset.

Close modal

If CHO is given during PA to treat pending or actual hypoglycaemia, the rise in glucose from treatment can trigger insulin delivery even during Exercise mode. Based on the current evidence, the general recommendation for this AID system is the same as that for other AID systems that have a higher glucose target for PA: to minimise excessive CHO feeding before and during the activity. We recommend that Exercise mode is enabled 1–2 h before the start of PA until the end of the activity to decrease IOB and reduce the risk of hypoglycaemia during activity [46] (D).

It may also be helpful for individuals with type 1 diabetes who regularly engage in PA and who may benefit from changes to various pump settings for more physically active days or days with more prolonged activity periods to create different Personal Profiles [8] (D). Personal profiles can be optimised in cases where glucose levels are expected to drop or rise during PA, and this may include adjustments to the basal insulin doses, insulin-to-CHO ratio, and/or ISF. For instances where glucose levels are expected to decrease during PA (e.g., walking, running, cycling), a Personal Profile can be created that can reduce insulin delivery (e.g., lower basal insulin doses; higher ISF; lower insulin-to-CHO ratio), which can be selected when deemed appropriate (Fig. 12).

Fig. 12.

Recommendations for use of the t:slim X2 Control-IQ system to manage glucose outcomes during PA. Consider adding a minimal manual bolus dose (e.g., 0.05 U) close to the onset of exercise to block the system from administering auto-correction doses for the next 60 min. Insulin delivery suspension with or without disconnection for prolonged periods (up to 120 min) may be required under some circumstances (e.g., swimming, diving, contact sports), although it is generally not recommended for most activities as several of these strategies cannot be implemented and/or require modification. ISF, insulin sensitivity factor; SG, sensor glucose. Glucose value: 7.0 mmol/L = 126 mg/dL. See online supplementary ESM 1 for version of this figure with glucose concentrations in mg/dL. This figure is available as part of a downloadable slideset.

Fig. 12.

Recommendations for use of the t:slim X2 Control-IQ system to manage glucose outcomes during PA. Consider adding a minimal manual bolus dose (e.g., 0.05 U) close to the onset of exercise to block the system from administering auto-correction doses for the next 60 min. Insulin delivery suspension with or without disconnection for prolonged periods (up to 120 min) may be required under some circumstances (e.g., swimming, diving, contact sports), although it is generally not recommended for most activities as several of these strategies cannot be implemented and/or require modification. ISF, insulin sensitivity factor; SG, sensor glucose. Glucose value: 7.0 mmol/L = 126 mg/dL. See online supplementary ESM 1 for version of this figure with glucose concentrations in mg/dL. This figure is available as part of a downloadable slideset.

Close modal

Based on consensus D, another possible strategy for this AID system is to consider adding a small manual bolus insulin dose (note that the minimum bolus is 0.05 U for this system) close to the onset of PA, which then disables the auto-bolus feature for the next 60 min, even if CHO are consumed and a rise in glucose occurs.

In other situations where glucose levels are expected to rise during PA (e.g., during high-intensity sprinting), a different Personal Profile can be created (e.g., higher basal insulin doses, lower ISF) (consensus D). Another option is to retain the usual AID settings (i.e., do not set “Exercise” mode) or consider putting the pump into Sleep mode for PA and turning Sleep mode off after PA to resume the usual AID settings post-exercise (consensus D). As a reminder, the system has a lower glucose target range in Sleep mode, but it does not deliver automated boluses during this time. As the targets are tighter, Sleep mode may be an option for instances when glucose levels tend to rise with PA, although research around the utility of this approach is warranted (consensus D). Figure 12 provides recommendations for managing glucose levels during PA using the t:slim X2 Control-IQ system.

There is a lack of evidence and recommendations on glucose management strategies during PA under special circumstances for people using AID technology [1‒4]. To better address some of these unique situations, Table 4 provides a summary of special PA circumstances, important considerations and possible strategies to help individuals with type 1 diabetes using AID systems. For additional details on specific considerations for PA, see online supplementary ESM 1.

Table 4.

Summary of special PA circumstances, important considerations, and possible strategies for individuals with type 1 diabetes using AID systems (consensus D)

Special circumstancesConsiderationsPossible strategies to limit PA dysglycaemia (consensus D)
Long-duration PA events (e.g., ultramarathon, ironman, triathlon, road bike racing, prolonged trekking/hiking) 
  • Low insulin delivery

  • Increased risk of ketones

  • Risk of hypoglycaemia during exercise

  • Risk of hyperglycaemia

 
  • Consider regular CHO feeding throughout activity

  • To prevent insulin deficiency during PA, may avoid higher glucose target for full duration of event

  • Avoid large, fully uncovered snacks during event that may drive the AID system to increase AID

  • Monitor CGM closely and consider setting higher alert thresholds on the CGM device where possible

 
Prolonged insulin pump disconnection during PA (e.g., ≥120 min pump disconnection for contact sports and/or water-based activities as described below) 
  • Low insulin delivery

  • Increased risk of ketones

  • Risk of hyperglycaemia

 
  • Suspend the AID system to inform algorithm of reduced insulin delivery, which will be viewable in data reports

  • If suspending the AID system, consider reconnecting the pump every hour for small amount of insulin delivery

  • If disconnected from the AID system, to prevent insulin deficiency during PA, consider reconnecting the pump and delivering ∼50% of the “usual basal” dose every hour

  • May need to switch to manual mode

  • May consider “un-tethered” approach combining long-acting insulin dose by injection/pen and pump (removed during PA) and/or consider switching to a patch pump

 
Competition stress (e.g., football game) 
  • Stress response

  • Hyperglycaemia before and during event

  • Delayed hypoglycaemia following event

 
  • Monitor IOB near competition start because the AID system is likely to increase insulin delivery with stress response and rising glucose levels

  • Stay hydrated

  • May want to avoid setting higher glucose target 1–2 h before competition

  • If glucose rises to >15.0 mmol/L with competition stress, consider a manual partial bolus correction (e.g., 50% of usual correction dose)

 
Water-based activities (e.g., swimming, surfing, scuba diving) 
  • Lack of device communication between CGM and AID controller (little to no automation)

  • Hypoglycaemia must be strictly avoided for safety

 
  • For the Omnipod 5 system, consider keeping the Pod and sensor closer together to permit increased Bluetooth communication and continued automation

  • Avoid high IOB at start of water-based activities

  • Consider insulin delivery suspension with or without disconnecting the AID system where possible

  • If suspending the AID system, consider reconnecting the pump every hour for small amount of insulin delivery

