Introduction: Fluid overload and congestion are common features in patients with heart failure and are associated with negative clinical outcomes. Therapies for these conditions are diuretic-centered but frequently fail to achieve patient-adequate hydration status, prompting the use of extracorporeal ultrafiltration. Artificial Diuresis 1 (AD1) is a miniaturized, portable, and wearable system designed to deliver isolated ultrafiltration with the finest degree of simplicity and practicality. Methods/Design: Single-center, crossover, randomized, open-label pilot study to investigate the safety and the efficacy (concerning ultrafiltration accuracy) of extracorporeal ultrafiltration with the device AD1 in comparison to isolated ultrafiltration with a traditional machine (PrisMaX). Patients with chronic kidney disease stage 5D (on hemodialysis) or intensive care patients presenting acute kidney injury stage 3D (requiring hemodialysis) will carry out a single session of isolated ultrafiltration with each of the machines. The safety primary outcomes will be the occurrence of adverse events. The efficacy primary outcome will be the accuracy of ultrafiltration rate (delivered/prescribed) on each of the devices. Conclusion: AD1 is a novel miniaturized device for extracorporeal ultrafiltration. This study will be the first-in-human use of AD1 in patients with fluid overload.

The prevalence of heart failure in Italy is 1.25% and this percentage is equally distributed in both genders. A rampant increase in prevalence is documented in the elderly population. Specifically, among persons aged 60 to 65 years, the prevalence is 1.47%, scaling up to 13% in nonagenarians. Moreover, the incidence of the syndrome is 1.99 per 1,000 person-years [1]. In other middle- or high-income countries, these numbers are similar [2]. Therefore, heart failure is a condition associated with major public health concerns, responsible for relevant financial and social burdens [3].

Acute exacerbations of chronic heart failure are frequent and are a leading cause of overall hospitalization and emergency room visits. Four out of 5 patients who seek hospital care because of acute heart failure present signs and symptoms of congestion [4, 5]. Staggering, the all-cause 1-year mortality for patients with acute heart failure is 23.6%. In contrast, for individuals with chronic heart failure, all-cause 1-year mortality is 6.4%, demonstrating the negative impact of acute exacerbations.

Current guidelines supported by robust evidence from clinical trials recommend a diuretic-centered approach for the management of this condition [6‒15]. Nonetheless, frequently, diuretic therapies fail, and mechanical extracorporeal ultrafiltration is required in the treatment of congestion [6, 16, 17]. The safety and efficacy of ultrafiltration have been demonstrated in numerous clinical trials [17].

Present-day machines utilized for extracorporeal ultrafiltration are either capable of performing other extracorporeal blood purification therapies (i.e., continuous renal replacement therapy [CRRT]) or dedicated to isolated ultrafiltration [18, 19]. Despite the evolution of these machines, burdensome personnel training is needed. Moreover, the execution of the therapy is not feasible in hospital wards, preventing its democratization and wide adoption. As a way to overcome these barriers, our group initiated a project of a portable and wearable miniaturized system for isolated ultrafiltration [20]. The innovative project and the device itself were named “Artificial Diuresis 1” (AD1) (shown in Fig. 1). We carried out in silico simulations, in vitro experiments, and in vivo animal experiments. After successfully passing through these necessary phases for hardware development, a pilot and exploratory study in humans could be planned (shown in Fig. 2). Herein, we describe the protocol for the first-in-human use of AD1.

Fig. 1.

Artificial Diuresis 1 (AD1). Dimensions of the novel portable, wearable, and miniaturized device for isolated ultrafiltration.

Fig. 1.

Artificial Diuresis 1 (AD1). Dimensions of the novel portable, wearable, and miniaturized device for isolated ultrafiltration.

Close modal
Fig. 2.

Phases of the development of a miniaturized portable device for extracorporeal ultrafiltration. a In silico 3D project drawing and performance simulations. b In vitro testing with the GALILEO machine. Ultrafiltration and circuit pressures were determined in different conditions such as access type and diameter, blood flow, and height of the collection bag (hydrostatic column). c In vivo experimental pig model. Three pigs carried out a single session each. Vital signs remained stable during the therapy; ultrafiltration was precise without technical problems. d In vivo, clinical safety and efficacy pilot trial.

