Abstract
Intima-media thickness (IMT) provides a surrogate end point of cardiovascular outcomes in clinical trials evaluating the efficacy of cardiovascular risk factor modification. Carotid artery plaque further adds to the cardiovascular risk assessment. It is defined as a focal structure that encroaches into the arterial lumen of at least 0.5 mm or 50% of the surrounding IMT value or demonstrates a thickness >1.5 mm as measured from the media-adventitia interface to the intima-lumen interface. The scientific basis for use of IMT in clinical trials and practice includes ultrasound physics, technical and disease-related principles as well as best practice on the performance, interpretation and documentation of study results. Comparison of IMT results obtained from epidemiological and interventional studies around the world relies on harmonization on approaches to carotid image acquisition and analysis. This updated consensus document delineates further criteria to distinguish early atherosclerotic plaque formation from thickening of IMT. Standardized methods will foster homogenous data collection and analysis, improve the power of randomized clinical trials incorporating IMT and plaque measurements and facilitate the merging of large databases for meta-analyses. IMT results are applied to individual patients as an integrated assessment of cardiovascular risk factors. However, this document recommends against serial monitoring in individual patients.
Introduction
Progressive, subclinical arterial wall alterations precede cardiovascular clinical events, which then reflect advanced atherosclerotic disease. The first morphological abnormalities of arterial walls can be visualized by B-mode ultrasonography. This high-resolution, noninvasive technique provides one of the best methods for the detection of early stages of atherosclerotic disease, based upon its simple nature, wide availability, and demonstrated capacity to depict arterial wall structure with better resolution than any other similar technique (e.g. magnetic resonance imaging or radiographic techniques). Accordingly, many studies have successfully applied intima-media thickness (IMT) as a technique to monitor arterial wall alterations based upon its association with cardiovascular risk factors and the incident cardiovascular disease [1,2,3,4,5,6]. However, these measurements are not strictly associated with individual risks of stroke, myocardial infarction and peripheral artery disease.
Diverse approaches for measuring IMT exist, creating confusion in the literature. In addition, unified criteria are needed for distinguishing atherosclerosis as seen in early plaque formation from thickening of the intima-media complex on the recognition that IMT reflects not only early atherosclerosis, but also nonatherosclerotic compensatory remodeling with largely medial hypertrophy as a result of smooth muscle cell hyperplasia and fibrocellular hypertrophy. This differentiation is important because epidemiological studies have shown that wall thickening as depicted by ultrasonographic measurements of IMT is different from atherosclerotic plaque regarding localization, natural history, risk factors and predictive value for vascular events.
As IMT is increasingly used in clinical trials to serve as a surrogate end point for best medical treatment, the success of interventions that lower risk factors for atherosclerosis can be documented. Further advancement for the field of IMT requires standardized methods, enabling homogenous data collection and analysis. This may help to improve the power of such studies and to facilitate the merging of large databases for meta-analyses.
This update of the last Mannheim IMT Consensus Document [7] addresses the important issue of standardization of carotid IMT measurements and seeks to clarify problems related to the classification of early atherosclerotic lesions.
Distinction between IMT and Plaque
In the absence of atherosclerotic plaques, B-mode ultrasound displays the vascular wall as a regular pattern that correlates with anatomical layers. The area of tissue starting at the luminal edge of the artery and ending at the boundary between the media and the adventitia represents the intima-media portion of this pattern.
This interface is well depicted by ultrasound. With increasing age, this pattern thickens in a uniform way within straight arterial segments. All known major vascular risk factors contribute to increased IMT thickening and its progression. As a mirror of these processes, IMT provides a tool to investigate normal aging and preclinical atherosclerosis. Ultrasound imaging further identifies later stages of atherosclerosis (plaque, stenosis, occlusion) which may also be identified either in the absence of or coincident with increased IMT. However, intermediate stages between increasing IMT and atherosclerotic plaque formation cannot clearly be differentiated neither by ultrasound nor by histological examination.
Such conditions are common at the bifurcation and the origin of the internal carotid artery, but occur only occasionally in the common carotid artery (CCA) where most studies are performed. Epidemiological and intervention studies have shown that although both share common risk factors of atherosclerosis, its natural history, patterns of risk factors and the prediction of cardiac and cerebral events are different for carotid IMT and carotid plaque [8,9,10].
Definitions for Ultrasound Characterization of IMT and Atherosclerotic Plaque
Ultrasound techniques identify:
(1) IMT is a double-line pattern visualized by echography on both walls of the CCA in a longitudinal image. Two parallel lines, which consist of the leading edges of two anatomical boundaries, form it: the lumen-intima and media-adventitia interfaces. This definition complies with the IMT correlation studies comparing this ultrasound pattern to carotid anatomy specimens [11] (fig. 1, fig. 2).
(2) Plaques are focal structures encroaching into the arterial lumen of at least 0.5 mm or 50% of the surrounding IMT value, or demonstrates a thickness >1.5 mm as measured from the intima-lumen interface to the media-adventitia interface.
These definitions permit classification of the vast majority of carotid lesions observed with ultrasound.