  • Note increased risk of hypoglycaemia with water-based sports; have fast-acting glucose readily available

 
Contact sports (e.g., wrestling, rugby, football, jiu-jitsu, water polo, platform diving) 
  • Stress response

  • Devices falling off or being knocked off

  • Lack of comfort

 
  • Consider location of device wear for increased comfort

  • Consider insulin delivery suspension with or without disconnecting the AID system where possible if concerned about damaging the pump

  • Use overlay patches and tape to secure the CGM and AID system in place

  • If disconnected from the AID system, to prevent insulin deficiency during PA, consider reconnecting the pump and delivering ∼50% of the “usual basal” dose every hour

  • May want to avoid setting higher glucose target 1–2 h before PA if hyperglycaemia occurs with contact sports

 
High ambient temperature (e.g., hiking in the heat) 
  • Increased perspiration

  • Increased insulin absorption

  • Increased energy expenditure

  • Increased risk of hypoglycaemia

  • Possible impact on CGM accuracy

 
  • Consider fingerstick blood glucose monitoring in cases where CGM accuracy may be impacted

  • Set higher glucose target 1–2 h before exercise until end of activity

  • Reduce mealtime or snack bolus insulin dose before and during activity (e.g., 25–33% reduction)

  • Use overlay patches and tape to secure the CGM and AID system in place

 
Low ambient temperature (e.g., skiing, cold exposure) 
  • Constriction of peripheral blood vessels, minimal sweat production, and increased metabolic heat production (e.g., shivering)

  • May be less risk for hypoglycaemia based on stress response to cold but likely depends on duration and intensity of activity

 
  • Concerns over glucose gel and/or insulin in pump freezing due to temperature so keep these items close to the body

  • Consider carrying hypoglycaemia treatment that cannot freeze (e.g., glucose tablets)

  • Loss of signal is possible with glucose metre and/or CGM so may need to warm up devices prior to testing

  • Monitor CGM closely and consider setting higher alert thresholds on the CGM device where possible

 
High-altitude environments (e.g., skiing, trekking, snowboarding) 
  • Hypoxia may impact decision-making

  • Impact on counterregulatory hormones

  • Increased dysglycaemia from hypoxia

  • Possible impact on CGM accuracy

 
  • Blood glucose metres may be inaccurate at high altitude; therefore, CGM could be recommended for combined use

  • Stay hydrated

  • Lack of evidence on strategies for using AID systems at high altitude

  • Short-term exposure to high altitude might induce insulin resistance through sympathetic stimulation, but insulin resistance will decrease after several days of exposure

  • Above 5,000 m, more cortisol will be produced and this may increase the risk of hyperglycaemia

 
Special circumstancesConsiderationsPossible strategies to limit PA dysglycaemia (consensus D)
Long-duration PA events (e.g., ultramarathon, ironman, triathlon, road bike racing, prolonged trekking/hiking) 
  • Low insulin delivery

  • Increased risk of ketones

  • Risk of hypoglycaemia during exercise

  • Risk of hyperglycaemia

 
  • Consider regular CHO feeding throughout activity

  • To prevent insulin deficiency during PA, may avoid higher glucose target for full duration of event

  • Avoid large, fully uncovered snacks during event that may drive the AID system to increase AID

  • Monitor CGM closely and consider setting higher alert thresholds on the CGM device where possible

 
Prolonged insulin pump disconnection during PA (e.g., ≥120 min pump disconnection for contact sports and/or water-based activities as described below) 
  • Low insulin delivery

  • Increased risk of ketones

  • Risk of hyperglycaemia

 
  • Suspend the AID system to inform algorithm of reduced insulin delivery, which will be viewable in data reports

  • If suspending the AID system, consider reconnecting the pump every hour for small amount of insulin delivery

  • If disconnected from the AID system, to prevent insulin deficiency during PA, consider reconnecting the pump and delivering ∼50% of the “usual basal” dose every hour

  • May need to switch to manual mode

  • May consider “un-tethered” approach combining long-acting insulin dose by injection/pen and pump (removed during PA) and/or consider switching to a patch pump

 
Competition stress (e.g., football game) 
  • Stress response

  • Hyperglycaemia before and during event

  • Delayed hypoglycaemia following event

 
  • Monitor IOB near competition start because the AID system is likely to increase insulin delivery with stress response and rising glucose levels

  • Stay hydrated

  • May want to avoid setting higher glucose target 1–2 h before competition

  • If glucose rises to >15.0 mmol/L with competition stress, consider a manual partial bolus correction (e.g., 50% of usual correction dose)

 
Water-based activities (e.g., swimming, surfing, scuba diving) 
  • Lack of device communication between CGM and AID controller (little to no automation)

  • Hypoglycaemia must be strictly avoided for safety

 
  • For the Omnipod 5 system, consider keeping the Pod and sensor closer together to permit increased Bluetooth communication and continued automation

  • Avoid high IOB at start of water-based activities

  • Consider insulin delivery suspension with or without disconnecting the AID system where possible

  • If suspending the AID system, consider reconnecting the pump every hour for small amount of insulin delivery

  • Note increased risk of hypoglycaemia with water-based sports; have fast-acting glucose readily available

 
Contact sports (e.g., wrestling, rugby, football, jiu-jitsu, water polo, platform diving) 
  • Stress response

  • Devices falling off or being knocked off

  • Lack of comfort

 
  • Consider location of device wear for increased comfort

  • Consider insulin delivery suspension with or without disconnecting the AID system where possible if concerned about damaging the pump

  • Use overlay patches and tape to secure the CGM and AID system in place

  • If disconnected from the AID system, to prevent insulin deficiency during PA, consider reconnecting the pump and delivering ∼50% of the “usual basal” dose every hour

  • May want to avoid setting higher glucose target 1–2 h before PA if hyperglycaemia occurs with contact sports

 
High ambient temperature (e.g., hiking in the heat) 
  • Increased perspiration

  • Increased insulin absorption

  • Increased energy expenditure

  • Increased risk of hypoglycaemia

  • Possible impact on CGM accuracy

 
  • Consider fingerstick blood glucose monitoring in cases where CGM accuracy may be impacted

  • Set higher glucose target 1–2 h before exercise until end of activity

  • Reduce mealtime or snack bolus insulin dose before and during activity (e.g., 25–33% reduction)

  • Use overlay patches and tape to secure the CGM and AID system in place

 
Low ambient temperature (e.g., skiing, cold exposure) 
  • Constriction of peripheral blood vessels, minimal sweat production, and increased metabolic heat production (e.g., shivering)