Fig. 2.

Phases of the development of a miniaturized portable device for extracorporeal ultrafiltration. a In silico 3D project drawing and performance simulations. b In vitro testing with the GALILEO machine. Ultrafiltration and circuit pressures were determined in different conditions such as access type and diameter, blood flow, and height of the collection bag (hydrostatic column). c In vivo experimental pig model. Three pigs carried out a single session each. Vital signs remained stable during the therapy; ultrafiltration was precise without technical problems. d In vivo, clinical safety and efficacy pilot trial.

Close modal

Trial Design

This is a single-center, crossover, randomized, open-label pilot study to assess the safety and efficacy of AD1 (Medica S.p.A. Medolla, Emilia-Romagna, Italy) in comparison with a standard CRRT machine (PrisMaX, Baxter Healthcare Corporation, Deerfield, IL, US), capable of performing continuous isolated ultrafiltration. The main goal is to analyze safety events during the treatment of isolated ultrafiltration in patients with fluid overload. This trial will not have a pre-defined run-in or washout phases. Each patient will perform only one treatment with either machine. Recruitment will utilize the convenience sampling strategy [21‒25]. The trial is registered as CESC Provincia di Vicenza_6.9.22(n.54/22)_1575_29.9.22.

Study Setting

Patients will be recruited from the Department of Nephrology, Dialysis, and Kidney Transplant and from the Department of Critical Care, both at San Bortolo Hospital, Vicenza, Veneto, Italy. Written consent will be obtained from all patients enrolled in the trial.

Eligibility Criteria

Inclusion Criteria

  • 1.

    Both genders.

  • 2.

    Established chronic kidney disease G5D patients on maintenance hemodialysis for at least 12 weeks, carrying out in-center sessions, presenting at least 2.5 kg over a pre-defined adequate body weight (dry weight).

  • 3.

    Established chronic kidney disease G5D patients on maintenance hemodialysis for at least 12 weeks, hospitalized in the intensive care unit, and carrying out sessions in this unit, presenting at least 2.5 kg over a pre-defined adequate body weight (dry weight).

  • 4.

    Intensive care unit patients presenting acute kidney injury stage 3D (requiring hemodialysis), in whom fluid accumulation is detected, and extracorporeal ultrafiltration is indicated according to the attending physician evaluation. In our unit, fluid accumulation is assessed by physical examination, point-of-care ultrasonography, venous excess Doppler ultrasound, other imaging such as chest radiography, chest computed tomography. Moreover, other variables evaluated for the assessment of congestion are central venous pressure, intra-abdominal pressure, and cumulative positive fluid balance, above 5–10% of presumed dry weight) 26‒33.

  • 5.

    Aged over 18 years.

Exclusion Criteria

  • 1.

    Planned renal transplant within the study intervention period.

  • 2.

    Planned conversion to peritoneal dialysis or transfer to another center.

  • 3.

    Pregnancy or breastfeeding.

  • 4.

    Indication for hemodialysis, hemodiafiltration, or hemoadsorption according to the attending physician.

  • 5.

    Patients with current infection by human immunodeficiency virus, hepatitis B, hepatitis C, and SARS-CoV-2.

  • 6.

    Impossibility of the patient or the next of kin to provide informed consent.

Discontinuation

  • 1.

    At the discretion of the treating physician.

  • 2.

    The decision by the patient or family to withdraw at any moment.

  • 3.

    Partial recovery of kidney function.

Missing Data

Missing data are expected at various time points during the trial. No data imputation techniques will be applied to any of the evaluated variables [34].

Data Monitoring

Data integrity will be enforced by various mechanisms, such as valid values and range checks. Access to the study data will be restricted by personal logging into the system data. All paper forms or electronic devices will be kept in locked cabinets. A complete backup will be regularly provided.