Physical and Structural Basics
Carotid examination includes visualization of common, internal and external carotid arteries. Continuous technical progress among ultrasound devices now provides higher spatial and density resolutions. Ongoing technical progress includes the emergence of three-dimensional imaging.
Ultrasound Principles of Physics and Technology
– The virtual biconcave lens of an ultrasound beam with its thicker nature on the near and distal part of the field leads to the optimal resolution in the mid part of this field.
– Establishing the perpendicular relationship between the ultrasound beam and the visualized structures provides optimal reflection of the incident ultrasound beam.
– The energy of the ultrasound beam decreases with the depth of insonation making the far field darker.
– The probe frequency has a direct relationship to resolution, and an inverse relationship to depth of ultrasound penetration.
– Linear transducers activate piezo-electric crystals simultaneously resulting in a synchronized propagation of the ultrasound wave. Sector scanners are typically activating crystals sequentially in predefined time intervals, causing asymmetric shifts and reception of skewed wavefronts. Linear ultrasound transducers provide best image quality for superficial arteries, while electronic sector scanners are better used for deep structures with less accessibility.
– B-mode signals which are displayed on the screen derive from the received raw radiofrequency signals of the probe.
Anatomic Structures and Physiological Principles
– Common and internal carotid arteries are homogeneous in structure and hemodynamics.
– Insonation is limited at the origin by the brachioarterial cephalic trunk on the right side and the aortic arch on the left side, the CCA ends at the bifurcation that is represented anatomically by the point of divergence of the CCA walls (fig. 1).
– The bifurcation is rather heterogeneous between individuals. At this site, remodeling varies according to the angles, curvatures and diameters of carotid arteries that influence shear stress constraints and modify wall thickness independently of any arterial disease or risk factor [12,13].
– Since atherosclerosis usually starts in the carotid bulb, delineation of the different segments is required.
Standards for Image Acquisition
Intima Media Thickness
Carotid arterial wall assessment may include the common, internal or bulb segments of the carotid artery. Whereas nearly all patients have their CCA imaged, successful imaging of the internal carotid artery and of the carotid bulb depends both upon the anatomical topography of the patient and on sonographer’s expertise.
Settings of the Ultrasound Device
(1) Standard equipment includes a high-resolution B-mode system operating in the black and white mode, preferentially with linear ultrasound transducers at frequencies >7 MHz.
(2) Appropriate depth of focus (e.g. 30–40 mm), and optimal frame rate 25 Hz (>15 Hz) provide optimal image quality and facilitate edge detection.
(3) Log gain compensation of approximately 60 db.
(4) Gain settings adjusted to obtain a symmetrical brightness on the near and far wall, or in the mid part of the field to eliminate intraluminal artifacts.
Examination Procedure
(1) Arterial wall segments should be assessed in a longitudinal view, strictly perpendicular to the ultrasound beam.
(2) Both walls should be clearly visualized in order to achieve diameter measurements. The optimal diameter should be obtained during diastole by automatic cine-loop detection or by looking for the minimal diameter during the cardiac cycle (fig. 4).
(3) Measurements obtained from radiofrequency signals are equivalent to those obtained with conventional video signals [14,15].
(4) Lateral probe position is recommended as it offers the best resolution in the midfield.
(5) Acquisition of multi-insonation angles increases procedure time consumption without benefit for measurements; however, in clinical trials they result in better reproducibility and facilitate statistical analysis [16].
(6) Horizontal arterial image display optimizes the interfaces between blood and vessel wall structures.
(7) Imaging of the carotid bifurcation provides a landmark essential in serial imaging.
(8) Longitudinal and cross-sectional views are required to visualize focal atherosclerosis.
Plaque acquisition should be done along the carotid tree, in longitudinal and cross-sectional views. Color flow Doppler imaging helps to identify low echogenic boundaries.
Measurement Location
Measurement of IMT should occur within a region free of plaque with a clearly identified double-line pattern. Such measurement will have increased accuracy, reproducibility [17], and ease of measurement using computer-based techniques. Such measurements will match a majority of reference values within studies performed throughout the world. The majority of reference values provide data from different countries [18,19,20].
This pattern is the only one that has been validated with anatomy [11].
(1) IMT should be measured preferably on the far wall of the CCA at least 5 mm below its end (fig. 3, 4, 5) which avoids interindividual variability induced by physiological remodeling and less gain dependence will exist. Values from the near wall depend in part on gain settings and are less reliable. If taken on the near wall, its value should be recorded separately from IMT of the far wall.
(2) IMT can be measured at the carotid bifurcation or ICA bulb in a region free of plaque, on a shorter length, taking caution of the large interindividual variability induced by remodeling and anatomic variations. These values must be recorded separately [21]. In study designs that include wall thickness, values obtained from different sites of the carotid arteries should be documented separately.
(3) Along 10 mm length of a straight arterial segment, a high-quality image acquisition is required for reproducible serial measurements. In case of vessel tortuosity, IMT measurements are only possible at a shorter vessel segment, especially in the carotid bifurcation or the ICA bulb.
(4) For plaque: Maximal thickness requires demonstration from 2 different angles of insonation, in longitudinal and cross-sectional views.