  • May be less risk for hypoglycaemia based on stress response to cold but likely depends on duration and intensity of activity

 
  • Concerns over glucose gel and/or insulin in pump freezing due to temperature so keep these items close to the body

  • Consider carrying hypoglycaemia treatment that cannot freeze (e.g., glucose tablets)

  • Loss of signal is possible with glucose metre and/or CGM so may need to warm up devices prior to testing

  • Monitor CGM closely and consider setting higher alert thresholds on the CGM device where possible

 
High-altitude environments (e.g., skiing, trekking, snowboarding) 
  • Hypoxia may impact decision-making

  • Impact on counterregulatory hormones

  • Increased dysglycaemia from hypoxia

  • Possible impact on CGM accuracy

 
  • Blood glucose metres may be inaccurate at high altitude; therefore, CGM could be recommended for combined use

  • Stay hydrated

  • Lack of evidence on strategies for using AID systems at high altitude

  • Short-term exposure to high altitude might induce insulin resistance through sympathetic stimulation, but insulin resistance will decrease after several days of exposure

  • Above 5,000 m, more cortisol will be produced and this may increase the risk of hyperglycaemia

 

Under some circumstances (e.g., prolonged periods of time in the water with pump suspension that can increase the risk of diabetic ketoacidosis), people may choose to switch to multiple daily injection (MDI) therapy for a certain period of time. However, this should be carried out only under the guidance of healthcare professionals and the care team.

In this joint EASD/ISPAD position statement, we provide both general strategies and AID device-specific recommendations to help healthcare professionals and people with type 1 diabetes use these emerging technologies more effectively for planned and unplanned PA. We stress that these recommendations should serve as a starting point for PA and that individual responses to activity should be learnt and discussed with the healthcare professional team. Strategies often require fine-tuning and individuals should be prepared for unpredictable glucose responses to PA, even after these strategies have been implemented. While there is individual variation in glycaemic responses to the diverse types of activities that individuals with type 1 diabetes perform, we hope that these evidence-informed recommendations can help individuals optimise glucose self-management in various PA settings.

The online version contains peer-reviewed but unedited supplementary material including a slideset of the figures for download, which is available to authorised users at https://doi.org/10.1007/s00125-024-06308-z.

We would like to thank the nine reviewers (three experts in the field of PA, AID, and type 1 diabetes; three people with type 1 diabetes using AID systems; and three parents of children/adolescents with type 1 diabetes using AID systems) for critical assessment of this manuscript prior to submission to the EASD CCA. We also want to thank K. Moser for support with the production of the figures and tables. We extend our gratitude to all individuals living with type 1 diabetes who have generously contributed their time and participated in research trials. Finally, we would like to acknowledge the following individuals who have contributed to clinical trials, data collection, and advancing the field of exercise and type 1 diabetes research: D. Morrison (University of Melbourne, Australia), B. Paldus (University of Melbourne, Australia), O. McCarthy (Steno Diabetes Center Copenhagen, Denmark), G. Ash (Yale University, USA), L. Nally (Yale University, USA), M. Clements (Children’s Mercy, Kansas City, USA), R. Gal (Jaeb Center for Health Research, USA), S. Patton (Nemours Children's Health, USA), N. Bratina (University of Ljubljana, Slovenia), F. Annan (University College London Hospitals NHS Foundation Trust, UK), C. Smart (John Hunter Children’s Hospital, Australia), E. Heyman (University of Lille, France), and S. Tagougui (University of Lille, France).

O.M. has received research funding from Sêr Cymru II COFUND Fellowship/European Union, Novo Nordisk, Abbott Diabetes Care, Sanofi, Dexcom, Team Novo Nordisk, SAIL, Brauerei Maisel, Medtronic, European Foundation for the Study of Diabetes (EFSD)/EASD, Falke, BISp., Ypsomed, DiabetesDE and Perfood. D.P.Z. has received an ISPAD-JDRF Research Fellowship and research support from Insulet and the Leona M. and Harry B. Helmsley Charitable Trust. P.A. has received research grants from Medtronic and Novo Nordisk. K.D. has received an ISPAD-JDRF Research Fellowship. T.B.’s institution has received research grant support from Abbott, Medtronic, Novo Nordisk, Sanofi, Novartis, Sandoz, Zealand Pharma, the Slovenian Research Agency, the National Institutes of Health (NIH) and the European Union. Yale School of Medicine has received research support for J.L.S. from Abbott Diabetes, the Jaeb Center for Health Research, JDRF, Insulet, Medtronic, NIH and Provention Bio. J.E.Y. has received in-kind research support from Dexcom and LifeScan Canada and support from Medtronic, Dexcom and Abbott. K.N.’s institution has received research funding from Zealand Pharma, Novo Nordisk, Medtronic and Dexcom. H.S. has received research support from Boehringer Ingelheim, Eli Lilly, Novo Nordisk, the European Union, the Austrian Research Promotion Agency (FFG) and the Austrian Science Fund (KLP3413324). B.B. has received research support from the NIDDK and Breakthrough T1D. KU Leuven has received research support for C.M. from Medtronic, Novo Nordisk, Sanofi and ActoBio Therapeutics. M.C.R. has received research grants and/or contract research support from the Jaeb Center for Health Research, Sanofi, Novo Nordisk, Eli Lilly, Dexcom, Insulet and Zucara Therapeutics. G.P.F. has received research funding from NIH, JDRF/BT1D, NSF, Medtronic, Dexcom, Abbott, Tandem, Insulet, Beta Bionics, and Lilly. E.A.D.’s institution has received research support from Medtronic, Dexcom and Tandem. N.O. has received research support from Dexcom, Medtronic, Roche Diabetes, Diabetes UK, Imperial College NIHR BRC, the Leona M. and Harry B. Helmsley Charitable Trust and UKRI EME. For P.G., KU Leuven received research grants from Dexcom (financial and non-financial support), Medtronic, Novo Nordisk, Roche, Sanofi, and Tandem. P.G. is the recipient of a senior clinical research fellowship from FWO, the Flemish Research Council (1861924N). R.M.B. has received research funding or product supply from Medtronic, Novo Nordisk, Sêr Cymru II COFUND European Union, Abbott Diabetes Care, Sanofi, Dexcom, Team Novo Nordisk, Supersapiens and Beneo. R.H. reports receiving speaker honoraria from Eli Lilly, Dexcom and Novo Nordisk, and license and/or consultancy fees from B. Braun and Abbott Diabetes Care, patents related to closed-loop, and being director at CamDiab. T.D. serves as Chief Medical Officer of Breakthrough T1D (formerly JDRF). None of the funders played a role in the development of this position paper.