The protocol was reviewed and approved by the Local Independent Ethics Committee (Comitato Etico per le Sperimentazioni Cliniche della Provincia di Vicenza 6.9.22(n.54/22)_1575_29.9.22) and submitted to Italian Ministry of Health for the final approval concerning use of devices still not CE-approved (Conformité Européenne). The study will be performed in accordance with the 2013 Fortaleza, Brazil, 7th Revision of the Declaration of Helsinki following good clinical practice standards (D.M. Sanità del July 15, 1997 e s.m.i.) [35]. Personal data will be treated to the General Data Protection Regulation (EU) 2016/679 and “Decreto Legislativo June 30, 2003 n.196 e ss.mm.ii. (Codice in materia dei dati personali).” The trial will be registered at ClinicalTrials.gov website.

Interventions/Study Procedures

For patients carrying out in-center regimens, isolated ultrafiltration sessions will have a duration from 4 to 6 h and the total fluid removal will range from 500 to 1,500 mL, according to the clinical need. For intensive care unit patients, sessions’ duration will vary from 6 to 12 h and total ultrafiltration volume will range from 1,000–1,200 mL, according to the clinical need. Systemic anticoagulation might be provided in the absence of contraindications. For patients already receiving anticoagulants, no anticoagulation will be used. After randomization, patients will be assigned to initiate the first treatment with one of the machines followed by a subsequent therapy with the alternative machine (shown in Fig. 3).

Fig. 3.

Graphical abstract summarizing the protocol. The study question is at the top. Each square displays a section of the protocol. The upper box in the left column, population, shows the planned sample size of 15 patients, including in-center CKD 5D and ICU patients presenting AKI 3D. Downward in the same column, the second box, design, demonstrates a randomized, crossover, and open-label study. In the same column, the last box, locations/settings, shows that the study will occur in San Bortolo Hospital, in which the IRRIV is also located. The upper two boxes of the middle column indicate recruitment strategy and devices (AD1 and PrisMaX) that will be utilized. In the middle box of the middle column, primary safety outcomes (adverse events) and primary efficacy outcome (ultrafiltration accuracy) are depicted. In the bottom box of the middle column, the procedures are demonstrated. In brief, the ultrafiltration range and the possible duration of the therapies are shown. This box also illustrates that the first session can be carried out with AD1 or with PrisMaX, followed by the alternative hardware in the subsequent session. Finally, the right column, the expected results are demonstrated. AD1, Artificial Diuresis 1; AKI 3D, acute kidney injury stage 3D (requiring dialysis); CKD 5D, chronic kidney disease stage 5D (requiring dialysis); IRRIV, International Renal Research Institute of Vicenza.

Fig. 3.

Graphical abstract summarizing the protocol. The study question is at the top. Each square displays a section of the protocol. The upper box in the left column, population, shows the planned sample size of 15 patients, including in-center CKD 5D and ICU patients presenting AKI 3D. Downward in the same column, the second box, design, demonstrates a randomized, crossover, and open-label study. In the same column, the last box, locations/settings, shows that the study will occur in San Bortolo Hospital, in which the IRRIV is also located. The upper two boxes of the middle column indicate recruitment strategy and devices (AD1 and PrisMaX) that will be utilized. In the middle box of the middle column, primary safety outcomes (adverse events) and primary efficacy outcome (ultrafiltration accuracy) are depicted. In the bottom box of the middle column, the procedures are demonstrated. In brief, the ultrafiltration range and the possible duration of the therapies are shown. This box also illustrates that the first session can be carried out with AD1 or with PrisMaX, followed by the alternative hardware in the subsequent session. Finally, the right column, the expected results are demonstrated. AD1, Artificial Diuresis 1; AKI 3D, acute kidney injury stage 3D (requiring dialysis); CKD 5D, chronic kidney disease stage 5D (requiring dialysis); IRRIV, International Renal Research Institute of Vicenza.

