Background/Purpose: The water jet technique dissects tissue while sparing cord-like structures such as blood vessels. The mechanism of such tissue-selective dissection has been unknown. The novel piezo actuator-driven pulsed water jet (ADPJ) system can achieve dissection with remarkably reduced water consumption compared to the conventional water jet; however, the system's characteristics and dissection capabilities on any organ have not been clarified. The purposes of this study were to characterize the physical properties of the novel ADPJ system, evaluate the dissection ability in swine organs, and reveal the mechanism of tissue-selective dissection. Methods: The pulsed water jet system comprised a pump chamber driven by a piezo actuator, a stainless steel tube, and a nozzle. The peak pressure of the pulsed water jet was measured through a sensing hole using a pressure sensor. The pulsed water jet technique was applied on swine liver in order to dissect tissue on a moving table using one-way linear ejection at a constant speed. The dissection depth was measured with light microscopy and evaluated histologically. The physical properties of swine liver were evaluated by breaking strength tests using tabletop universal testing instruments. The liver parenchyma was also cut with three currently available surgical devices to compare the histological findings. Results: The peak pressure of the pulsed water jet positively correlated with the input voltage (R2 = 0.9982, p < 0.0001), and this was reflected in the dissection depth. The dissection depth negatively correlated with the breaking strength of the liver parenchyma (R2 = 0.6694, p < 0.0001). The average breaking strengths of the liver parenchyma, hepatic veins, and Glisson's sheaths were 1.41 ± 0.45, 8.66 ± 1.70, and 29.6 ± 11.0 MPa, respectively. The breaking strength of the liver parenchyma was significantly lower than that of the hepatic veins and Glisson's sheaths. Histological staining confirmed that the liver parenchyma was selectively dissected, preserving the hepatic veins and Glisson's sheaths in contrast to what is commonly observed with electrocautery or ultrasonic instruments. Conclusions: The dissection depth of liver tissue is well controlled by input voltage and is influenced by the moving velocity and the physical properties of the organ. We showed that the device can be used to assure liver resection with tissue selectivity due to tissue-specific physical properties. Although this study uses an excised organ, further in vivo studies are necessary. The present work demonstrates that this device may function as an alternative tool for surgery due to its good controllability of the dissection depth and ability of tissue selectivity.

1.
Ronnie TP, Sheung TF, Chung ML, Chi LL, Chi ML, Wai KY, Chun Y, John W: Improving perioperative outcome expands the role of hepatectomy in management of benign and malignant hepatobiliary diseases: analysis of 1,222 consecutive patients from a prospective database. Ann Surg 2004;240:698-710.
2.
Sitzmann JV, Greene PS: Perioperative predictors of morbidity following hepatic resection for neoplasm: a multivariate analysis of a single surgeon experience with 105 patients. Ann Surg 1994;219:13-17.
3.
Yamamoto J, Kosuge T, Takayama T, Shimada K, Yamasaki S, Ozaki H, Yamaguchi N, Mizuno S, Makuuchi M: Perioperative blood transfusion promotes recurrence of hepatocellular carcinoma after hepatectomy. Surgery 1994;115:303-309.
4.
David AK, Jennifer S, Leah BP, Mithat G, William RJ, Ronald PD, Scott T, David W, Leslie HB, Yuman F: Influence of transfusions on perioperative and long-term outcome in patients following hepatic resection for colorectal metastases. Ann Surg 2003;237:860-870.
5.
Rau HG, Meyer G, Jauch KW, Cohnert TU, Buttler E, Schildberg FW: Liver resection with the water jet: conventional and laparoscopic surgery. Chirurg 1996;67:546-551.
6.
Rau HG, Wichmann MW, Schinkel S, Buttler E, Pickelmann S, Schauer R, Schildberg FW: Surgical techniques in hepatic resections: ultrasonic aspirator versus Jet-Cutter. A prospective randomized clinical trial. Zentralbl Chir 2001;126:586-590.
7.
Takayama T, Makuuchi M, Kubota K, Harihara Y, Hui AM, Sano K, Ijichi M, Hasegawa K: Randomized comparison of ultrasonic vs clamp transection of the liver. Arch Surg 2001;136:922-928.
8.
Arita J, Hasegawa K, Kokudo N, Sano K, Sugawara Y, Makuuchi M: Randomized clinical trial of the effect of a saline-linked radiofrequency coagulator on blood loss during hepatic resection. Br J Surg 2005;92:954-959.