O.M. has received lecture fees from Medtronic, Eli Lilly, Novo Nordisk, Sanofi, TAD Pharma, Theras, Diatec, Ypsomed, Dexcom, AstraZeneca, Insulet and Perfood; and is on the advisory board for Sanofi, TAD Pharma, Glaice, Perfood, Medtronic and Dexcom. D.P.Z. has received honoraria for speaking engagements from Ascensia Diabetes Care, Insulet Canada, Dexcom Canada and Medtronic; and is on an advisory board for Dexcom and the Diabetes Research Hub. P.A. has taken part in advisory boards of Eli Lilly, Insulet, Medtronic, Novo Nordisk, Roche and Sanofi. K.D. has received honoraria for speaking engagements from Abbott, Eli Lilly, Medtronic, Novo Nordisk and Pfizer; is on an advisory board for Medtronic and Novo Nordisk. S.E.H. has received speaker’s honoraria from Eli Lilly, Sanofi, Medtronic, Insulet, Dexcom and Ypsomed. J.P. has received speaker’s honoraria from Dexcom and Abbott and sits on the advisory board for Roche. T.B. has served on advisory panels of Novo Nordisk, Sanofi, Eli Lilly, Boehringer, Medtronic, Abbott and Indigo Diabetes and has received honoraria for participating on the speakers bureaus of Eli Lilly, Novo Nordisk, Medtronic, Abbott, Sanofi, Dexcom, Aventis, AstraZeneca and Roche. J.L.S. reports serving, or having served, on advisory panels for Cecelia Health, Insulet, Medtronic Diabetes, StartUp Health Diabetes Moonshot and Vertex and having served as a consultant to Abbott Diabetes, Insulet, Medtronic Diabetes, Vertex and Zealand. R.R.-L. has received consulting/advisory panel honoraria from Abbott, AstraZeneca, Bayer, Boehringer Ingelheim, Dexcom, Eli Lilly, HLS Therapeutics, INESSS, Insulet, Janssen, Medtronic, Merck, Novo Nordisk, Pfizer and Sanofi-Aventis; honoraria for conferences from Abbott, AstraZeneca, Boehringer Ingelheim, CPD Network, Dexcom, CMS Canadian Medical & Surgical Knowledge Translation Research Group, Eli Lilly, Janssen, Medtronic, Merck, Novo Nordisk, Sanofi-Aventis, Tandem and Vertex Pharmaceutical; consumable gifts (in kind) from Eli Lilly and Medtronic; and purchase fees from Eli Lilly in the field of automated insulin delivery. J.E.Y. has received speaker’s fees from Dexcom and Abbott. J.K.M. is a member of the advisory boards of Abbott Diabetes Care, Becton-Dickinson/Embecta, Biomea, Eli Lilly, Medtronic, Novo Nordisk, Pharmasens, Roche Diabetes Care, Sanofi and Viatris; has received speaker honoraria from Abbott Diabetes Care, A. Menarini Diagnostics, Becton-Dickinson/Embecta, Boehringer Ingelheim, Eli Lilly, MedTrust, Novo Nordisk, Roche Diabetes Care, Sanofi, Servier and Ypsomed; and is a shareholder of decide Clinical Software and elyte Diagnostics. M.T. has received honoraria for speaking engagements from Ely Lilly, Medtronic and Ypsomed and is on advisory boards for Abbott Diabetes Care and Sanofi. K.N. owns shares in Novo Nordisk and has been a paid consultant for Novo Nordisk and Medtronic and has received speaker and advisory board honorarium for her institution from Abbott, Medtronic, Novo Nordisk, Insulet and Convatec. H.S. has received speaker’s honoraria or is on the advisory board for Amgen, Amarin, Boehringer Ingelheim, Cancom, Eli Lilly, Daiichi Sankyo and Novo Nordisk. M.C.R. serves on advisory panels for Zealand Pharma, Zucara Therapeutics and Indigo Diabetes; acts as a consultant for the Jaeb Center for Health Research; has given lectures sponsored by Dexcom, Novo Nordisk and Sanofi; and is a shareholder, or holds stocks in, Supersapiens and Zucara Therapeutics. B.A.B. has received consulting fees from Ypsomed, Arecor and Medtronic. C.M. serves or has served on the advisory panel for Novo Nordisk, Sanofi, Eli Lilly, Novartis, Dexcom, Boehringer Ingelheim, Bayer, Roche, Medtronic, Insulet, Biomea Fusion, SAB Bio and Vertex. Financial compensation for these activities has been received by KU Leuven. C.M. serves or has served on the speakers bureau for Novo Nordisk, Sanofi, Eli Lilly, Medtronic and Boehringer Ingelheim. Financial compensation for these activities has been received by KU Leuven. C.M. is president of EASD. All external support provided to EASD can be found at www.easd.org. D.N.O. has received honoraria from Medtronic, Insulet, Abbott Diabetes Care, Novo Nordisk and Sanofi, and research support from Medtronic, Insulet, Dexcom, Roche, GlySens, BioCapillary and Endogenex, and is on advisory boards for Medtronic, Insulet, Abbott Diabetes Care, Ypsomed, Novo Nordisk and Sanofi. GPF has served as a speaker/consultant/ad board member for Medtronic, Dexcom, Abbott, Tandem, Insulet, Beta Bionics, Sequel and Lilly. E.A.D.’s institution has received speaker honorariums from Eli Lilly and E.A.D. has sat on an advisory panel for Tandem. N.S.O. has participated in advisory groups for Dexcom, Medtronic and Roche Diabetes and received fees for speaking from Astra Zeneca, Sanofi, Dexcom, Tandem, Medtronic and Roche Diabetes. P.G. has served on the advisory board for Insulet and Ypsomed. P.G. reports consulting fees from Abbott, Bayer and Medtronic, and honoraria for speaking from Abbott, Bayer, Dexcom, Insulet, Medtronic, Novo Nordisk, Vitalaire and Ypsomed (financial compensation received by KU Leuven). PG has received support for attending (virtual) conferences/meetings from Medtronic, Novo Nordisk, Roche and Sanofi (financial compensation received by KU Leuven). R.M.B. has received lecture fees from Abbott Diabetes Care, Novo Nordisk, Medtronic, Eli Lilly and Sanofi. T.D. has received speaker, advisory panel or research support from Abbott, Dexcom, Eli Lilly, Insulet, Medtronic, Novo Nordisk, Roche, Sanofi, Ypsomed and Vitalaire. He is a shareholder of DreaMed Ltd. The authors declare that there are no other relationships or activities that might bias, or be perceived to bias, their work.