Close modal

In case of intraprocedural hypotension, ultrafiltration can be momentarily discontinued by clamping the ultrafiltrate line (see Fig. 4). In this situation, the blood pump is not stopped, and the blood still runs through the circuit, minimizing the risk of filter coagulation in case of blood stagnation which occurs when the blood pump is stopped. Moreover, a 1,000-mL saline bag is connected to the venous line of the extracorporeal circuit via a 3-way stopcock if fluid bolus infusion is required to revert an episode of refractory hypotension. The amount of fluid delivered will vary at discretion of the attending physician.

Fig. 4.

Components of the Artificial Diuresis 1 (AD1) hardware. a External case with a membrane keyboard placed on a polycarbonate cover. b Disposable circuit displaying inlet and outlet lines, hollow fiber filter, ultrafiltrate line (effluent line), and the collection bag. c Opened device without the circuit cassette. The main components are the peristaltic blood pump, blood leak detector, flow sensors, air detector, inlet and outlet line pressure detector. d Opened device with the disposable circuit in place.

Fig. 4.

Components of the Artificial Diuresis 1 (AD1) hardware. a External case with a membrane keyboard placed on a polycarbonate cover. b Disposable circuit displaying inlet and outlet lines, hollow fiber filter, ultrafiltrate line (effluent line), and the collection bag. c Opened device without the circuit cassette. The main components are the peristaltic blood pump, blood leak detector, flow sensors, air detector, inlet and outlet line pressure detector. d Opened device with the disposable circuit in place.

Close modal

The venous access for patients with acute kidney injury stage 3D will be a temporary venous access with a diameter >11 Fr. For the maintenance hemodialysis patients, the venous access will be a tunneled catheter or a temporary catheter, both with a diameter >11 Fr. Peripheral accesses, arteriovenous fistulas, or arteriovenous grafts will not be utilized. In a previous in vitro study carried out by our group, AD1 was tested in a closed-loop circuit primed with whole blood. The catheters tested were 7 Fr, 10 Fr, and 12 Fr. The experiments with each of the catheters were successful and accurate ultrafiltration was achieved [36].

Primary Objective

Primary safety outcomes will include the incidence of clinical events such as intraprocedural hypotension, bleeding events, hemolysis, hypothermia, electrolyte imbalance, anaphylactoid reactions, and circuit clotting. Primary efficacy outcome will be the percentual variation between prescribed versus delivered total ultrafiltration volume.

Secondary Objective

Secondary safety outcomes will include the incidence of clinical events such as cramps, nausea, vomiting, headache, fever, chills, chest pain, and pruritus and also technical problems such as filter hollow fiber rupture and identification of air in the circuit. Secondary efficacy outcome will be the measurement of user-friendliness and technical complications.

Primary Outcomes (Primary Objective Measurement)

Primary safety outcomes will assess intraprocedural hypotension employing invasive or noninvasive arterial pressure volume, air embolism through clinical evaluation and peripheral oxygen saturation monitoring, hemolysis by a change in the effluent color, hypothermia via intermittent or continuous temperature monitoring, electrolyte imbalances by biochemical blood sampling, circuit clotting by visual inspection, and anaphylactoid reactions by physicians’ judgment. Primary efficacy outcomes will be measured by collecting the effluent volume of the AD1 or PrisMaX.

Secondary Outcomes (Secondary Objective Measurement)

Secondary safety outcomes will be presented as percentual of the occurrence of events during all the therapies. Secondary efficacy outcomes will be assessed by the completion of a user satisfaction questionnaire and the measurement of technical issues that required nurses’ intervention (troubleshooting) summarize in Table 1.

Table 1.