9.
Koo BN, Kil HK, Choi JS, Kim JY, Chun DH, Hong YW: Hepatic resection by the Cavitron ultrasonic surgical aspirator increases the incidence and severity of venous air embolism. Anesth Analg 2005;101:966-970.
10.
Lesurtel M, Selzner M, Petrowsky H, McCormack L, Clavien PA: How should transection of the liver be performed? A prospective randomized study in 100 consecutive patients: comparing four different transection strategies. Ann Surg 2005;242:814-823.
11.
Lupo L, Gallerani A, Panzera P, Tandoi F, Di-Palma G, Memeo V: Randomized clinical trial of radiofrequency-assisted vs clamp-crushing liver resection. Br J Surg 2007;94:287-291.
12.
Oertel J, Gaab MR, Warzok R, Piek J: Waterjet dissection in the brain: review of the experimental and clinical data with special reference to meningioma surgery. Neurosurg Rev 2003;26:168-174.
13.
Papachristou DN, Barters R: Resection of the liver with a water jet. Br J Surg 1982;69:93-94.
14.
Oertel J, Gaab MR, Knapp A, Essig H, Warzok R, Piek J: Water jet dissection in neurosurgery: experimental results in the porcine cadaveric brain. Neurosurgery 2003;52:153-159.
15.
Une Y, Uchino J, Horie T, Sato Y, Ogasawara K, Kakita A, Sano F: Liver resection using a water jet. Cancer Chemother Pharmacol 1989;23:S74-S77.
16.
Seto T, Yamamoto H, Takayama K, Nakagawa A, Tominaga T: Characteristics of an actuator-driven pulsed water jet generator to dissecting soft tissue. Rev Sci Instrum 2011;82:055105.
17.
Okitani A, Oneda Y, Kubo T, Ishii T, Suzuki R, Awata T, Sunada Y, Yamashita Y, Migita K, Matsuishi M, Hatae K: Softening mechanism of vacuum-cooked squid muscle. J Jpn Soc Food Sci Technol 2008;55:170-176.
18.
Seki S, Iwamoto H: Disruptive forces for swine heart, liver, and spleen: their breaking stresses. J Trauma 1998;45:1079-1083.
19.
Terzis AJ, Nowak G, Rentzsch O, Arnold H, Diebold J, Baretton G: A new system for cutting brain tissue preserving vessels: water jet cutting. Br J Neurosurg 1989;3:361-366.
20.
Kaehler GF, Sold MG, Fischer K, Post S, Enderle M: Selective fluid cushion in the submucosal layer by water jet: advantage for endoscopic mucosal resection. Eur Surg Res 2007;39:93-97.
21.
Yahagi N, Neuhaus H, Schumacher B, Neugebauer A, Kaehler GF, Schenk M, Fischer K, Fujishiro M, Enderle MD: Comparison of standard endoscopic submucosal dissection (ESD) versus an optimized ESD technique for the colon: an animal study. Endoscopy 2009;41:340-345.
22.
Aroussi AA, Sami IM, Leguerrier A, Verhoye JP: The blower: a useful tool to complete thrombectomy of the mechanical prosthetic value. Ann Thorac Surg 2006;81:1911-1912.
23.
Lipshitz I, Bass R, Loewenstein A: Cutting the cornea with a waterjet keratome. J Refract Surg 1996;12:184-186.
24.
Vollmer CM, Dixon E, Sahajpal A, Cattral MS, Grant DA, Gallinger S, Taylor BR, Greig PD: Water-jet dissection for parenchymal division during hepatectomy. HPB (Oxford) 2006;8:377-385.
25.
Schurr MO, Wehrmann M, Kunert W, Melzer A, Lirici MM, Trapp R, Kanehira E, Buess G: Histologic effects of different technologies for dissection in endoscopic surgery: Nd:YAG laser, high frequency and water-jet. Endosc Surg Allied Technol 1994;2:195-201.
26.
Oertel J, Gaab MR, Pillich DT, Schroeder HW, Warzok R, Piek J: Comparison of waterjet dissection and ultrasonic aspiration: an in vivo study in the rabbit brain. J Neurosurg 2004;100:498-504.
27.