All authors of this joint EASD/ISPAD position statement substantially contributed to conception and design, acquisition of data or analysis and interpretation of data, drafted the article or revised it critically for important intellectual content and approved the final version to be published.

Additional Information

Othmar Moser and Dessi P. Zaharieva contributed equally to this work.This article is being simultaneously published in Diabetologia (https://doi.org/10.1007/s00125-024-06308-z) and Hormone Research in Paediatrics (https://doi.org/10.1159/000542287) by the EASD and ISPAD. This position statement was reviewed for EASD by its Committee on Clinical Affairs, and approved by the Boards of EASD and ISPAD.

All data used within this position statement are included in the manuscript.

1.
Adolfsson
P
,
Taplin
CE
,
Zaharieva
DP
,
Pemberton
J
,
Davis
EA
,
Riddell
MC
, et al
.
ISPAD clinical practice consensus guidelines 2022: exercise in children and adolescents with diabetes
.
Pediatr Diabetes
.
2022
;
23
(
8
):
1341
72
.
2.
Riddell
MC
,
Gallen
IW
,
Smart
CE
,
Taplin
CE
,
Adolfsson
P
,
Lumb
AN
, et al
.
Exercise management in type 1 diabetes: a consensus statement
.
Lancet Diabetes Endocrinol
.
2017
;
5
:
377
90
.
3.
Colberg
SR
,
Sigal
RJ
,
Yardley
JE
,
Riddell
MC
,
Dunstan
DW
,
Dempsey
PC
, et al
.
Physical activity/exercise and diabetes: a position statement of the American diabetes association
.
Diabetes Care
.
2016
;
39
(
11
):
2065
79
.
4.
Moser
O
,
Riddell
MC
,
Eckstein
ML
,
Adolfsson
P
,
Rabasa-Lhoret
R
,
van den Boom
L
, et al
.
Glucose management for exercise using continuous glucose monitoring (CGM) and intermittently scanned CGM (isCGM) systems in type 1 diabetes: position statement of the European Association for the Study of Diabetes (EASD) and of the International Society for Pediatric and Adolescent Diabetes (ISPAD) endorsed by JDRF and supported by the American Diabetes Association (ADA)
.
Diabetologia
.
2020
;
63
(
12
):
2501
20
.
5.
Phillip
M
,
Nimri
R
,
Bergenstal
RM
,
Barnard-Kelly
K
,
Danne
T
,
Hovorka
R
, et al
.
Consensus recommendations for the use of automated insulin delivery technologies in clinical practice
.
Endocr Rev
.
2023
;
44
(
2
):
254
80
.
6.
Zaharieva
DP
,
Morrison
D
,
Paldus
B
,
Lal
RA
,
Buckingham
BA
,
O’Neal
DN
.
Practical aspects and exercise safety benefits of automated insulin delivery systems in type 1 diabetes
.
Diabetes Spectr
.
2023
;
36
(
2
):
127
36
.
7.
O’Neal
DN
,
Zaharieva
DP
,
Morrison
D
,
McCarthy
O
,
Nørgaard
K
.
Exercising safely with the MiniMed™ 780G automated insulin delivery system
.
Diabetes Technol Ther
.
2024
;
26
(
S3
):
84
96
.
8.
Zimmer
RT
,
Auth
A
,
Schierbauer
J
,
Haupt
S
,
Wachsmuth
N
,
Zimmermann
P
, et al
.
(Hybrid) closed-loop systems: from announced to unannounced exercise
.
Diabetes Technol Ther
.
2023
.
9.
Sherr
JL
,
Heinemann
L
,
Fleming
GA
,
Bergenstal
RM
,
Bruttomesso
D
,
Hanaire
H
, et al
.
Automated insulin delivery: benefits, challenges, and recommendations. A consensus report of the joint diabetes technology working group of the European association for the study of diabetes and the American diabetes association
.
Diabetologia
.
2023
;
66
(
1
):
3
22
.
10.
Garg
SK
,
McVean
JJ
.
Development and future of automated insulin delivery (AID) systems
.
Diabetes Technol Ther
.
2024
;
26
(
S3
):
1
6
.
11.
Garcia-Tirado
J
,
Colmegna
P
,
Villard
O
,
Diaz
JL
,
Esquivel-Zuniga
R
,
Koravi
CLK
, et al
.
Assessment of meal anticipation for improving fully automated insulin delivery in adults with type 1 diabetes
.
Diabetes Care
.
2023
;
46
(
9
):
1652
8
.
12.
Shalit
R
,
Minsky
N
,
Laron-Hirsh
M
,
Cohen
O
,
Kurtz
N
,
Roy
A
, et al
.
Unannounced meal challenges using an advanced hybrid closed-loop system
.
Diabetes Technol Ther
.
2023
;
25
(
9
):
579
88
.
13.
Parent
C
,
Lespagnol
E
,
Berthoin
S
,
Tagougui
S
,
Heyman
J
,
Stuckens
C
, et al
.
Barriers to physical activity in children and adults living with type 1 diabetes: a complex link with real-life glycemic excursions
.
Can J Diabetes
.
2023
;
47
(
2
):
P124
32
.
14.
Zaharieva
DP
,
Turksoy
K
,
McGaugh
SM
,
Pooni
R
,
Vienneau
T
,
Ly
T
, et al
.
Lag time remains with newer real-time continuous glucose monitoring technology during aerobic exercise in adults living with type 1 diabetes
.
Diabetes Technol Ther
.
2019
;
21
(
6
):
313
21
.
15.
Li
A
,
Riddell
MC
,
Potashner
D
,
Brown
RE
,
Aronson
R
.
Time lag and accuracy of continuous glucose monitoring during high intensity interval training in adults with type 1 diabetes
.
Diabetes Technol Ther
.
2019
;
21
(
5
):
286
94
.
16.
Guillot
FH
,
Jacobs
PG
,
Wilson
LM
,
Youssef
JE
,
Gabo
VB
,
Branigan
DL
, et al
.
Accuracy of the Dexcom G6 glucose sensor during aerobic, resistance, and interval exercise in adults with type 1 diabetes
.
Biosensors
.
2020
;
10
(
10
):
138
.
17.
Da Prato
G
,
Pasquini
S
,
Rinaldi
E
,
Lucianer