User questionnaire

Items Score:1: unaccomplished despite using operator's manual2: accomplished under operator's manual guidance3: accomplished without operator's manual support 
Getting started a) Check battery charge; b) battery installation; c) power on; d) home screen 
Connecting set a) Set visual inspection (filter, lines, bags); b) load disposables; c) blood line self-loaded to blood pump 
Priming and connecting the patient a) Blood line connections; b) running the treatment; c) blood flow adjustment; d) battery replacement; e) treatment start; f) stopping the treatment; g) treatment conclusion 
Blood returning and disconnecting the patient a) Arterial line disconnection; b) blood return; c) venous line disconnection 
Disconnecting and discarding the set a) Remove disposable set; b) return to home screen; c) power off; c) remove battery; d) recharge battery 
Items Score:1: unaccomplished despite using operator's manual2: accomplished under operator's manual guidance3: accomplished without operator's manual support 
Getting started a) Check battery charge; b) battery installation; c) power on; d) home screen 
Connecting set a) Set visual inspection (filter, lines, bags); b) load disposables; c) blood line self-loaded to blood pump 
Priming and connecting the patient a) Blood line connections; b) running the treatment; c) blood flow adjustment; d) battery replacement; e) treatment start; f) stopping the treatment; g) treatment conclusion 
Blood returning and disconnecting the patient a) Arterial line disconnection; b) blood return; c) venous line disconnection 
Disconnecting and discarding the set a) Remove disposable set; b) return to home screen; c) power off; c) remove battery; d) recharge battery 

Sample Size Calculation

The sample size was calculated based on the primary outcome. We did not find evidence from medical literature concerning trials about ultrafiltration with a similar design. Based on previous studies about human device testing carried over along more than 40 years [18] in our hospital and expert input from the manufacturer, we estimated that 15 patients would be required. This sample size was considered adequate to increase the probability of occurrence and detection of the safety outcomes already specified. Moreover, concerning the primary efficacy endpoints, we considered that 15 treatments on each machine could demonstrate the accuracy of AD1 with a variation of less than 10% between delivered versus prescribed ultrafiltration total ultrafiltration volume.

Randomization

Fifteen opaque sealed envelopes will be provided; in eight of them, the designated treatment will be ultrafiltration with AD1 and in the remaining seven, ultrafiltration with the CRRT machine.

Allocation Concealment

A box containing the envelopes will be secured in an office staff room. None of the staff members will be aware of the trial’s progress. Contact will be made by phone call. The opened envelopes will progressively be discarded.

Statistical Analysis

Continuous variables will be described according to central tendency measures (mean and median) and dispersion measures (standard variation, confidence interval, interquartile range). Data will be presented as means ± standard deviation for continuous variables with normal distribution and median and interquartile range for non-normally distributed variables. Normality will be assessed a priori by visual inspection of histograms and by performing statistical tests such as the Kolmogorov-Smirnov or Shapiro-Wilk. Categorical variables will be presented as percentages and absolute numbers.

The comparison of independent categorical variables and dependent continuous variables with normal distribution will apply a t test. If the distribution of the dependent continuous variable is non-normal, the Mann-Whitney Wilcoxon test will be used. In the comparison between independent categorical variables and dependent categorical variables, the χ2 or Fisher’s exact test will be applied. Data will be processed in IBM SPSS for Windows (IBM Corp., Armonk, NY, US).

In the proposed single-center, crossover, randomized, open-label pilot study, we plan to pragmatically assess the safety and efficacy of a new miniaturized ultrafiltration device in comparison with a traditional machine. The defined sample size will likely be reached because the eligibility criteria are broad and our hospital has a large intensive care unit and a large hemodialysis center. In both settings, condition of fluid overload is prevalent. The physicians and nurses have the needed technical expertise and other first-in-human uses of machines were carried out in our center. Thus, we expect that staff training will occur at a fast pace.

If safety and efficacy outcomes are achieved, researchers in the field of cardiorenal medicine and intensive care will certainly be encouraged to carry out randomized controlled trials. We believe this is a groundbreaking and innovative investigation, justifying the application of human and financial resources to this project. The long-term goal will be to assess the beneficial adjuvant effects of a miniaturized portable device associated with diuretics to improve the management of fluid overloaded patients in acute settings. In addition, we suppose that the exploration of extracorporeal ultrafiltration in the elective management of cardiorenal patients, in whom fluid accumulation is highly prevalent, is highly needed. Therefore, there is a strong rationale to investigate the efficacy of the device as an elective therapy.