Sato C, Nakano T, Nakagawa A, Yamada M, Yamamoto H, Kamei T, Miyata G, Sato A, Fujishima F, Nakai M, Niinomi M, Takayama K, Tominaga T, Satomi S: Experimental application of pulsed laser-induced water jet for endoscopic submucosal dissection: mechanical investigation and preliminary experiment in swine. Dig Endosc 2012;25:255-263.
28.
Ohki T, Nakagawa A, Hirano T, Hashimoto T, Menezes V, Jokura H, Uenohara H, Sato Y, Saito T, Shirane R, Tominaga T, Takayama K: Experimental application of pulsed Ho:YAG laser-induced liquid jet as a novel rigid neuroendoscopic dissection device. Lasers Surg Med 2004;34:227-234.
29.
Tschan CA, Tschan K, Krauss JK, Oertel J: First experimental results with a new waterjet dissector: Erbejet 2. Acta Neurochir 2009;151:1473-1482.
30.
Rau HG, Duessel AP, Wurzbacher S: The use of water-jet dissection in open and laparoscopic liver resection. HPB 2008;10:275-280.
31.
Foldyna J, Sitek L, Svehla B, Svehla S: Utilization of ultrasound to enhance high-speed water jet effects. Ultrason Sonochem 2004;11:131-137.
32.
Oertel J, Gen M, Krauss JK, Zumkeller M, Gaab MR: The use of water jet dissection in endoscopic neurosurgery. Technical note. J Neurosurg 2006;105:928-931.
33.
Shi H, Jiang SJ, Li B, Fu DK, Xin P, Wang YG: Natural orifice transluminal endoscopic wedge hepatic resection with a water-jet hybrid knife in a non-survival porcine model. World J Gastroenterol 2011;17:926-931.
34.
Nakagawa A, Hirano T, Komatsu M, Sato M, Uenohara H, Ohyama H, Kusaka Y, Shirane R, Takayama K, Yoshimoto T: Holmium: YAG laser-induced liquid jet knife: possible novel method for dissection. Lasers Surg Med 2002;31:129-135.
35.
Kang MO, Won-Beom C, Seong HP, Eunsil Y, Sung GL, Young-Suk L, Han CL, Young-Hwa C, Yung SL, Dong-Jin S: Diagnosis of hepatic steatosis and fibrosis by transient elastography in asymptomatic healthy individuals: a prospective study of living related potential liver donors. J Gastroenterol 2007;42:382-388.
36.
David AK, Yuman F, Arief S, Mithat G, Peter JA, David SK, Ronald PD, Michael DA, Leslie HB, William RJ: Impact of steatosis on perioperative outcome following hepatic resection. J Gastrointest Surg 2003;7:1034-1044.
37.
Jean-Nicolas V, Timothy MP, Dario R, Tsung-Teh W, Daria Z, Paulo MH, Henry QX, Cathy E, Gregory YL, Mari MK, Mauro R, Andrea M, Lorenzo C, Steven AC, Eddie KA: Chemotherapy regimen predicts steatohepatitis and an increase in 90-day mortality after surgery for hepatic colorectal metastases. J Clin Oncol 2006;24:2065-2072.
38.
Rubbia-Brandt L, Mentha G, Terris B: Sinusoidal obstruction syndrome is a major feature of hepatic lesions associated with oxaliplatin neoadjuvant chemotherapy for liver colorectal metastases. J Am Coll Surg 2006;202:199-200.
Copyright / Drug Dosage / Disclaimer
Copyright: All rights reserved. No part of this publication may be translated into other languages, reproduced or utilized in any form or by any means, electronic or mechanical, including photocopying, recording, microcopying, or by any information storage and retrieval system, without permission in writing from the publisher.
Drug Dosage: The authors and the publisher have exerted every effort to ensure that drug selection and dosage set forth in this text are in accord with current recommendations and practice at the time of publication. However, in view of ongoing research, changes in government regulations, and the constant flow of information relating to drug therapy and drug reactions, the reader is urged to check the package insert for each drug for any changes in indications and dosage and for added warnings and precautions. This is particularly important when the recommended agent is a new and/or infrequently employed drug.
Disclaimer: The statements, opinions and data contained in this publication are solely those of the individual authors and contributors and not of the publishers and the editor(s). The appearance of advertisements or/and product references in the publication is not a warranty, endorsement, or approval of the products or services advertised or of their effectiveness, quality or safety. The publisher and the editor(s) disclaim responsibility for any injury to persons or property resulting from any ideas, methods, instructions or products referred to in the content or advertisements.
You do not currently have access to this content.