T
,
Donà
S
,
Santi
L
, et al
.
Accuracy of CGM systems during continuous and interval exercise in adults with type 1 diabetes
.
J Diabetes Sci Technol
.
2022
;
16
(
6
):
1436
43
.
18.
Moser
O
,
Sternad
C
,
Eckstein
ML
,
Szadkowska
A
,
Michalak
A
,
Mader
JK
, et al
.
Performance of intermittently scanned continuous glucose monitoring systems in people with type 1 diabetes: a pooled analysis
.
Diabetes Obes Metab
.
2022
;
24
(
3
):
522
9
.
19.
Moser
O
,
Mader
J
,
Tschakert
G
,
Mueller
A
,
Groeschl
W
,
Pieber
TR
, et al
.
Accuracy of continuous glucose monitoring (CGM) during continuous and high-intensity interval exercise in patients with type 1 diabetes mellitus
.
Nutrients
.
2016
;
8
(
8
):
489
.
20.
Oliva-Lozano
JM
,
Alacid
F
,
López-Miñarro
PA
,
Muyor
JM
.
What are the physical demands of sexual intercourse? A systematic review of the literature
.
Arch Sex Behav
.
2022
;
51
(
3
):
1397
417
.
21.
Yardley
JE
.
Reassessing the evidence: prandial state dictates glycaemic responses to exercise in individuals with type 1 diabetes to a greater extent than intensity
.
Diabetologia
.
2022
;
65
(
12
):
1994
9
.
22.
Toghi-Eshghi
SR
,
Yardley
JE
.
Morning (fasting) vs afternoon resistance exercise in individuals with type 1 diabetes: a randomized crossover study
.
J Clin Endocrinol Metab
.
2019
;
104
(
11
):
5217
24
.
23.
Fahey
AJ
,
Paramalingam
N
,
Davey
RJ
,
Davis
EA
,
Jones
TW
,
Fournier
PA
.
The effect of a short sprint on postexercise whole-body glucose production and utilization rates in individuals with type 1 diabetes mellitus
.
J Clin Endocrinol Metab
.
2012
;
97
(
11
):
4193
200
.
24.
Justice
TD
,
Hammer
GL
,
Davey
RJ
,
Paramalingam
N
,
Guelfi
KJ
,
Lewis
L
, et al
.
Effect of antecedent moderate-intensity exercise on the glycemia-increasing effect of a 30-sec maximal sprint: a sex comparison
.
Physiol Rep
.
2015
;
3
(
5
):
e12386
.
25.
Myette-Coté
É
,
Molveau
J
,
Wu
Z
,
Raffray
M
,
Devaux
M
,
Tagougui
S
, et al
.
A randomized crossover pilot study evaluating glucose control during exercise initiated 1 or 2 h after a meal in adults with type 1 diabetes treated with an automated insulin delivery system
.
Diabetes Technol Ther
.
2023
;
25
(
2
):
122
30
.
26.
McCarthy
OM
,
Christensen
MB
,
Kristensen
KB
,
Schmidt
S
,
Ranjan
AG
,
Bain
SC
, et al
.
Automated insulin delivery around exercise in adults with type 1 diabetes: a pilot randomized controlled study
.
Diabetes Technol Ther
.
2023
;
25
(
7
):
476
84
.
27.
Dovc
K
,
Bergford
S
,
Fröhlich-Reiterer
E
,
Zaharieva
DP
,
Potocnik
N
,
Müller
A
, et al
.
A comparison of faster insulin aspart with standard insulin aspart using hybrid automated insulin delivery system in active children and adolescents with type 1 diabetes: a randomized double-blind crossover trial
.
Diabetes Technol Ther
.
2023
;
25
(
9
):
612
21
.
28.
Moser
O
,
Muller
A
,
Aberer
F
,
Aziz
F
,
Kojzar
H
,
Sourij
C
, et al
.
Comparison of insulin glargine 300 U/mL and insulin degludec 100 U/mL around spontaneous exercise sessions in adults with type 1 diabetes: a randomized cross-over trial (ULTRAFLEXI-1 study)
.
Diabetes Technol Ther
.
2023
;
25
(
3
):
161
8
.
29.
Laffel
LM
,
Aleppo
G
,
Buckingham
BA
,
Forlenza
GP
,
Rasbach
LE
,
Tsalikian
E
, et al
.
A practical approach to using trend arrows on the Dexcom G5 CGM system to manage children and adolescents with diabetes
.
J Endocr Soc
.
2017
;
1
(
12
):
1461
76
.
30.
Moser
O
,
Eckstein
ML
,
Mueller
A
,
Birnbaumer
P
,
Aberer
F
,
Koehler
G
, et al
.
Pre-exercise blood glucose levels determine the amount of orally administered carbohydrates during physical exercise in individuals with type 1 diabetes-A randomized cross-over trial
.
Nutrients
.
2019
;
11
(
6
):
1287
.
31.
Moser
O
,
Mueller
A
,
Andritz
E
, et al
.
Performance of automated insulin delivery systems during a winter ski camp in children and adolescents with type 1 diabetes: the InSki trial
.
Diabetologia
.
2024
;
67
(
1 Suppl l
):
490
. (Abstract).
32.
Paldus
B
,
Morrison
D
,
Zaharieva
DP
,
Lee
MH
,
Jones
H
,
Obeyesekere
V
, et al
.
A randomized crossover trial comparing glucose control during moderate-intensity, high- intensity, and resistance exercise with hybrid closed-loop insulin delivery while profiling potential additional signals in adults with type 1 diabetes
.
Diabetes Care
.
2022
;
45
(
1
):
194
203
.
33.
Morrison
D
,
Zaharieva
DP
,
Lee
MH
,
Paldus
B
,
Vogrin
S
,
Grosman
B
, et al
.
Comparable glucose control with fast-acting insulin aspart versus insulin aspart using a second-generation hybrid closed-loop system during exercise
.
Diabetes Technol Ther
.
2022
;
24
(
2
):
93
101
.
34.
Moser
O
,
Sanfilippo
S
,
Grothoff
A
,
Hoffmann
S
,
Wachsmuth
N
,
Voit
T
, et al
.
Sicherheit und Effizienz des mylife CamAPS FX Hybrid Closed-Loop Systems während maximaler Ausbelastungstests bei Kindern und Jugendlichen mit Diabetes Mellitus Typ 1: eine explorative präliminäre Analyse der InLoopChild Studie