The protocol was reviewed and approved by the Local Independent Ethics Committee (Comitato E. per le Sperimentazioni Cliniche della Provincia di Vicenza 6.9.22 (n.54/22) 1575_29.9.22). Informed signed consent will be obtained from all the participants. The informed consent form will be accessible online in the supplementary material section (for all online suppl. material, see https://doi.org/10.1159/000530943). The study will be performed in accordance with the 2013 Fortaleza, Brazil, 7th Revision of the Declaration of Helsinki following good clinical practice standards.

Alessandra Brendolan, Anna Lorenzin, Luca Sgarabotto, Valentina Corradi, Nicola Marchionna, and Monica Zanella do not have any potential conflicts of interest. Claudio Ronco has received research grants, funding for lectures, been consultant or advisory board member for Asahi, Astute, B. Braun, Baxter, bioMérieux, Bioporto, CytoSorbents, Estor, Fresenius Medical Care, General Electric (GE), Jafron, Medtronic, and Toray. Thiago Reis has received funding for lectures, been consultant or advisory board member for AstraZeneca, B. Braun, Baxter, bioMérieux, Boehringer Ingelheim, Contatti Medical (CytoSorbents), Eurofarma, Jafron, Lifepharma, Medtronic, and Nova Biomedical.

No funding was received.

Thiago Reis: methodology (equal); writing – original draft (lead); and writing – review and editing (lead). Luca Sgarabotto, Anna Lorenzin, Valentina Corradi, Nicola Marchionna: methodology (equal); writing – original draft (supporting); and writing – review and editing (equal). Alessandra Brendolan: methodology (lead); writing – original draft (lead); and writing – review and editing (equal). Monica Zanella: methodology (equal); writing – original draft (equal); and writing – review and editing (equal). Claudio Ronco: conceptualization (lead); funding acquisition (lead); investigation (equal); methodology (equal); project administration (lead); resources (lead); writing – original draft (equal); writing – review and editing (equal); and supervision. All authors were responsible for critical revision of the manuscript and approved the final version before submission.

Additional Information

Trial registration number: CESC Provincia di Vicenza_6.9.22(n.54/22)_1575_29.9.22.

Data are available on request due to privacy or other restrictions.