Methodology Standards
IMT
(1) Edge detection systems provide accurate measurements of IMT [22]. Although manual measurements may lead to valid measurements, they require rigorous quality control and quality assurance. Manual and semimanual reading methods require more time than automated systems. Automated systems can provide the mean maximal value of 150 measurements performed on a 10-mm segment of CCA instantaneously.
Manual measurements are more observer dependent than semi-automatic systems.
(2) CCA inter-adventitial and intraluminal diameters must also be measured as IMT is significantly correlated with arterial diameter [23].
(3) IMT measurements options include the mean, maximum, composite measures from both sides and different arterial sites. Mean IMT values averaged across the entire distance are less susceptible to outliers, whereas the maximal IMT may reflect more advanced stages with focal thickening or plaque formation. The maximal value may be misleading due to the effect of measurement and sampling errors.
IMT values from the left and right side can be averaged although there is a significant difference between the left and right CCA IMT, with higher values on the left side [24]. Composite scores including both plaque and IMT should be avoided.
(4) Each vascular laboratory must perform quality control of their equipment periodically by (phantom scans) and reliability studies of scans and measurements for ultra-sonographers and readers. Reporting requirements include the intra-class correlation coefficients to be evaluated for intra- and interobserver variability, both for IMT and plaque measurements.
Plaque
(1) Plaque location, thickness, area and number, scanned in longitudinal and cross-sections must be recorded. The incremental value of recording texture (density, echogenicity, shadow) remains uncertain pending more research.
Guidelines and Use in Clinical Practice
Guidelines that support the use of IMT for the assessment of cardiovascular risk include: ACC/AHA Guideline on the initial detection of cardiovascular risk: Recommends assessment as a class IIA (‘is reasonable to perform’) in individual at intermediate coronary heart disease risk [25].
The National Cholesterol Education program Adult Treatment Panel III identified IMT as a method to detect subclinical atherosclerosis and guide selection and intensification of lipid management [26].
The ESH/ESC European practice guidelines in Spain include IMT to detect target organ damage induced by hypertension [27].
The American Society of Echocardiography [28] and the Society of Atherosclerosis Imaging and Prevention [29] published appropriate use criteria.
CIMT and plaque presence are recommended for the initial detection of CHD risk in asymptomatic patients:
– at intermediate risk
– in the setting of 2 or more NCEP risk factors
– with metabolic syndrome
– in the setting of a family history of premature CHD
– with a known CAC score of zero and FRS 11–20%.
Reference Values and Societies Guidelines
The differences observed in IMT values in a healthy population from different countries are related to many factors, including methodological variability and differences in cardiovascular risk and factor profiles [30,31,32].
Since 2000, several guidelines suggest that carotid artery B-mode ultrasound imaging is safe, noninvasive and relatively inexpensive allowing to assess subclinical atherosclerosis [33,34,35] in asymptomatic persons >45 years old and could add incremental information to traditional risk factor assessment [36,37,38,39]. Reference values require matching to specific imaging protocols utilized in large sample size population as much as possible.
Risk calculators may integrate IMT measurement with risk factor measurements. The ARIC study has published such a calculator incorporating IMT and plaque assessment to determine an adjusted Framingham risk. CCAIMT ‘normal values’ in the absence of plaque should help to characterize populations at intermediate risk.
Use of IMT in Research
Epidemiological and interventional trials evaluating cardiovascular diseases should include IMT and plaque to better characterize the study population.
Ancillary studies in research should be sufficiently powered to evaluate the predictive value of IMT and plaque for incident clinical events.
In clinical trials, several additional requirements are recommended to assure data harmonization. They can be performed through an expert IMT/Plaque panel as usually represented in the Steering Committees [40].
Clinical trials using IMT and plaque as intermediate outcomes need to address the following:
– Study design.
– Study quality control.
– Full image monitoring.
– Statistical plan analysis.
– Data monitoring (from CRO and Core Lab).
– Result analysis and definition:
(1) Inclusion criteria for the participating ultrasound centers.
(2) Equipment installation if necessary.
(3) Sonographers training and performance control.
(4) Certification of the centers (equipment and sonographers).
(5) Quality assurance and quality control.
(6) Central reading facilities.
– Evaluation of data variability through intra-class correlation coefficient.
– Choice of end points depends on:
(1) Primary or secondary prevention status.
(2) Epidemiological or interventional studies.
IMT and plaque measurements including maximal or mean IMT, plaque thickness, area and volume, and plaque score may all be useful as imaging outcomes.
All these procedures and choices can reduce measurement variability, which is a key parameter for a high-quality study, statistical power and sample size determination.
Conclusions
Carotid IMT and plaques are different phenotypes indicating increased vascular risk. Plaque presence demonstrates a higher risk and therefore overrides IMT predictive values. However, IMT without plaque remains a significant marker of an increased risk of vascular events and significantly predicts plaque occurrence. The continuity of vascular wall changes is best monitored in CCA IMT studies, different from discontinuous focal lesions (plaque) which are characteristic of atherosclerotic disease. Therefore, the distinction between IMT and plaque must be clearly specified in the scanning protocols.