.
Diabetolgie und Stoffwechsel
.
2024
;
19
(
S1
):
S41
.
(Abstract) [in German]
.
35.
Messer
LH
,
Buckingham
BA
,
Cogen
F
,
Daniels
M
,
Forlenza
G
,
Jafri
RZ
, et al
.
Positive impact of the bionic pancreas on diabetes control in youth 6-17 Years old with type 1 diabetes: a multicenter randomized trial
.
Diabetes Technol Ther
.
2022
;
24
(
10
):
712
25
.
36.
Bionic Pancreas Research Group
;
Russell
SJ
,
Beck
RW
,
Damiano
ER
,
El-Khatib
FH
,
Ruedy
KJ
, et al
.
Multicenter, randomized trial of a bionic pancreas in type 1 diabetes
.
N Engl J Med
.
2022
;
387
(
13
):
1161
72
.
37.
Russell
SJ
,
El-Khatib
FH
,
Nathan
DM
,
Magyar
KL
,
Jiang
J
,
Damiano
ER
.
Blood glucose control in type 1 diabetes with a bihormonal bionic endocrine pancreas
.
Diabetes Care
.
2012
;
35
(
11
):
2148
55
.
38.
Russell
SJ
,
El-Khatib
FH
,
Sinha
M
,
Magyar
KL
,
McKeon
K
,
Goergen
LG
, et al
.
Outpatient glycemic control with a bionic pancreas in type 1 diabetes
.
N Engl J Med
.
2014
;
371
(
4
):
313
25
.
39.
Russell
SJ
,
Hillard
MA
,
Balliro
C
,
Magyar
KL
,
Selagamsetty
R
,
Sinha
M
, et al
.
Day and night glycaemic control with a bionic pancreas versus conventional insulin pump therapy in preadolescent children with type 1 diabetes: a randomised crossover trial
.
Lancet Diabetes Endocrinol
.
2016
;
4
(
3
):
233
43
.
40.
Fushimi
E
,
Aiello
EM
,
Cho
S
,
Riddell
MC
,
Gal
RL
,
Martin
CK
, et al
.
Online classification of unstructured free-living exercise sessions in people with Type 1 Diabetes
.
Diabetes Technol Ther
.
2024
;
26
(
10
):
709
19
.
41.
McClure
RD
,
Alcántara-Cordero
FJ
,
Weseen
E
,
Maldaner
M
,
Hart
S
,
Nitz
C
, et al
.
Systematic review and meta-analysis of blood glucose response to high-intensity interval exercise in adults with type 1 diabetes
.
Can J Diabetes
.
2023
;
47
(
2
):
171
9
.
42.
Zaharieva
DP
,
Messer
LH
,
Paldus
B
,
O’Neal
DN
,
Maahs
DM
,
Riddell
MC
.
Glucose control during physical activity and exercise using closed loop technology in adults and adolescents with type 1 diabetes
.
Can J Diabetes
.
2020
;
44
(
8
):
740
9
.
43.
Mattsson
S
,
Jendle
J
,
Adolfsson
P
.
Carbohydrate loading followed by high carbohydrate intake during prolonged physical exercise and its impact on glucose control in individuals with diabetes type 1-an exploratory study
.
Front Endocrinol
.
2019
;
10
:
571
.
44.
Franc
S
,
Benhamou
PY
,
Borot
S
,
Chaillous
L
,
Delemer
B
,
Doron
M
, et al
.
No more hypoglycaemia on days with physical activity and unrestricted diet when using a closed-loop system for 12 weeks: a post hoc secondary analysis of the multicentre, randomized controlled Diabeloop WP7 trial
.
Diabetes Obes Metab
.
2021
;
23
(
9
):
2170
6
.
45.
Hanaire
H
,
Franc
S
,
Borot
S
,
Penfornis
A
,
Benhamou
PY
,
Schaepelynck
P
, et al
.
Efficacy of the Diabeloop closed-loop system to improve glycaemic control in patients with type 1 diabetes exposed to gastronomic dinners or to sustained physical exercise
.
Diabetes Obes Metab
.
2020
;
22
(
3
):
324
34
.
46.
Lee
MH
,
Vogrin
S
,
Paldus
B
,
Jayawardene
D
,
Jones
HM
,
McAuley
SA
, et al
.
Glucose and counterregulatory responses to exercise in adults with type 1 diabetes and impaired awareness of hypoglycemia using closed-loop insulin delivery: a randomized crossover study
.
Diabetes Care
.
2020
;
43
(
2
):
480
3
.
47.
Berget
C
,
Sherr
JL
,
DeSalvo
DJ
,
Kingman
RS
,
Stone
SL
,
Brown
SA
, et al
.
Clinical implementation of the Omnipod 5 automated insulin delivery system: key considerations for training and onboarding people with diabetes
.
Clin Diabetes
.
2022
;
40
(
2
):
168
84
.
48.
Hughes
MS
,
Kingman
RS
,
Hsu
L
,
Lal
RA
,
Buckingham
BA
,
Zaharieva
DP
.
Swimming with the Omnipod 5 automated insulin delivery system: connectivity in the water
.
Diabetes Care
.
2023
;
46
(
8
):
e148
50
.
49.
Turner
LV
,
Sherr
J
,
Zaharieva
DP
,
Baran
JD
,
Bode
BW
,
Brown
SA
, et al
.
326-OR: activity feature attenuates drop in glucose for individuals with type 1 diabetes using Omnipod 5 automated insulin delivery system
.
Diabetes
.
2024
;
73
(
Suppl ment_1
):
326
OR
. (Abstract).
50.
Tagougui
S
,
Taleb
N
,
Legault
L
,
Suppère
C
,
Messier
V
,
Boukabous
I
, et al
.
A single-blind, randomised, crossover study to reduce hypoglycaemia risk during postprandial exercise with closed-loop insulin delivery in adults with type 1 diabetes: announced (with or without bolus reduction) vs unannounced exercise strategies
.
Diabetologia
.
2020
;
63
(
11
):
2282
91
.
51.
McCarthy
OM
,
Christensen
MB
,
Kristensen
KB
,
Schmidt
S
,
Ranjan
AG
,
Bain
SC
, et al
.
Glycemia around exercise in adults with type 1 diabetes using automated and nonautomated insulin delivery pumps: a switch pilot trial
.
Diabetes Technol Ther
.
2023
;
25
(
4
):
287
92
.
52.
Morrison
D
,
Paldus
B
,
Zaharieva
DP
,
Lee
MH
,
Vogrin
S
,
Jenkins
AJ
, et al
.