1.
Piccinni
C
,
Antonazzo
IC
,
Simonetti
M
,
Mennuni
MG
,
Parretti
D
,
Cricelli
C
.
The burden of chronic heart failure in primary care in Italy
.
High Blood Press Cardiovasc Prev
.
2017 Jun
24
2
171
8
.
2.
Khatibzadeh
S
,
Farzadfar
F
,
Oliver
J
,
Ezzati
M
,
Moran
A
.
Worldwide risk factors for heart failure: a systematic review and pooled analysis
.
Int J Cardiol
.
2013 Sep
168
2
1186
94
.
3.
Ziaeian
B
,
Fonarow
GC
.
Epidemiology and aetiology of heart failure
.
Nat Rev Cardiol
.
2016 Jun
13
6
368
78
.
4.
Chioncel
O
,
Mebazaa
A
,
Harjola
V-P
,
Coats
AJ
,
Piepoli
MF
,
Crespo-Leiro
MG
.
Clinical phenotypes and outcome of patients hospitalized for acute heart failure: the ESC Heart Failure Long-Term Registry
.
Eur J Heart Fail
.
2017 Oct
19
10
1242
54
.
5.
Koratala
A
,
Ronco
C
,
Kazory
A
.
Diagnosis of fluid overload: from conventional to contemporary concepts
.
Cardiorenal Med
.
2022
;
12
(
4
):
141
54
.
6.
McDonagh
TA
,
Metra
M
,
Adamo
M
,
Gardner
RS
,
Baumbach
A
,
Böhm
M
.
2021 ESC Guidelines for the diagnosis and treatment of acute and chronic heart failure
.
Eur Heart J
.
2021 Sep
42
36
3599
726
.
7.
Edwards
NC
,
Price
AM
,
Steeds
RP
,
Ferro
CJ
,
Townend
JN
Birmingham Cardio-Renal Group
.
Management of heart failure in patients with kidney disease: updates from the 2021 ESC guidelines
.
Nephrol Dial Transpl
.
2023 Jan
gfad011
.
8.
Trulls
JC
,
Morales-Rull
JL
,
Casado
J
,
Carrera-Izquierdo
M
,
Snchez-Marteles
M
,
Conde-Martel
A
.
Combining loop with thiazide diuretics for decompensated heart failure: the CLOROTIC trial
.
Eur Heart J
.
2023 Feb
44
5
411
21
.
9.
Mullens
W
,
Damman
K
,
Harjola
V-P
,
Mebazaa
A
,
Brunner-La Rocca
H-P
,
Martens
P
.
The use of diuretics in heart failure with congestion: a position statement from the Heart Failure Association of the European Society of Cardiology: diuretics in heart failure
.
Eur J Heart Fail
.
2019 Feb
21
2
137
55
.
10.
Mullens
W
,
Dauw
J
,
Martens
P
,
Verbrugge
FH
,
Nijst
P
,
Meekers
E
.
Acetazolamide in acute decompensated heart failure with volume overload
.
N Engl J Med
.
2022 Sep
387
13
1185
95
.
11.
Schulze
PC
,
Bogoviku
J
,
Westphal
J
,
Aftanski
P
,
Haertel
F
,
Grund
S
.
Effects of early empagliflozin initiation on Diuresis and kidney function in patients with acute decompensated heart failure (EMPAG-HF)
.
Circulation
.
2022 Jul
146
4
289
98
.
12.
Felker
GM
,
Lee
KL
,
Bull
DA
,
Redfield
MM
,
Stevenson
LW
,
Goldsmith
SR
.
Diuretic strategies in patients with acute decompensated heart failure
.
N Engl J Med
.
2011 Mar
364
9
797
805
.
13.
Palazzuoli
A
,
Ruocco
G
,
Vescovo
G
,
Valle
R
,
Di Somma
S
,
Nuti
R
.
Rationale and study design of intravenous loop diuretic administration in acute heart failure: diur-ahf
.
ESC Heart Fail
.
2017 Nov
4
4
479
86
.
14.
Martens
P
,
Dauw
J
,
Verbrugge
FH
,
Nijst
P
,
Meekers
E
,
Augusto
SN
.
Decongestion with acetazolamide in acute decompensated heart failure across the spectrum of left ventricular ejection fraction: a prespecified analysis from the ADVOR trial
.
Circulation
.
2023 Jan
147
3
201
11
.
15.
Palazzuoli
A
,
Ruocco
G
,
Severino
P
,
Gennari
L
,
Pirrotta
F
,
Stefanini
A
.
Effects of metolazone administration on congestion, diuretic response and renal function in patients with advanced heart failure
.
J Clin Med
.
2021 Sep
10
18
4207
.
16.
Kazory
A
,
Sgarabotto
L
,
Ronco
C
.
Extracorporeal ultrafiltration for acute heart failure
.
Cardiorenal Med
.
2022 Nov
17.
Reis
T
,
Ronco
F
,
Ostermann
M
.
Diuretics and ultrafiltration in heart failure
.
Cardiorenal Med
.
2023 Jan
18.
Ronco
C
.
Continuous renal replacement therapy: forty-year anniversary
.
Int J Artif Organs
.
2017 Jun
40
6
257
64
.
19.
Costanzo
MR
,
Ronco
C
,
Abraham
WT
,
Agostoni
P
,
Barasch
J
,
Fonarow
GC
.
Extracorporeal ultrafiltration for Fluid Overload in heart failure: current status and prospects for further research
.
J Am Coll Cardiol
.
2017 May
69
19
2428
45
.
20.
Ronco
C
,
Brendolan
A
,
Sgarabotto
L
.
Rationale and need for simpler and effective miniaturized bedsideultrafiltration devices
.
Cardiorenal Med
.
2022 Dec
21.
Evans
SR
.
Clinical trial structures
.
J Exp Stroke Transl Med
.
2010 Jan
3
1
8
18
.
22.
Stratton
SJ
.
Population research: convenience sampling strategies
.
Prehosp Disaster Med
.
2021 Aug
36
4
373
4
.
23.
Julious
SA
.
Sample size of 12 per group rule of thumb for a pilot study
.
Pharmaceut Statist
.
2005 Oct
4
4
287
91
.
24.
Thabane
L
,
Ma
J
,
Chu
R
,
Cheng
J
,
Ismaila
A
,
Rios
LP
.
A tutorial on pilot studies: the what, why and how
.
BMC Med Res Methodol
.
2010 Dec
10
1
1
.
25.
Lancaster
GA
,
Dodd
S
,
Williamson
PR
.
Design and analysis of pilot studies: recommendations for good practice
.
J Eval Clin Pract
.
2004 May
10
2
307
12
.
26.
Gheorghiade
M
,
Follath
F
,
Ponikowski
P
,
Barsuk
JH
,
Blair
JEA
,
Cleland
JG
.
Assessing and grading congestion in acute heart failure: a scientific statement from the acute heart failure committee of the heart failure association of the European society of cardiology and endorsed by the European society of intensive care medicine
.
Eur J Heart Fail
.
2010 May
12
5
423
33
.
27.
Malbrain
MLNG
,
Martin
G
,
Ostermann
M
.
Everything you need to know about deresuscitation
.
Intensive Care Med
.
2022 Dec
48
12
1781
6
.
28.
Boorsma
EM
,
ter Maaten
JM
,
Damman
K
,
Dinh
W
,
Gustafsson
F
,
Goldsmith
S
.
Congestion in heart failure: a contemporary look at physiology, diagnosis and treatment
.
Nat Rev Cardiol
.
2020 Oct
17
10
641
55
.
29.
Koratala
A
,
Ronco
C
,
Kazory
A
.
The promising role of lung ultrasound in assessment of volume status for patients receiving maintenance renal replacement therapy
.
Blood Purif
.
2020
;
49
(
6
):
643
6
.
30.
Koratala
A
,
Kazory
A
.
Point of care ultrasonography for objective assessment of heart failure: integration of cardiac, vascular, and extravascular determinants of volume status
.
Cardiorenal Med
.
2021
;
11
(
1
):
5
17
.
31.
Safadi
S
,
Murthi
S
,
Kashani
KB
.
Use of ultrasound to assess hemodynamics in acutely ill patients
.
Kidney360
.
2021 Aug
2
8
1349
59
.
32.
Taleb Abdellah
A
,
Koratala
A
.
Nephrologist-performed point-of-care ultrasound in acute kidney injury: beyond hydronephrosis
.
Kidney Int Rep
.
2022 Jun
7
6
1428
32
.
33.
Romero-González
G
,
Manrique
J
,
Slon-Roblero
MF
,
Husain-Syed
F
,
De la Espriella
R
,
Ferrari
F
.
PoCUS in nephrology: a new tool to improve our diagnostic skills
.
Clin Kidney J
.
2023
;
16
(
2
):
218
29
. sfac203.
34.
Dziura
JD
,
Post
LA
,
Zhao
Q
,
Fu
Z
,
Peduzzi
P
.
Strategies for dealing with missing data in clinical trials: from design to analysis
.
Yale J Biol Med
.
2013 Sep
86
3
343
58
.
35.
World Medical Association
.
Declaration of Helsinki: ethical principles for medical research involving human subjects
.
JAMA
.
2013 Nov
310
20
2191
.
36.
Sgarabotto
L
,
Kazory
A
,
Brendolan
A
,
Di Lullo
L
,
Zanella
M
,
Ronco
C
.
The science of extracorporeal ultrafiltration: introducing a novel miniaturized device
.
Cardiorenal Med
.
2023 Feb