Late afternoon vigorous exercise increases postmeal but not overnight hypoglycemia in adults with type 1 diabetes managed with automated insulin delivery
.
Diabetes Technol Ther
.
2022
;
24
(
12
):
873
80
.
53.
Jayawardene
DC
,
McAuley
SA
,
Horsburgh
JC
,
Gerche
AL
,
Jenkins
AJ
,
Ward
GM
, et al
.
Closed-loop insulin delivery for adults with type 1 diabetes undertaking high-intensity interval exercise versus moderate-intensity exercise: a randomized, crossover study
.
Diabetes Technol Ther
.
2017
;
19
(
6
):
340
8
.
54.
Franc
S
,
Daoudi
A
,
Pochat
A
,
Petit
MH
,
Randazzo
C
,
Petit
C
, et al
.
Insulin-based strategies to prevent hypoglycaemia during and after exercise in adult patients with type 1 diabetes on pump therapy: the DIABRASPORT randomized study
.
Diabetes Obes Metab
.
2015
;
17
(
12
):
1150
7
.
55.
Reddy
R
,
Wittenberg
A
,
Castle
JR
,
El Youssef
J
,
Winters-Stone
K
,
Gillingham
M
, et al
.
Effect of aerobic and resistance exercise on glycemic control in adults with type 1 diabetes
.
Can J Diabetes
.
2019
;
43
(
6
):
406
14.e1
.
56.
Riddell
MC
,
Peters
AL
.
Exercise in adults with type 1 diabetes mellitus
.
Nat Rev Endocrinol
.
2023
;
19
(
2
):
98
111
.
57.
Chico
A
,
Moreno-Fernández
J
,
Fernández-García
D
,
Solá
E
.
The hybrid closed-loop system Tandem t:slim X2TM with control-IQ technology: expert recommendations for better management and optimization
.
Diabetes Ther
.
2024
;
15
(
1
):
281
95
.
58.
Adolfsson
P
,
Hanas
R
,
Zaharieva
DP
,
Dovc
K
,
Jendle
J
.
Automated insulin delivery systems in pediatric type 1 diabetes: a narrative review
.
J Diabetes Sci Technol
.
2024
;
18
(
6
):
1324
33
.
59.
Schoelwer
MJ
,
Robic
JL
,
Gautier
T
,
Fabris
C
,
Carr
K
,
Clancy-Oliveri
M
, et al
.
Safety and efficacy of initializing the control-IQ artificial pancreas system based on total daily insulin in adolescents with type 1 diabetes
.
Diabetes Technol Ther
.
2020
;
22
(
8
):
594
601
.
60.
Breton
MD
,
Cherñavvsky
DR
,
Forlenza
GP
,
DeBoer
MD
,
Robic
J
,
Wadwa
RP
, et al
.
Closed-loop control during intense prolonged outdoor exercise in adolescents with type 1 diabetes: the artificial pancreas ski study
.
Diabetes Care
.
2017
;
40
(
12
):
1644
50
.
61.
Ekhlaspour
L
,
Forlenza
GP
,
Chernavvsky
D
,
Maahs
DM
,
Wadwa
RP
,
Deboer
MD
, et al
.
Closed loop control in adolescents and children during winter sports: use of the Tandem Control-IQ AP system
.
Pediatr Diabetes
.
2019
;
20
(
6
):
759
68
.
62.
Mameli
C
,
Rigamonti
A
,
Felappi
B
,
Nicolucci
A
,
Lucisano
G
,
Baresi
S
, et al
.
Performance of Tandem control IQ during outdoor physical activity in children and adolescents with type 1 diabetes
.
Diabetes Technol Ther
.
2024
;
26
(
2
):
112
8
.
63.
Bergford
S
,
Riddell
MC
,
Gal
RL
,
Patton
SR
,
Clements
MA
,
Sherr
JL
, et al
.
Predicting hypoglycemia and hyperglycemia risk during and after activity for adolescents with type 1 diabetes
.
Diabetes Technol Ther
.
2024
;
26
(
10
):
728
38
.
64.
Moser
O
,
Tschakert
G
,
Mueller
A
,
Groeschl
W
,
Pieber
TR
,
Obermayer-Pietsch
B
, et al
.
Effects of high-intensity interval exercise versus moderate continuous exercise on glucose homeostasis and hormone response in patients with type 1 diabetes mellitus using novel ultra-long-acting insulin
.
PLoS One
.
2015
;
10
(
8
):
e0136489
.
65.
Riddell
MC
,
Zaharieva
DP
,
Tansey
M
,
Tsalikian
E
,
Admon
G
,
Li
Z
, et al
.
Individual glucose responses to prolonged moderate intensity aerobic exercise in adolescents with type 1 diabetes: the higher they start, the harder they fall
.
Pediatr Diabetes
.
2019
;
20
(
1
):
99
106
.
66.
Roy-Fleming
A
,
Taleb
N
,
Messier
V
,
Suppère
C
,
Cameli
C
,
Elbekri
S
, et al
.
Timing of insulin basal rate reduction to reduce hypoglycemia during late post-prandial exercise in adults with type 1 diabetes using insulin pump therapy: a randomized crossover trial
.
Diabetes Metab
.
2019
;
45
(
3
):
294
300
.
67.
Riddell
MC
,
Li
Z
,
Gal
RL
,
Calhoun
P
,
Jacobs
PG
,
Clements
MA
, et al
.
Examining the acute glycemic effects of different types of structured exercise sessions in type 1 diabetes in a real-world setting: the type 1 diabetes and exercise initiative (T1DEXI)
.
Diabetes Care
.
2023
;
46
(
4
):
704
13
.
68.
Bergford
S
,
Riddell
MC
,
Jacobs
PG
,
Li
Z
,
Gal
RL
,
Clements
MA
, et al
.
The type 1 diabetes and EXercise initiative: predicting hypoglycemia risk during exercise for participants with type 1 diabetes using repeated measures random forest
.
Diabetes Technol Ther
.
2023
;
25
(
9
):
602
11
.
69.
Li
Z
,
Calhoun
P
,
Rickels
MR
,
Gal
RL
,
Beck
RW
,
Jacobs
PG
, et al
.
Factors affecting reproducibility of change in glucose during exercise: results from the type 1 diabetes and EXercise initiative
.
J Diabetes Sci Technol
.
2024
:
19322968241234687
.
70.
Riddell
MC
,
Scott
SN
,
Fournier
PA
,
Colberg
SR
,
Gallen
IW
,
Moser
O
, et al
.
The competitive athlete with type 1 diabetes
.
Diabetologia
.
2020
;
63
(
8
):
1475
90
.